TurboMass Software User’s Guide
Report Generation Window
In this window you can select the Forms you want printed and the Sample List(s) to
be used as input. The selection of Tentatively Identified Compounds from the initial
qualitative processing will also be made within this window. This window displays a
comprehensive view for generating reports using the EPA Forms 1 through 8 that
apply to GC/MS analyses.
The process of generating the Forms will involve the reprocessing a Sample List by
the TurboMass software (as with other Communiqué reports) but there are several
unique aspects in this case:
1. The Sample List is queued for reprocessing from this window rather than
from the Sample List window.
2. The Start Sample List Run dialog does not appear. The reprocessing always
uses the following options:
Auto Quantify Samples = On (Quantify Samples only)
Generate Communiqué Reports = On
Quantify, Qualify, and Generate Reports = After Each Run (or summary
reports at the end of the sample list)
3. Which sample rows are processed will be indicated by the user on the
Sample List tab. Individual rows can be turned on or off for processing.
4. The Report Method supplied to the Report Manager (and hence the
template(s) provided to Communiqué will not be that in the Report Method
column of the Sample List but will instead depend on the Forms selected by
the user and the Submitter and Task associated with the sample. The Report
Method in the Sample List can be execute in the usual manner (Run/Start
from the Sample List).
596