SD trouble
北
京
Is the DAS
power supply voltage
proper?
NG
Defective DAS
power supply
司
公
限 cn
有 m.
展 co
发 a. 54
术 ti 4/
技 en /4
捷 ig 34
安 il 40
勤 .d 00
天 ww -65
w
0
OK
OK
REF ch NG (>50)
X-ray optical system in
normal operation?
01
NG
Trouble in the
X-ray optical
system
Offset SD checked by DCA
Offset SD
checked with the
REF detector cable
disconnected
OK
OK
Defective REF
detector
NG
Exchange the INTEG PWB in the
first slot with a PWB in another
slot and perform check.
OK
Defective INTEG PWB in slot 1
Check the outputs of the REF
detector and the main detector
using the DCA.
Is the mean value for 21 channels within
the range of the standards?
OK
NG
NG
Trouble in
X-ray system
X-ray tube current OK?
OK
Trouble in the reconstruction system
and scan control
Figure 2.5-4 SD troubleshooting flow
No. 2D201-103EN*E
113
NG
Trouble in the
main detector