Download User Guide
Transcript
Finetuning ● Replace the eyepiece with your CCD camera ● run continuous exposures (around 1sec?) eventually in binning 2x2 or 3x3. ● Position the slit to the center of your image by adjusting: ○ doublet focus if you do not see the slit (changing from eyepiece to camera requires some adjustment) ○ grating angle ○ main mirror angle ● Playing with those three adjustments, you should quickly find the ideal positions. ● Eventually, try to put the slit closer to one edge to avoid vignetting. Shift the mirror screw slightly toward the slit trappe so the slit image is at 1/3rd of the CCDimage. This way, you optimize the position and reduce risk of vignetting. This is not mandatory but good to do... Note where your spectrum is on the CCD - this is where you will have to find your star. Main mirror adjusting screw Screw high Screw low Screw nominal Zero order Lhires III – User Guide (English) – DC0004A 17 / 43