Unable to operate the panel switch.
Symptom
Unable to operate the
panel switch.
Check Items
See
Is the LOCK LED illuminated?
Release the panel operation lock.
Is the REMOTE LED illuminated?
To control the PAT from the panel, press
the LOCAL switch to enable local mode
Is the PAT being controlled via the
RS232C, GPIB, USB or LAN
interface?
The PAT does not switch to Was a local lockout (llo) specified to
local mode even when I
the communication interface?
press the LOCAL switch.
A-8
Remedy
Use a communication command to
clear the local lockout (llo).
3-24
3-25
3-25