A.3 MS FAQs
153
A.2 COLUMN FAQs
1. Which type of column should I use in my methods development?
Obviously, the liquid phase depends on what you are trying to separate.
Protocols usually specify the type of the column and its working phase.
Since you specified methods development, I would select a bonded-phase
capillary column. Bonded-phase columns have a longer lifetime with
less column bleed into the MS detector. Bonded-phase monolith columns
are touted as the column of the future, but they still need development
work. Capillary columns can be shortened for cleaning the column head
and coiled and squeezed to fit temperature-controlled column heaters.
2. Why can’t I use column switching as a control variable in GC like I can
in HPLC? Column switching in conventional GC systems requires
disconnecting the interface to the mass spectrometer and breaking the
vacuum. Vacuum restoration can take up to 4 h. Aligent has announced
a microfluidic device called QuickSwap to divert carrier gas flow away
from the mass spectrometer allowing the vacuum to be maintained
while the column is replaced. This is controlled by their EPC system,
but if it can be adapted it may offer a general method for rapid column
switching in GC/MS methods development.
A.3 MS FAQs
1. Which mass spectrometer is best; quadrupole, ion trap, or time-offlight? Each analyzer has its own strength. Quadrupoles were the first
mass spectrometers used for chromatography, still represent the bulk of
laboratories systems, and are the least expensive. Look for a bench-top
quadrupole that can give you chromatograms labeled with peak
molecular weights. Linear ion trap analyzers are the most flexible
allowing you to do simple molecular weights as well as precursor and
product fragmentations studies. They also can store enough ions to give
excellent sensitivity for trace analysis and can trap specific ions for
fragmentation analysis. For large molecule studies you are going to
want a time-of-flight system with an electrospray interface.
2. How high of a vacuum should I be getting? You will need at least
105 Torr before you begin running chromatography and should see
106 Torr on a clean system with a turbo pump. I have talked to people
who get 107 Torr with an oil diffusion pump and clean oil. Sometimes
you need to push down on the lid on the mass spectrometer containment