Isolate the error
channel by DCA
and sinogram
Channel
dependent?
NO
Data error
present?
NO
YES
YES
01
Four
consecutive
error
channels?
Reference
channel
(Channel R)?
NO
NO
Move the faulty
INTEG PWB to
another position.
Fanbeam-like
streak?
YES
Defective cables
connecting units
YES (SS/ADI2, KGTS,
GCIF2,
OPCONT2) or
their connectors
NO
Bit dependent?
NO Defective
YES
Defective TIMING
I/O PWB or I/O
cable or
OPCONT2 PWB
YES
YES
Channels
moving?
Defective
INTEG
PWB
Defective main
detector
NO
OK
Check the offset
data with the
gantry stopped.
Exchange the INTEG
PWB in the first slot
with a PWB in another
slot and perform
check.
DAS supply
voltage OK?
Offset data
abnormal?
YES
NG
Adjust supply voltages.
Check the I/O
and GND cables.
NO
Defective main detector
Defective FPA/ADC
PWB or TIMING I/O
PWB
Did the
YES
defective channel
move?
NO
Defective REF
detector (or its
output cable)
Defective
INTEG PWB in
slot 1
Figure 2.5-2 Streak shower troubleshooting flow
No. 2D201-103EN*E
111
OPCONT2
or SS/ADI2