WindSonic M
Doc No 1405 PS 0030 Issue 1
August 2012
Output rate (not displayed)
WindSonic M
Doc No 1405 PS 0030 Issue 1
August 2012
9.2 Gill format – UV, Continuous
The WindSonic M samples continuously and delivers wind information at ¼, ½, 1 (default
setting), 2, or 4 outputs / second.
In this mode, the output is given as signed (i.e. positive or negative) speeds along the
‘U’ (= South – North) axis and the ‘V’ (= East – West) axis.
Status
This indicates either
Correct operation Code 00 and A (NMEA).
or Error codes See Section 11.5 for explanation of codes.
Checksum
This enables the host system to check that the data has been correctly received. This is the
EXCLUSIVE – OR of the bytes between (and not including) the <STX> and <ETX>
characters.
<STX>Q, +001.59, - 002.74, M, 00, <ETX> 2D
WindSonic
node
address
U axis
speed and
polarity
Where:
<STX>
WindSonic M node address
U axis
V axis
Units
Status
V axis
speed and
polarity
Units
=
=
=
=
=
=
Status
Check
sum
Start of string character (ASCII value 2)
Unit identifier
speed & polarity
speed & polarity
Units of measure (knots, m/s etc)
Anemometer status code (see Section 12.5
for further details)
<ETX>
= End of string character (ASCII value 3)
Checksum
= This is the EXCLUSIVE – OR of the bytes
between (and not including) the <STX>
and <ETX> characters
The WindSonic M unit identifier, Units, and Checksum are as described in Section 0 above.
Figure 5 shows the polarity of U and V if the wind components along the U and V axis are
blowing in the direction of the respective arrows.
Figure 5 UV Polarity
30
31