Trouble Shooting Program Start After starting OptoControl the program tries to communicate with the streak camera. In case the communication is not possible the box below appears. Make sure the streak camera is in Standby mode or On mode. Verify the correct RS-232 cable or network cable is used and plugged correctly. Hint: If a peer-to-peer connection between the streak camera and the PC is used, make sure to use a cross-over network cable. Also verify whether the OptoAnalyse software had been installed in a multi camera configuration. In this case the INIT.INI file contains the serial number of the streak camera. INIT.INI file configuration and physically connected device have to fit. IP and Mask of Streak Camera unknown If no communication with the camera is possible verify the correct IP setting of the OptoControl Program. Set the IP and Mask according to the indications on the rear of the streak camera main unit. IP Conflict An IP conflict might appear in case the static IP of the streak camera is used by another device in the same network. Consult the network administrator to prevent other devices to use the static IP of the streak camera. Alternatively a peer-to-peer connection can be used to avoid any conflicts. Password Unknown If no communication with the camera is possible verify the correct IP setting of the OptoControl program. Set the IP and Mask according to the indications on the rear of the streak camera main unit. Two Ethernet Interfaces on the PC If more than one Ethernet interface is installed in the PC make sure to use different sub-networks as underlined below. Example: IP1: 192.254.127.254 OptoAnalyse Ver. 3.50 Ref. 3991-SU-01-I Page 100