View the display section
of the thermoregulator.
NO
京
北
Defective heat controller,
heater, or thermal sensor
L or LL size
39°C to 41°C
YES
司
公
限 cn
有 m.
展 co
发 a. 54
术 ti 4/
技 en /4
捷 ig 34
安 il 40
勤 .d 00
天 ww -65
w
0
Check the FOV.
SS, S, or M size
NO
Is the FMD
output normal?
01
YES
NO
Proper FMD
alignment?
Check the appearance of
slits (wedge slit, beam
trimmer slit).
Realign the FMD
YES
Problem in the FMD
Wedge slit
Flaw ont beam trimmer
Slit flaw?
Trouble in X-ray optical
system
Defective beam trimmer
No flaw
NO
In proper
alignment?
For an INTEG PWB that
has an abnormal channel,
exchange it with the
adjacent normal PWB and
check the data again.
Improper alignment
YES
YES
NO
Error on a
few channels?
Data
transmitted right in
TEST mode?
YES
Error channels
moved?
YES
Erroneous
data every 16
channels?
Defective main detector
NO
Failure of the
OPCONT2 PWB or
subsequent section,
or of the TIMING I/O
PWB, or of cables
for each unit, or
connector failure
YES Defective INTEG
PWB
NO
NO
Are offset and
offset SD
normal?
Defective INTEG PWB
YES
Defective TIMING I/O PWB
or FPA/ADC PWB or main
detector. Or insufficient
dose (increase mAs and
perform image acquisition)
Figure 2.5-1 Ring image troubleshooting flowchart
No. 2D201-103EN*E
110
NO Defective INTEG
PWB