Appendix D - Essential performance Power Input 100-240V ac Ultrasound Frequency Maximum intensity Maximum output power Output modes Pulse duration Treatment timer Large transducer Small transducer 34 1.1 MHz ±5% and 3.4 MHz ±5% 1.5 W/cm2 (±20%) in CW 3.0 W/cm2 (±20%) pulsed modes 6 W (±20%) average CW and pulsed 1:1, 1:2, 1:4 and 1:9 2 ms (±10%) 0 to 30 minutes ERA 4cm2 (±20%) (IEC61689), collimated BNR<5 ERA 0.6cm2 (±20%) (IEC61689), divergent at 1MHz ERA 0.4cm2 (±20%) (IEC61689), collimated at 3MHz BNR<5 Primo Therasonic 360 & 460