Checking the Calibration Span, Continued Checking Calibration Span Follow the steps in the table below to check the Span of the EPAM-5000 Monitor. Note: Failure to follow this procedure in its entirety may cause an incorrect “k” value reading. Step 1 2 3 4 5 6 7 8 Action Turn monitor on. Run monitor on a 1 (one) second sample rate for (5) minutes. See Chapter 3 Page 3-9 Sampling. With unit running select down arrow from keypad to activate Auto Purge. An “X” will appear in the upper right hand corner to indicate sensor is purging. See Chapter 3 Page 3-6 Auto Zero Purge if more detailed information is needed. After Auto Purge is complete (x is no longer present on screen) select enter to stop monitor from running. Performing the Span Check. Step Action 5-1 Select Span Check from Main Menu. 5-2 Select Yes to proceed with span check (all data will be lost) or select no if you want to download stored data. 5-3 After selecting yes calibrate sensor will appear. Scale factor must reset to 1000. If not press up or down arrow as needed. Remove Auto Zero filter and impactor sleeve. Insert the span reference into the sensor head (see diagram 5.1 on page 5-2). Allow 2-3 minutes for reading to stabilize. Note: Be sure the locating pin on the calibration reference slides into the locating hole on the sensor head. Also Push down on the calibration reference to be sure it is aligned properly. Slightly twist CS-105 clockwise when positioned in alignment hole. This will ensure the CS-105 is in the exact position and ensure reproducibility when “K” value is displayed on the EPAM-5000 monitor. Observe the printed “k” value on the calibration reference. Note: The printed “k” value should match the concentration value shown on the EPAM-5000 LCD to within + 10%. Continued on next page 5 - 3