Download Field Service Manual Rev-01
Transcript
Medtronic CryoCath LP Technical Reference Manual (For Hospital Biomedical Engineers) TRM-002 MDT.2 CryoConsole Rev-00 Manufacturer: Medtronic CryoCath LP 16771 Chemin Ste-Marie Kirkland, Quebec, Canada H9H 5H3 Tel: (514) 694 1212 Fax: (514) 694 7075 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 TABLE OF CONTENTS SECTION I I.1 I.2 I.3 I.4 I.5 I.6 PURPOSE ........................................................................................................................................... 5 SCOPE ............................................................................................................................................... 5 DEFINITIONS AND ACRONYMS ......................................................................................................... 5 WARNINGS/PRECAUTIONS ............................................................................................................... 6 GENERAL DESCRIPTION ................................................................................................................... 7 TECHNICAL SPECIFICATIONS ......................................................................................................... 13 SECTION II II.1 II.2 II.3 II.4 II.5 II.6 II.7 GENERAL INFORMATION....................................................................................... 5 FUNCTIONAL SYSTEM DESCRIPTION .............................................................. 17 PURPOSE .................................................................................................................................... 17 SCOPE ........................................................................................................................................ 17 DEFINITIONS AND ACRONYMS................................................................................................... 17 SYSTEM OVERVIEW ................................................................................................................... 17 SOFTWARE INTERFACE .............................................................................................................. 18 MECHANICAL SYSTEM............................................................................................................... 20 ELECTRICAL SYSTEM................................................................................................................. 21 SECTION III OPERATION AND CLINICAL SEQUENCE ......................................................... 22 SECTION IV TROUBLESHOOTING AND MAINTENANCE..................................................... 22 IV.1 IV.2 IV.3 IV.4 IV.5 PURPOSE .................................................................................................................................... 22 SCOPE ........................................................................................................................................ 22 DEFINITIONS AND ACRONYMS................................................................................................... 23 MAINTENANCE .......................................................................................................................... 23 SOFTWARE INTERFACE .............................................................................................................. 27 APPENDIX A– SYSTEM NOTICES...................................................................................................... 31 APPENDIX B– ADDITIONAL WARNING MESSAGES ................................................................... 39 APPENDIX C– GUIDELINE & SPECIFICATIONS FOR N2O REFILLING .................................. 40 Medtronic CryoCath LP Page 1 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Index of Figures Figure 1 - The CryoConsole Gen III and Accessories ..................................................................7 Figure 2 - The CryoConsole Gen IV and Accessories ..................................................................8 Figure 3 - North American tank and European tank .....................................................................8 Figure 4 - The CryoConsole Gen III features ..............................................................................10 Figure 5 - The CryoConsole Gen IV features .............................................................................10 Figure 6 - Foot switch connector.................................................................................................11 Figure 7 - Main panel ..................................................................................................................23 Figure 8 - System Setup panel ...................................................................................................24 Figure 9 - Password panel ..........................................................................................................24 Figure 10 - System Maintenance panel ......................................................................................25 Figure 11 - Password Setup panel..............................................................................................25 Figure 12 - Date & Time Setup panel .........................................................................................25 Figure 13 - Site Setup panel .......................................................................................................26 Figure 14 - Backup Patient Files panel .......................................................................................26 Figure 15 - Floppy Drive .............................................................................................................27 Figure 16 - Message List ............................................................................................................28 Figure 17 - Danger and Warning System Notice ........................................................................29 Figure 18 - Caution and Notice System Notice...........................................................................29 Medtronic CryoCath LP Page 2 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Index of Tables Table 1 - CryoConsole Specifications.........................................................................................13 Table 2 - Accessories Specifications ..........................................................................................15 Table 3 - Notice series ................................................................................................................30 Medtronic CryoCath LP Page 3 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 THIS MANUAL DOES NOT REPLACE THE OPERATOR’S MANUAL, ITS APPENDICES, AND THE INSTRUCTIONS FOR USE FOR EACH RESPECTIVE CATHETER. THE TERM CATHETER IS USED THROUGHOUT THIS MANUAL TO DESCRIBE BOTH CARDIAC CRYOABLATION CATHETERS AS WELL AS SURGICAL CRYOABLATION DEVICES. Medtronic CryoCath LP Page 4 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Section I I.1 Rev-00 General Information Purpose The purpose of this section is to give general information of the CryoConsole functionality, characteristic and accessories available. I.2 Scope This manual applies to the Medtronic CryoCath LP MDT.2 CryoAblation System Gen III (North American Ref#103A2, European Ref#103E2) as well as Gen IV system (North American Ref#104A2, European Ref#104E2). The CryoAblation Data Manager versions supported by this manual are 3.03 and 3.04. I.3 Definitions and Acronyms The following definitions and acronyms are used in this section: MDT IEC ISO UL TLV NIOSH ECG CSA MDD LRQ 7F/9F CE Sccm PSIa KPa Medtronic CryoCath LP Medtronic CryoCath LP International Electro technical Commission International Organization for Standardization Underwriters Laboratories Threshold Limit Value National Institute for Occupational Safety and Health Electrocardiogram Canadian Standards Association Medical Device Directive Lois Refondues du Québec 7 French / 9 French Conformité Européenne Standard Cubic Centimeters per Minute Pound per Square Inch Absolute Kilopascal Page 5 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV I.4 Rev-00 Warnings/Precautions DO NOT ATTEMPT to service the MDT.2 CryoConsole. Only technicians, fully trained by MDT, can service the CryoConsole. Training courses and custom made tools are available at MDT. Only the catheters, accessories, gas cylinders and spare parts that have been obtained from Medtronic CryoCath LP are to be used with the CryoConsole. Even though the CryoConsole has passed the requirements of IEC 601-1 (UL report 75HN) it is the user’s responsibility after installation and maintenance to verify and ensure that the CryoConsole meets the applicable electrical safety requirements. Electrical Safety Test Test performed by: Print Name Medtronic Field Engineer Hospital BioMedical Engineer Signature Date (yyyy.mm.dd) Nitrous Oxide Usage ¾ The MDT.2 CryoConsole operates on high-pressure liquid N2O delivery. Store and handle the N2O cylinder according to hospital gas management policy. Make sure to know their location for future needs. ¾ When using the CryoConsole during a procedure or during servicing, Nitrous oxide (N2O) must be safely vented through the connected scavenging according to policies & procedures of each hospital in each country. Refer to the MDT.2 CryoConsole Operator’s Manual for proper venting as described in Section IV – CryoConsole Pre-procedure set-up. ¾ Use only N2O tanks supplied by Medtronic CryoCath LP. ¾ Use only Medtronic CryoCath LP supplied scavenging hose. ¾ Do NOT restrict or kink the scavenging hose. ¾ Do NOT disconnect the catheter or the scavenging hose from the MDT.2 CryoConsole until ALL N2O is safely vented from the system (a few seconds after the end of the last injection). ¾ Cylinders are NOT for use on any other equipment. Medtronic CryoCath LP Page 6 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 ¾ Threshold Limit Value (TLV) for N2O is 25ppm as recommended by the National Institute for Occupational Safety and Health (NIOSH). ¾ Exposures to N2O should be minimized to prevent short-term behavioural and long-term reproductive health effects. ¾ Do not open the N2O gas tank valve until it is properly connected and tightened to the CryoConsole. ¾ Do not disconnect the high-pressure lines from the cylinder until the pressure gauges read ‘0’ psi. Close the N2O gas tank valve and allow the N2O to safely vent from the system and lines, by turning the vent valve on the tank side couterclockwise, before disconnecting the high-pressure lines. I.5 General Description The Medtronic CryoCath LP CryoAblation System consists of three principal components: ¾ ¾ ¾ MDT.2 CryoConsole with CryoAblation Data Manager software The selected CryoAblation catheter Supplies and Accessories, including: o Sterile/single use Coaxial Umbilical o Sterile/single use Electrical Umbilical o Non-sterile Auto Connection Box o Non-sterile ECG Cable o Gas Scavenging Hose o Refrigerant Tank (N2O cylinder) o Power Cord o Foot switch (Optional) Figure 1 - The CryoConsole Gen III and Accessories Medtronic CryoCath LP Page 7 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Figure 2 - The CryoConsole Gen IV and Accessories Note: Figure 1 and Figure 2 show the North American set-up (Nitrous Oxide tank) The European tank differs from the North American tank in size, weight and capacity. Figure 3 shows the visual difference between the two. North American tank European tank Figure 3 - North American tank and European tank The CryoConsole only operates in temperatures between 15°C and 30°C (60°F and 86°F), an altitude ranging from sea level to 2400 meters (8000 feet), and at a maximum humidity of 75%. Medtronic CryoCath LP Page 8 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 I.5.1. Description of the System MDT.2 CryoConsole The CryoConsole Gen III reference (103A2 and 103E2) is a type CF, 115/230v, 50/60Hz, 10amp, with respect to electric shock, fire, mechanical hazards in accordance with UL 2601-1: 1997 and CAN/CSA C22.2 No. 601.1, control number 75HN and IEC #601-1. The CryoConsole Gen IV model (104A2 and 104E2) is a type CF, 115/230v, 50/60Hz, 10amp, with respect to electric shock, fire, mechanical hazards in accordance with CAN/CSA C22.2 No. 601.1-M90 file #220418 and IEC #601-1:1998. Medtronic CryoCath LP recommends that the biomedical engineer of the site perform regular electrical leakage current measurements in order to ensure that the system leakage current does not exceed the allowable current as required by the applicable standards (IEC #601-1). The CryoConsole houses the electrical, mechanical components and the software for controlling and recording the CryoTherapy procedure. It stores and controls delivery of liquid N2O under high pressure through the coaxial umbilical to the catheter, recovers the refrigerant vapour from the catheter under vacuum, and disposes the refrigerant through the hospital scavenging system. Pressurized liquid N2O is delivered from the CryoConsole through an ultra fine, robust injection tube, to the tip of the catheter Inside the tip, the liquid N2O vaporizes as it encounters heat from the surrounding tissue Medtronic CryoCath LP The CryoConsole has numerous safety systems to prevent any potential hazards The warmed vapor is returned to the CryoConsole through a lumen maintained under vacuum Page 9 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Figure 4 - The CryoConsole Gen III features Pressurized liquid N2O is delivered from the CryoConsole through an ultra fine, robust injection tube, to the tip of the catheter Inside the tip, the liquid N2O vaporizes as it encounters heat from the surrounding tissue The CryoConsole has numerous safety systems to prevent any potential hazards The warmed vapor is returned to the CryoConsole through a lumen maintained under vacuum Figure 5 - The CryoConsole Gen IV features The hardware controls the safety monitoring system while the software provides the interface to the patient information, procedure temperature, temperature set point, time set point and procedure data information. Numerous safety systems exist to control the safe delivery of liquid N2O to the catheter and the return of the vapour. 1. Inside the catheter Vacuum System ....... The shaft is always under vacuum unless blood is detected. Leak Detection .......... The leak detector detects blood ingress. 2. At the catheter handle Blood Detection......... An infrared sensor inside the handle detects blood. 3. Inside the CryoConsole Flow Detection .......... The system monitors the flow of the returned N2O vapour and will shut off the injection if the flow is outside the parameterized range. Fluid Detection .......... An ultrasound detector will disable the system if fluid is detected inside the console vacuum line. Vent System.............. The vent system sucks back the refrigerant. Medtronic CryoCath LP Page 10 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 When pushing the “Injection On” button for 2 second, the button lights up green, and the refrigerant injection begins. Injection stops when the red “Injection Off” button is pressed, when the timer runs out or when a system notice is triggered. I.5.2. Accessories The accessories include a sterile single use coaxial umbilical, a sterile single use electrical umbilical, an auto connection box and an ECG cable. During a medical intervention, the nonsterile accessories must be handled in such a way not to introduce them into the sterile field. The electrical umbilical provides the electrical connection between the CryoConsole and the catheter through the auto connection box. The coaxial umbilical provides the refrigerant during injection and recovers the vapour from the catheter to the CryoConsole. The auto connection box and the ECG cable provide an interface between the catheter and a standard EP system. The auto connection box has a mechanical relay that automatically connects the thermocouples upon injection and disconnects them when injection stops and the temperature at the distal end of the catheter reaches 32°C or above. The purpose of this feature is to reduce the noise superimposed on the intra-cardiac signal recording. I.5.2.1. Optional Accessory The foot switch serves as a remote control to start and stop the injection. The injection will start if the foot switch is pressed for more than 2.5 seconds. To stop the injection, the switch must be pressed for 0.5 second. The foot switch connector is located on the bottom back of the CryoConsole as shown in Figure 6. Using the foot switch does not change the functionality of the injection button. The reference number for the Gen III footswitch is 103FS and the reference number for the Gen IV footswitch is 104FS. Foot switch connector Figure 6 - Foot switch connector I.5.2.2. Connection Sequence Connect the scavenging hose, the power cord and the optional foot switch (if applicable). To connect the accessories in preparation for a case, first connect the non-sterile accessories to the CryoConsole: ¾ ¾ Auto connection box ECG Cable Medtronic CryoCath LP Page 11 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Then introduce the sterile items into the sterile field and connect them from the sterile field: ¾ ¾ ¾ Sterile electrical umbilical Sterile coaxial cable CryoAblation catheter Under no circumstances should any sterile coaxial or electrical umbilical be resterilized or re-used. All sterile products, once used, are to be discarded with accordance to hospital procedures. Should the customer or Medtronic CryoCath LP Technologies request that any accessory be returned to Medtronic CryoCath LP for inspection, it must be disinfected, cleaned and packaged as per Medtronic CryoCath LP Quality guidelines. A Return Goods Authorization (RGA) number is necessary and is distributed by Medtronic CryoCath LP Customer Service. Non-sterile accessories should be cleaned with a dry cloth and stored with the CryoConsole after use. Only accessories provided by Medtronic CryoCath LP can be used with the CryoConsole. Do not expose catheter or accessories to fluids or solvents. Medtronic CryoCath LP Page 12 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV I.6 Rev-00 Technical Specifications MDT.2 CryoConsole (REF# 103A2, 103E2, 104A2, 104E2) Regulatory Approvals: The MDT.2 CryoConsole Gen III reference (103A2 and 103E2) is a type CF, 115/230v, 50/60Hz, 10A, with respect to electric shock, fire, mechanical hazards in accordance with UL 2601-1:1994, CAN/CSA C22.2 No. 601.1, IEC 601-01, and control number 75 HN. The CryoConsole Gen IV reference (104A2 and 104E2) is a type CF (defibrillation proof), 110/120/230v, 50/60Hz, 10A, with respect to electric shock, fire, mechanical hazards in accordance with UL 2601-1:1997, EN/IEC 60601-1, CAN/CSA C22.2 No. 601.1-M90 CSA file #220418. MDD CLASS IIb - CE Mark Certificate of Conformity No: 0088/ LRQ 0959680 PMA approved device PO20045 MDL 32473 (Canada) REF# MDT.2 CryoConsole After power up, normal operating parameters are: NA: 110V±10% 10 amps CE: 230V±10% 5 amps At power up, the equipment could draw a transient current of 10% greater than the nominal values Operational Parameters: 103E2 / 103A2 / 104E2 / 104A2 15°C to 30°C (60°F to 86°F) Maximum of 75% humidity Altitude from sea level to 2400 meters (8000 feet) Dimensions: Gen III Gen IV Metric Imperial Metric H: 127 cm H: 50” H: 157 cm D: 82 cm D: 32” D: 58.5 cm W: 54 cm W: 21” W: 54.6 cm Weight: Gen III Gen IV Metric Imperial Metric Imperial H: 62” D: 23” W: 21.5” Imperial 145 kg 320 lbs 131 kg 289 lbs (Without N2O tank) (Without N2O tank) (Without N2O tank) (Without N2O tank) Table 1 - CryoConsole Specifications Medtronic CryoCath LP Page 13 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Accessories The ECG cable and auto connection box provide an interface between the Medtronic CryoCath LP and a standard EP system (via 2mm touch proof connectors). The scavenging hose is used to evacuate N2O from the CryoConsole to the hospital gasscavenging outlet. The optional foot switch is a remote control to initiate and stop injection of refrigerant. Copies of all the Regulatory Approvals are available upon request. Medtronic CryoCath LP Page 14 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV REF# Accessories 2037A Auto connection box 2035W 103FS(GenIII) 104FS(GenIV) ECG cable Footswitch (optional) Rev-00 Metric L: 14 cm W: 7 cm H: 5 cm Imperial L: 5.5” W: 2.5” H: 1.8” 0.5 kg (approx.) 1.1 lbs. (approx.) L: 81 cm L: 32” L: 17 cm W: 15 cm H: 2.5 cm L: 6.5” W: 6.0” H: 1.0” Cord Length: 6.7m Cord Length: 12’ L (1035C): 20’ L (1035D): 25’ L (1035E): 30’ L (1035F): 40’ ID: 0.3” OD: 0.55” 1035C (1035CW) 1035D (1035DW) 1035E (1035EW) 1035F (1035FW) Yellow Scavenging hose (White Scavenging hose) L (1035C): 6.0 m L (1035D): 7.5 m L (1035E): 9.0m L (1035F): 10.5m ID: 0.75 cm OD: 1.375 cm 1038N Power cord North America L: 2.5 m L: 8’ 2” 1038E Power cord EU L: 2.5 m L: 8’ 2” 1038U Power cord UK L: 2.5 m L: 8’ 2” 1038Y Power cord Italy L: 2.5 m L: 8’ 2” Table 2 - Accessories Specifications Refrigerant Tank The refrigerant is liquid nitrous oxide (N2O). ¾ ¾ European N2O refrigerant tank – REF # 103NE North American N2O refrigerant tank – REF # 103N2 Medtronic CryoCath LP Page 15 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 The CryoConsole operates with a refrigerant tank that contains high-pressure liquid nitrous oxide (N2O). Store and handle the nitrous oxide refrigerant tanks according to the hospital gas storage/management policy. Exposure to nitrous oxide should be minimized to prevent shortterm behavioural and long-term reproductive health effects. Do not use Medtronic CryoCath LP refrigerant tanks on any other equipment. Use only the N2O refrigerant tank obtained by Medtronic CryoCath LP. The Threshold Limit Value (TLV) for nitrous oxide is 25ppm as recommended by NIOSH, National Institute for Occupational Safety and Health. N2O Specifications Description: Cryogenic liquid nitrous oxide. Molecular Formula: N2O (UN 1070) 99.5% liquid phase with humidity level less than 50ppm. Gas Scavenging Requirements The CryoConsole produces Nitrous Oxide (N2O) waste gas that should be removed from the system. There are three options for scavenging. Selecting the appropriate scavenging method is the hospital’s responsibility. Use only the scavenging hose obtained by Medtronic CryoCath LP. Exhaust N2O to Vacuum Scavenging System (Active Scavenging) This is the Medtronic CryoCath LP recommended method for removing waste gas. The scavenging hose should be attached to a hospital evacuation or vacuum system. The hospital system should meet the following requirements: Gas to be scavenged .............................................................. Nitrous Oxide (N2O) Scavenging refrigerant flow .................................................... 5000 sccm N2O Scavenging vacuum level ....................................................... ≥ 15 in Mercury (Hg), or ≥ 380mm Hg, or ≤ 7.4 psia, or ≤ 51 KPa Scavenging hose length provided .......................................... approximately 6.1m/ 20’ up to 18.3m/ 60 ‘ Scavenging connector ........................................................... DISS (Some adaptors for common connection systems are available) Exhaust N2O into a Non-recirculating General Ventilation System (Passive Scavenging) If a non-recirculating general ventilation system is present in the hospital, the discharge end of the scavenging hose can be placed up in an exhaust vent and the N2O will be vented out of the building. Care should be taken to ensure that the tubing is placed far enough into the duct system to prevent the N2O from re-entering the room. Exhaust N2O Directly to Outdoors (Passive Scavenging) Since the gas is under positive pressure, the discharge end of the scavenging hose can be exhausted directly outdoors. This can be accomplished by placing the end of the hose out a window or attached to a pipe placed through a wall or ceiling to the outdoors. Care must be taken to ensure that the discharge location is NOT an area where people are normally present or in close proximity to an outside air intake or window where the N2O could re-enter the building. The standard scavenging hose length provided by Medtronic CryoCath LP is 10.5m/40’ long. If a longer hose is required, please contact Medtronic CryoCath LP Customer Service. Medtronic CryoCath LP Page 16 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Section II Functional System Description II.1 Purpose The purpose of this section is to give a basic knowledge of the functional system of the MDT.2 CryoAblation System, therefore ensuring that appropriate actions are taken. II.2 Scope This section briefly describes the functional system of the CryoConsole, introducing the software interface, the mechanical system, and the electrical system. II.3 Definitions and Acronyms The following definitions and acronyms are used in this section MDT EP LCD CDM PID CPU ISA BUS AC DC IDE II.4 Medtronic CryoCath LP Electrophysiology Liquid Crystal Display CryoAblation Data Manager Proportional-Integral-Derivative Central Processing Unit Industry Standard Architecture BUS Alternating current Direct current Integrated Drive Electronics System Overview The CryoConsole is a combination of mechanical assemblies controlled by electronic circuitry. Control interfaces are available on the front and back panels of the unit allowing the system to be integrated in EP Lab set-up. The catheter electrical connection is fed to the system through an isolation interface located on the patient board. This electrical interface provides a protection against electrical hazards and protects the CryoConsole unit from defibrillator discharges. The user interfaces with the CryoConsole through a LCD touch screen, two indicators and two controls located on the front panel of the CryoConsole. The LCD touch screen shows a series of menus for operation of the CryoConsole and display of relevant information. The user starts the application of a procedure by pushing the hardware injection “ON” button for 2 seconds and stops the procedure at any time by pressing the hardware injection “OFF” button. Medtronic CryoCath LP Page 17 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 When the optional foot switch is in use, the user starts the application of a procedure by depressing the footswitch for 2-3 seconds and stops the procedure at any time by depressing the footswitch for 0.5 seconds. The watchdog system is the focal point of any data sent and received either to or from the software user interface. The watchdog system is the brain of the system control, which continually monitors the mechanical assembly state, user requests and process failure and/or warning detection. It also acts as an interface between the software/user and the mechanical plumbing, which ensures patient and user safety should the software fail. II.5 Software Interface II.5.1. Software Interface Overview The CryoConsole includes the software called CryoAblation Data Manager (CDM) and the software called Boot Strap Module (v3.02 and above). The main role of the software is to acquire and display system data from the hardware, set procedure parameters, save procedure data on the system hard disk and offer some automatic operational features. At system start-up, the software initiates a system self test which ensures proper functionality of the watchdog system and other system and safety features. Should the self-test fail, the CDM disables the system in order to prevent the user from proceeding. II.5.2. Boot Strap Module The role of the Boot Strap Module is to detect and authenticate software upgrade MDT diskettes and then launch the upgrade. This application facilitates software upgrades without the need for a field service representative intervention. This software is installed on CryoConsole depending on the approved country of installation. II.5.3. CryoAblation Data Manager (CDM) From the main interface panel, the software provides four major functional areas for the user to access: the CryoTherapy Panel, the Review Patient Records Panel, the Service System Panel and the Shut Down System button. Selecting the Begin CryoTherapy button brings up the Patient Data Panel where the patient information is entered and saved to the hard disk. Through this panel, the user initiates a CryoTherapy procedure. When a CryoTherapy procedure is requested, the software ensures that a valid catheter is connected to the CryoConsole. When it is connected, the CDM downloads a catheter identification code stored on a smart chip located inside the catheter handle. The catheter code is used to identify a catheter specific file located on the hard disk of the CryoConsole. This file allows the software to program the user interface on the procedure panel and set their default values, and contains parameters required for the proper operation of the identified catheter. During a procedure, the software continuously displays relevant information on the screen. The temperature profiles are saved to the hard disk for future recall. Data recording starts at the Medtronic CryoCath LP Page 18 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 beginning of an injection and stops when the temperature reaches +10oC after stopping the injection. This feature allows the recording of the temperature thawing profile. If a warning or failure occur, the user is immediately notified. If a system failure is detected during the procedure, the software stops saving the temperature profiles and opens a separate file where it logs the current state of the system. This system state file contains the most recent 99 system failures. Depending on the type of catheter connected, the CryoConsole controls the catheter in two different ways: Temperature Control (only during CryoMapping): During a procedure, pressurized refrigerant is delivered in order to reach and maintain a predetermined ablation temperature. This continuous temperature control assures a compensation for heat load changes due to the blood movement during cardiac cycles or due to changes in the contact between the catheter tip and the tissue. Temperature control is performed using two control loops: a hardware pressure control loop and a software temperature control loop. The CDM controls the tip temperature by reading the current tip temperature and determines the required delivery pressure set point using a special algorithm based on digital proportional-integral-derivative (PID) compensator. The pressure set point is sent to the hardware pressure controller, which assures a continuous control of a proportional valve based on a pressure transducer reading. Flow Control (for all catheters during CryoAblation): During a CryoTherapy procedure, pressurized refrigerant is delivered to reach and maintain a specific pre-determined flow value. This flow value ensures optimal tip temperatures required for ablation. As in the temperature control case, two control loops are used: A hardware pressure control loop and the software flow control loop. The CDM controls the refrigerant flow by reading the flow meter and determining the required delivery pressure set point to maintain that flow using a special algorithm based on digital proportional-integral-derivative (PID) compensator. Should the CDM fail, a redundant hardware system is implemented as part of the Watchdog system to monitor the pressure set point sent by the CDM. Furthermore, the Watchdog system monitors the state of the CPU and the software in order to detect a freeze or abnormal loops and takes action by placing the mechanical plumbing into a safe mode. Another important feature of the Watchdog system is to block the analog set point controlled by the CDM whenever an injection OFF is requested or a failure is generated. If the Review Patient Records option is selected, a file manager panel is displayed. This allows the user to select from a list of previously run procedures. The selected file may then be opened for viewing, copied to a floppy disk, or deleted from the hard disk. The third functional area, which can be selected from the main panel, is the Service System Panel. This is to aid certified service personnel in performing routine system maintenance and to help technical personnel in troubleshooting the system. To prohibit unauthorized personnel from accessing restricted system information, several troubleshooting panels are password protected. Medtronic CryoCath LP Page 19 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 The last functional area is the Shut Down System button. Before turning off the power of the CryoConsole, make sure to wait until the message “It is now safe to turn off your computer” is displayed on the screen. II.5.4. Catheter Files Each catheter used with the CryoConsole requires different settings of operations. Therefore, the data for each catheter is stored in a catheter file. When a catheter is approved for use with the CryoConsole, the associated catheter file is installed in the appropriate directory. II.5.5. Hardware File The CryoConsole uses a binary file called Hardware file which contain the necessary parameters for interpreting analog signals including the appropriate tank formula for the calculation of remaining N2O. The hardware file is dependant of the CDM software version and the CDM software is configured to load the appropriate file based on the console configuration (North American, European, or other). II.6 Mechanical System II.6.1. Mechanical System Overview The mechanical assembly delivers a controlled pressurized refrigerant to the catheter, recovers the refrigerant vapour from the catheter and evacuates the residual refrigerant from the catheter at the end of the injection. The mechanical system can be split into the following three subsystems: ¾ Refrigerant delivery plumbing ¾ Refrigerant recovery system ¾ Refrigerant evacuation system The delivery system uses liquid N2O as refrigerant, which is sub-cooled to an appropriate temperature prior to being injected inside the catheter through the injection umbilical tube. The system recovers the N2O vapour from the catheter and sends it to the hospital scavenging system through a scavenging hose. Electromechanical valves, pressure transducers and flow meter are used in order to control and monitor the status of different areas of the mechanical assembly. II.6.2. Pressure Transducers A pressure transducer is a device that converts input pressure into low-level electrical/electronic signals (volts). In the CryoConsole, we use three pressure transducers. The first transducer is used to measure the tank pressure. The second transducer is used to measure the pressure delivered by the Medtronic CryoCath LP Page 20 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 proportional valve. That pressure is fed to the PID system that will be used as a feedback to compensate and adjust the proportional valve according to either the temperature or the flow (depending on the catheter used). The third transducer is used to measure the pressure out of the catheter, thus verifying that N2O gas is flowing out of the catheter and that no blockage occurs inside the catheter. II.6.3. Regulators A pressure regulator is a device that adjusts pressure at a desired value. Since a pressure of 760 psig is enough to ablate at -75oC, we use two regulators, first to bring down the pressure to 840 psig and second to bring it down to 810 psig. II.6.4. Solenoid Valves Solenoid valves are electro-mechanical devices that use a solenoid to control valve activation. In the CryoConsole, there are five solenoid valves performing specific duties. ¾ The first valve is the injection solenoid valve turning on and off the injection circuit. ¾ The second valve is the switch II.7 Electrical System II.7.1. Electrical System Overview The LCD touch screen and the front panel implement the user interface with the CryoConsole. The LCD touch screen is a graphical user interface, which displays CryoConsole status, data and accepts user data and control inputs via screen touches. The front panel consists of controls and displays that show CryoConsole status. The side IDE panel consists of a mouse, keyboard and analog connectors. The mouse and keyboard connectors are connected to the CPU board through extension cables. Those connectors should only be used by Field Service representatives. Also found on this panel is an hour meter that indicates the number of hours the CryoConsole has been powered on. This information is useful for maintenance purposes. The 1.44 Mb floppy disk drive is mounted in the procedure panel and is used to download patient data on disks and to upload upgrades to the CDM software and catheter identification files. Medtronic CryoCath LP Page 21 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 II.7.2. Power Distribution The power entry module with voltage selector, thermistors, slow blow fuses, line filter and power switch are rated at 250V operating voltage. The power supply, isolation transformer and line filter are rated at 50/60 Hz. The voltage selector on the primary side of the isolation transformer is used to enable the 120v/240v conversion. II.7.2.1. Thermistors The thermistors are only installed on the power inlet cables of CryoConsole powered by 220V. The thermistors purpose is to protect against any existing inrush current when the system is first switched on. Section III Operation and Clinical Sequence Please refer to the Operator’s Manual, its appendices, as well as the appropriate Instructions For Use for the respective catheter. Section IV Troubleshooting and Maintenance IV.1 Purpose This section proposes troubleshooting steps to follow when the CryoConsole is not operating correctly. IV.2 Scope This section is separated in two main sections. These sections are the maintenance and the software interface. Medtronic CryoCath LP Page 22 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 IV.3 Definitions and Acronyms The following definitions and acronyms are used in this section CDM CPU DAQ FPGA HDD NIBBLE CryoAblation Data Manager Central Processing Unit Data acquisition board Field Programmable Gate Array Hard Disk Drive Four binary digits or half of a byte. A nibble can be conveniently represented by one hexadecimal digit. IV.4 Maintenance In addition to act as an interface for the user, the CryoAblation Data Manager software installed on the CryoConsole is an interface for the Field Service Representative during a maintenance job. The representative can access the maintenance section by touching the Service System button located on the main panel, shown in Figure 7. Some panels require the use of a keyboard attached to the maintenance panel above the N2O cylinder. The connector type of the keyboard must be a type AT. Figure 7 - Main panel IV.4.1. System Setup panel This screen is used to monitor the tank level read by the load cell as well as the tank pressure read by the pressure transducer. The user can return to the Main Panel by pressing the Return to Main Panel button or access the System Maintenance panel by pressing the Advanced Troubleshooting button. Figure 8 shows the System Setup panel. Medtronic CryoCath LP Page 23 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Figure 8 - System Setup panel IV.4.2. Password panel When pressing the Advanced Troubleshooting button, the system notice #12303 - The CryoAblation console maintenance should be done by a qualified person appears. The CryoConsole requires a password to proceed. The low level access password is “user”. Only trained service engineers will have access to all required passwords. The user enters the password and presses Ok. Figure 9 - Password panel IV.4.3. System Maintenance panel In this panel, the user can choose between six different sections: ¾ ¾ ¾ ¾ ¾ Password Setup Date & Time Setup Touch Screen Setup Site Setup Back Up Patient Files Figure 10 shows the System Maintenance panel. Medtronic CryoCath LP Page 24 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Figure 10 - System Maintenance panel IV.4.4. Password Setup panel The purpose of this panel is to change the default password used to access the System Maintenance panel. Figure 11 shows the Password Setup panel. Do NOT change the password. Figure 11 - Password Setup panel IV.4.5. Date & Time Setup panel The Date & Time Setup panel allows the user to change the date and time of the CryoConsole. Figure 12 shows the Date & Time Setup panel. Figure 12 - Date & Time Setup panel IV.4.6. Touch Screen Setup panel The purpose of this panel is to calibrate the touch screen when the precision is out of range. When the user pushes on the Touch Screen Setup button, the system gives instructions on how to calibrate the touch screen. Medtronic CryoCath LP Page 25 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV IV.4.7. Rev-00 Site Setup panel The purpose of this panel is to: ¾ Choose the language of the CryoAblation Data Manager ¾ Enter the site name ¾ Update the physician, outcome and diagnosis list of the CryoAblation Data Manager To set the language, select a language from the Language drop-down list. To enter the site name, write the site name in the New Site text box and press the Update Site button. This name will appear on the home screen after power-up. To update one of the three lists, press the desired button (Physician, Outcome, Diagnosis), enter the physician name, outcome or diagnosis in the text box, and press the Add button. To delete an entry, select the entry in the list and press Delete. Refer to Figure 13. Figure 13 - Site Setup panel IV.4.8. Back Up Patient File The purpose of this panel is to make a back up of all the patient data currently stored on the hard drive. When downloaded from this panel, the files are saved in binary format. These files should be downloaded and sent to MDT on a regular basis. Figure 14 - Backup Patient Files panel Medtronic CryoCath LP Page 26 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 These files help us understand any problem encountered during an injection, see behaviors that may have not appeared during an injection, and helps us identify any possible behavioural patterns. To download these files: 1) Navigate to the System Maintenance panel 2) Select Backup Patient Files 3) Insert a floppy disk in the floppy drive 4) Press Backup. The console will start downloading the patient files that have a check mark to the floppy disk. As the files get saved on the floppy, the checkmark is removed. The console may request you to change the disk if it gets full. The downloading process is finished when all the checkmarks disappear and the green light on the floppy drive turns off. Green Light Figure 15 - Floppy Drive IV.5 Software Interface IV.5.1. General information The CryoConsole includes software called CryoAblation Data Manager (CDM). The main role of the CDM is to acquire and display system data from the hardware, set procedure parameters, save procedure data on the system hard disk and offer some automatic operational features. At system start-up, the software initiates a system self test which ensures proper functionality of the watchdog system and other system and safety features. Should the self-test fail, the CDM disables the system in order to prevent the user from proceeding and displays a coded error message on the screen. IV.5.2. Power up test Medtronic CryoCath LP Page 27 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 At start-up, CDM performs a series of tests to verify the CryoConsole hardware and proper operation. If a test fails, two error codes are generated and displayed on the CryoConsole screen. The first series of power up tests is related to the software integrity, including the testing of the system and configuration files. These files contain important information that could affect the functionality of the system. The second series of tests is related to the interface between the software and the hardware. These tests verify the proper operation of the data acquisition, digital and analog input/output lines. The data acquisition test is crucial to the performance of other tests; if this test fails, the remaining tests are not performed. The third series of tests is related to the functionality of the FPGA and the functionality of major safety features. The FPGA component is part of the Watchdog board. IV.5.3. CDM Codes and Messages The CDM software uses codes and messages to inform the user the status of the CryoConsole. Always note the message number before calling Technical Support in order to resolve the issue faster. IV.5.3.1.1. System notice hierarchy A message to the user is called a system notice. That system notice can be of four different types, depending on the message content. Starting from the least important and going to the most important messages are the notice, caution, the warning and the danger. They respectively are blue, yellow, orange and red. Depending on the source of the notice, the software will assign a different importance to this notice. To prevent the user from missing a system notice, the first screen appearing to the user is the Message List notice showing all of the system notices. Figure 16 shows the Message List screen Figure 16 - Message List Medtronic CryoCath LP Page 28 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 The system notices includes a different presentation of the information. It presents the notice with three distinctive fields: Problem, Consequence and Action. Figure 17 and Figure 18 show the four notice types. A blue notice is for information only and no action is required in order to get into the Therapy panel and start and injection. A yellow notice requires an action in order to go into the Therapy panel. An orange notice is a failure that will stop an injection or prevent the user from starting one. This notice can be reset and no immediate action needs to be taken. A red notice is a critical console failure that will cause the termination of the procedure. The console needs to be shutdown and Medtronic CryoCath LP Technical Support must be informed. Figure 17 - Danger and Warning System Notice Figure 18 - Caution and Notice System Notice Medtronic CryoCath LP Page 29 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 The notices are part of a numbered series. Table 3 shows the numbered series in relation with the source of the notice. Series 10000 11000 12000-13000-14000 15000 16000 17000 18000 19000 20000 21000 22000-23000-24000-25000 50000 60000 Source of the notice Language confirmation panel System Start-up Failure System Warning File Reading Error Disk Space File Writing Error System Alert System Error Directory Error System Self Test Failure Message Panel Hardware Failure Hardware Warning Table 3 - Notice series IV.5.3.1.2. Hardware System Notice The hardware system notices are the most common errors that are seen in the field while troubleshooting a problem. The watchdog board is the hardware responsible for triggering a hardware failure or a hardware warning. The associated system notices are the 50000 and 60000 series notices. Whenever the console fails the Power up test, it will give you yellow system notice, #11200, saying “There is a problem with the system Error1: X-X-X Erro2: Y-Y-Y). Be sure to note these two error numbers before calling Technical Support. Table 5 in Appendix A– System Notices shows all of the system notices, the problem, the consequence and recommended action. Table 6 in Appendix B– Additional Warning Messages shows warning messages not included in the system notices. Medtronic CryoCath LP Page 30 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Appendix A– System Notices . Notice Problem Consequence Action Description 10300 The language you selected is not currently available. The language was not changed. Select a different language. Language selected is not currently available 11200 There is a problem with the system (Error1: X, Error2: Y) You cannot continue the procedure. Shut down the system and restart it. If the problem persists, call Medtronic CryoCath LP customer service. Shutdown system and call service technician. Error1 = X Error2 = Y 12200 There is a problem with the system. You must take action in order to continue the procedure. 12201 There is a problem with the system. You must take action in order to continue the procedure. There is a problem with the system file. There is a problem with the system file. You cannot continue the procedure. You cannot continue the procedure. 12204 There is a problem with the catheter file. You cannot use the catheter. 12205 There is a problem with the catheter You must take action in order to continue the procedure. 12206 The catheter may be outside the patient, causing the catheter tip temperature to be too low. You must take action in order to continue the procedure. 12300 There is a problem with the language library: (%s) The language setting has changed back to English. 12301 There is a problem with the system file. The system functionality is limited. 12202 12203 Disconnect the electrical umbilical cable from the catheter and then reconnect it. If the problem persists, call Medtronic CryoCath LP customer service. Disconnect the electrical umbilical cable from the catheter and then reconnect it. If the problem persists, call Medtronic CryoCath LP customer service. Call Medtronic CryoCath LP customer service. Call Medtronic CryoCath LP customer service. Disconnect the catheter and reconnect it. If the problem persists, call Medtronic CryoCath LP customer service. Disconnect the catheter electrical umbilical cable and reconnect it. If problem persists, replace the catheter. Check the catheter placement. Try selecting the language again. If the problem persists, call Medtronic CryoCath LP customer service. Install or create a new system file. 12302 There are no messages Click Ok to continue 12303 The CryoAblation console maintenance should be done by a qualified person. Click Ok to continue The system could not save the password. The password was not changed. Try to change the password again. If the problem persists, call Medtronic CryoCath LP customer service. There is a problem with the message file. The system could not save You cannot view the messages. The language was not 12304 12305 12306 Medtronic CryoCath LP You are not able to continue the procedure, Flow detection initialization error Flow Curve Programming Error (Faulty Tank/Temperature programming) Incompatible System File Incompatible Failure File Invalid Catheter Data File Catheter Compatibility / Integrity Unacceptable catheter use / Catheter Compatibility Can't load language library, Library Name System file error The Failure Database is empty The CryoAblation console maintenance should be done by a qualified person. Field Password Saving Error Click Ok to continue. Failure file not found Select a different Error Saving Language Page 31 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Rev-00 Problem Consequence Action Description your language selection. changed. Code 12307 There is a problem with the configuration file. The system could not save the temperature calibration parameters. 12308 The system could not open the patient file. You cannot view the patient file. 12309 The system could not open the patient file. You cannot view the patient file. 12310 The system could not open the patient file. You cannot view the patient file. 12311 The system could not open the patient file. You cannot view the patient file. 12312 The patient file is empty. language, or install or create a new configuration file. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Call Medtronic CryoCath LP customer service. Install or create a new messages file. Install or create a new configuration file. 12313 12314 12315 There is a problem with the message file. There is a problem opening the configuration file. There is a problem with the catheter file. You cannot view the patient file. The system cannot update the file. The system cannot update the file. The system cannot load the catheter parameters. You cannot continue without entering a catheter name. 12316 The catheter name field is empty. 12317 The temperature has not yet reached 20-30 degrees. You cannot calibrate the temperature yet. 12318 There is an error copying the patient file. You cannot copy the patient file to diskette. 12319 There is an error copying flow curve data. You cannot copy the flow curve data to diskette. 12320 The patient name field is empty. You cannot continue without entering a patient name. 12321 There is no diskette in the drive. 12322 Catheter file copying error: %s Cannot copy catheter file to the floppy 12323 There is a problem with the catheter file %s. You cannot use the catheter. 13300 The system cannot load the touch screen calibration file. You cannot calibrate the touch screen. Medtronic CryoCath LP Delete this file. Enter the catheter name. Calibration Saving Error Patient file not found Patient file Empty (file check code is incorrect) Patient Record Reading Error (Pt file cannot be opened) Invalid Patient File (file check code is incorrect) Patient Record Empty Failure file error (setcdmver) Configuration file error (setcdmver) Invalid Catheter Data File, Please delete it Please enter Catheter Name Try again. If the problem persists, call Medtronic CryoCath LP customer service. If the problem persists, contact Medtronic CryoCath LP customer service. Temperature calibration can not be performed, Please check patient board temperature File copying error, General input output error with the following (patient Filename) File copying error, General input output error with the following (flow curve data Filename) Enter the patient's name in order to continue. Please enter patient name in order to continue Insert a diskette and press Ok. Try again, and if problem persists call Medtronic CryoCath LP customer service Disconnect the catheter and reconnect it. If the problem persists, call Medtronic CryoCath LP customer service. No disk in the floppy drive, Please insert a disk File copying error. General input output error with the following (Filename) Wait until the board reaches the appropriate temperature. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Catheter file error. Following file is not a valid catheter file (filename) Could not load Calibration Library Page 32 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Problem Consequence You cannot continue without entering a console ID. You may not be able to complete another procedure. 14300 The console ID is empty. 14301 The hard drive is almost full. 15300 There is a problem finding the file. You cannot copy the file to diskette. 15301 There is a problem finding the file. You cannot copy the file to diskette. 15302 There is a problem copying the file. You cannot copy the file to diskette. 15303 There is a problem copying the file. You cannot copy the file to diskette. 15304 There is a problem finding the file. You cannot copy the file to diskette. 15305 There is a problem copying the file. You cannot copy the file to diskette. 16300 16301 There is no diskette in the drive. There is not enough space on the diskette. Action Description Enter the console ID now. Console ID is not set yet, Please, do it now Free some space on the hard drive before the next procedure. The hard disk is almost full, Free some space before the next procedure Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Insert a diskette and press Ok. Insert another diskette and press Ok. 17300 There is a problem copying the file. You cannot copy the file to diskette. 17301 There is a problem copying the file. You cannot copy the file to diskette. 17302 There is a problem copying the file. You cannot copy the file to diskette. 17303 There is a problem copying the file. You cannot copy the file to diskette. 17304 There is a problem copying the file. You cannot copy the file to diskette. 17305 There is a problem copying the file. You cannot copy the file to diskette. 17306 There is not enough space on the diskette. 18300 There is a problem with the Medtronic CryoCath LP Rev-00 You cannot continue the Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. To continue, insert another diskette and press Ok. Otherwise, press Cancel. Shut down the system and File not found or directory in path not found (copy fail file, pt db, conv ptdb) File not found, please remove disk (copy cath db) General I/O error occurred (copy fail file, mkdir, copy conv ptdb) Insufficient memory to complete operation (copy fail file, mkdir, copy conv ptdb) Invalid path (copy fail file) Access denied (copy fail file & ptdb) No disk in the floppy drive, Please insert a disk Not enough disk space, Change disk One of the path components not found (Copy ptdb & MkDir) General I/O error occurred Insufficient memory to complete operation Invalid path Access denied Specified path is a directory, not a file. Disk is full (w/ chg disk in Archive & ConvDb) You are not able to Page 33 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Rev-00 Problem Consequence Action Description system. procedure. restart it. If the problem persists, call Medtronic CryoCath LP customer service. continue the procedure, Reboot System 19300 There is a problem copying the file. You cannot copy the file to diskette. 19301 There is a problem finding the file. You cannot copy the file to diskette. 19302 There is a problem finding the file. You cannot copy the file to diskette. 19303 There is a problem copying the file. You cannot copy the file to diskette. 19304 There is a problem copying the file. You cannot copy the file to diskette. 19305 There is a problem copying the file. You cannot copy the file to diskette. 19306 There is a problem copying the file. You cannot copy the file to diskette. 20300 There is a problem copying the file. You cannot copy the file to diskette. 20301 There is a problem copying the file. You cannot copy the file to diskette. 20302 There is a problem copying the file. You cannot copy the file to diskette. 20303 There is a problem finding the file. You cannot copy the file to diskette. 20304 There is a problem copying the file. You cannot copy the file to diskette. 20305 There is a problem copying the file. You cannot copy the file to diskette. 21200 There is a problem with the system. You cannot continue the procedure. 22400 System Shut Down Please wait while your Medtronic CryoCath LP Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Shut down the system and restart it. If the problem persists, call Medtronic CryoCath LP customer service. Access to floppy disk is denied, check disk (copy fail file and conv db) File not found or directory in path not found General I/O error occurred Insufficient memory to complete operation Invalid path (for either of the file names). Specified source path is a directory, not a file (copy fail file) Disk is full One of the path components not found General I/O error occurred Insufficient memory to complete operation Invalid path Access denied Disk is full You are not able to continue the procedure, Call a service technician Please wait while your Page 34 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Problem Consequence 22402 Target Temperature Not Attained System Shut Down system shuts down Minimum temperature is warmer than %d °C Please close gas tank 22403 System Startup 22401 Rev-00 Action Make sure the refrigerant tank has been opened. Catheter integrity test, injection will be enabled shortly. System self test - Please wait ... Refilling Reservoir, injection will be enabled shortly. 22500 22501 22502 Injection will be enabled in X seconds System Startup - Self test delay 23300 Are you sure you want to exit? If you exit, the system will close this patient file. As a result, you would not be able to continue this patient's procedure and would have to create a new patient file. 23301 The patient file will not fit on one diskette. You will need multiple diskettes. To continue, press Ok. Otherwise, press Cancel. 23302 Are you sure you want to end the procedure? If you end procedure, the system returns to the main screen and closes the patient file. To end the procedure, press yes. Otherwise, press No. 23400 File Delete Are you sure you want to delete the file? 23401 System Shut Down 23402 Patient Database 23403 Temperature Calibration 23404 Catheter File Database 23405 Analog Data 24300 There is not enough space on the diskette. Error in File Selection 25401 Copy Failed Click Ok to retry, otherwise click Cancel and copy 'c:\\cryosys\\system\\Analo Are you sure you want to exit from this panel ? Not enough disk space, You will need multiple disks, do you wish to continue? (Conv db copy to disk) End Procedure, By ending the procedure you will return to the main screen and close the patient file. Are you sure you want to end the procedure? (Response to user selecting abort in failure panel) File Delete, Are you sure you want to delete the file? (pt db) System Shutdown, Are you sure you want to quit? Database Conversion, Database already exists, overwrite? System Saving, Are you sure you want to save these data? Warning, Do you want to overwrite the existing file (Overwriting catheter file) Overwrite analog data file To continue, insert another diskette and press Ok. Otherwise, press Cancel. 25400 Medtronic CryoCath LP To exit, press Yes. Otherwise, press No. Are you sure you want to quit? This patient file already exists. Do you want to overwrite it? Are you sure you want to save the temperature calibration? This catheter file already exists. Do you want to overwrite it? Overwrite the existing analog data file? Click Ok to retry, otherwise click Cancel and copy 'c:\\cryosys\\system\\Analo gData.txt' manually Description system shuts down Minimum temperature is warmer than X °C Please close gas tank Make sure the refrigerant tank is opened 2 turns counter clockwise Change disk then press Ok//Conv or archive, ok to continue else cancel Error in File Selection, Click OK to Retry or Cancel and copy c:\cryosys\system\AnalogD ata.txt manually Copy Failed, Click OK to Retry or Cancel and copy c:\cryosys\system\AnalogD Page 35 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Problem Consequence Rev-00 Action gData.txt' manually 50000 50001 The electrical umbilical cable is not connected properly. The vacuum is disabled due to a problem with the coaxial umbilical cable. Injection has stopped. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. 50002 The coaxial umbilical cable is not connected properly. 50003 The pressure is low in the system. 50005 Catheter leak is detected. 50006 Blood is detected inside the catheter handle 50007 There is fluid in the console. You cannot continue the procedure. 50008 There is a problem with the system. Injection has stopped. 50009 There is a problem with the system. Injection has stopped. 50010 There is a problem with the system and the temperature could drop below acceptable limits. You must take action in order to continue the procedure. 50011 The path of the refrigerant is obstructed. Injection has stopped. You must take action in order to continue the procedure. 50012 The path of the refrigerant is obstructed. Injection has stopped. You must take action in order to continue the procedure. 50013 The refrigerant level is too low to continue. You cannot continue the procedure. 50014 There is a problem with the console. You must take action in order to continue the procedure. 50016 The CryoMapping temperature exceeded the PRESET range of Injection has stopped. You must take action in order to continue the procedure. Medtronic CryoCath LP Description ata.txt manually Disconnect both ends of the electrical umbilical cable and reconnect them. Disconnect both ends of the coaxial umbilical cable and reconnect them. Disconnect both ends of the coaxial umbilical cable and reconnect them. 817 Vacuum connection failure Ensure refrigerant tank is open. 819.1Low gas pressure X psi, ensure refrigerant tank is open Replace the catheter 821 Catheter leak is detected Replace the catheter 822 Blood is detected inside the catheter handle The system is inoperable. Please shut it down and call Medtronic CryoCath LP customer service. Try again. If the problem persists, shut down the system and restart it. If there is still a problem, call Medtronic CryoCath LP customer service. Try again. If the problem persists, shut down the system and restart it. If there is still a problem, call Medtronic CryoCath LP customer service. Disconnect the catheter electrical umbilical cable and reconnect it. If problem persists, call Medtronic CryoCath LP customer service. Disconnect the coaxial umbilical cable from the console, make sure that the port it was attached to is clear of obstruction, and reconnect the coaxial umbilical cable to the console. If problem persists, replace the coaxial umbilical cable. If there is still a problem, replace the catheter. Replace the refrigerant tank. Try again. If the problem persists, shut down the system and restart it. If there is still a problem, call Medtronic CryoCath LP customer service. Position the catheter tip so that it makes better contact with the tissue. 816 Catheter electrical connection failure Baseline Flow 823 Fluid is detected inside console vacuum line 824 Software Failure 825 Failure has been initiated by the CPU. 826Temperature Threshold programming 819.2 Ensure the inside of console coaxial port is clear 819.3 Ensure integrity of refrigerant path in catheter/umbilical 829 Refrigerant level failure 830 Sub cooler compressor failure 832 Improve catheter tiptissue contact Page 36 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice Problem Consequence Rev-00 Action Description temperature. 50017 The console is running hot. Injection has stopped. You cannot continue the procedure. 50018 The refrigerant is not cold enough. Injection has stopped. You cannot continue the procedure. 50021 Problem with scavenging Injection stopped. You must take action in order to continue the procedure. 50022 There is a problem with the console You must take action in order to continue the procedure 50024 There is a problem with the refrigerant vent. You must take action in order to continue the procedure. 50025 There is a problem with the injection process. Injection will not start. You must take action in order to continue the procedure. 50026 There is a problem with the injection process. Injection has stopped. You must take action in order to continue the procedure. 50027 There is a problem with the injection process. Injection has stopped. You must take action in order to continue the procedure. 50028 There is a problem with the injection process. Injection has stopped. You must take action in order to continue the procedure. 50029 There is too much refrigerant flowing. 50030 There is too much refrigerant flowing. 50031 There is a problem with the injection process. 60000 The coaxial umbilical cable is not connected properly. 60001 The electrical umbilical cable is not connected Medtronic CryoCath LP Injection has stopped. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. Injection has stopped. You must take action in order to continue the procedure. You must take action in order to continue the procedure. You must take action in order to continue the Check whether the fans are working. Shut down the system and wait 10 minutes before restarting it. If the problem persists, call Medtronic CryoCath LP customer service. Wait 10 minutes before continuing the procedure. If the problem persists, call Medtronic CryoCath LP customer service. Ensure that the hospital scavenging system is turned on and that the scavenging hose is securely attached. Try again. If the problem persists, shut down the system and restart it. If there is still a problem, call Medtronic CryoCath LP customer service. Disconnect the coaxial umbilical cable from the console and make sure that the port it was attached to is clear of obstruction. Make sure there is no obstruction in the catheter or coaxial umbilical cable, and reconnect it. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Replace the coaxial umbilical cable. Position the catheter tip so that it makes better contact with the tissue. Try again. If the problem persists, call Medtronic CryoCath LP customer service. Disconnect both ends of the coaxial umbilical cable and reconnect them. Disconnect both ends of the electrical umbilical 833 Console temperature is high 834 Refrigerant temperature is high 837 Scavenging system failure 838 Vacuum pump failure 840 Vent system failure 841 User Request backup interface failure. 842 Injection valve (S1) failure. 843 Vent valve (S2) failure. 844 Vacuum valve (S3) failure. 820.1 Change coaxial umbilical 820.2 Improve catheter tiptissue contact 847 Vent valve (S6) failure. 848 Baseline Flow 849 Check electrical connector Page 37 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Notice 60002 60003 60004 Rev-00 Problem Consequence Action properly. The vacuum is disabled due to a problem with the coaxial umbilical cable. procedure. You must take action in order to continue the procedure. You may run out of refrigerant cable and reconnect them. Disconnect both ends of the coaxial umbilical cable and reconnect them. Low refrigerant level Problem with scavenging You must take action in order to continue the procedure. Description 850 Check vacuum connector Replace refrigerant tank 851 Check refrigerant tank level Ensure that the hospital scavenging system is turned on and that the scavenging hose is securely attached. 852 Check scavenging system Table 5- System Notices Medtronic CryoCath LP Page 38 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Appendix B– Additional Warning Messages WARNING Messages Check refrigerant tank level Check mechanical connectors Possible Corrective Action • Verify the cylinder level and cylinder pressure in the system set-up panel. • Center the cylinder on the plate support (Load Cell). • Verify the hose connection has no bends or kinks. • Verify the actual cylinder weight, using a hospital’s calibrated scale. • Replace the cylinder after completion of the procedure. • Verify the coaxial connections • Replace the coaxial umbilical • Replace the catheter Note: This animation can occur in the patient screen when the baseline flow is around the threshold value of 100 sccm. This usually means there is a leak in the system or the flow meter needs to be calibrated. Vacuum Connection Panel Electrical Connector Animation Scavenging Connector Animation Upon pressing “OK” field and the warning persists: • Verify coaxial connections, then “OK”. If the warning persists, • Replace the coaxial umbilical, then “OK”. If the warning persists, • Replace the catheter • • • • Verify the electrical connections Replace the ECG box Replace the Electrical Umbilical Replace the catheter • Verify the scavenging connection at the CryoConsole and wall outlet. • Verify the hospital’s scavenging system is working (in case of an active system). Table 6- Warning messages corrective action Medtronic CryoCath LP Page 39 of 41 TRM-002 MDT.2 CRYOCONSOLE – GEN IV Rev-00 Appendix C– Guideline & Specifications for N2O refilling ¾ The refrigerant to be refilled is N2O (UN 1070) 99.5% liquid phase with a humidity level less than 50ppm. It is important that the refiller supply the user with a “Certificate of Analysis” (at each refilling), which would provide proof that the refilled refrigerant met or exceeded the specifications indicated above. It is also important that the ‘Certificate of Analysis’ refers to the gas quality after it has been put into the refrigerant tank. ¾ The refrigerant must be refilled in the supplied Medtronic CryoCath LP N2O tank. The approximate amount of refrigerant to refill is 10lbs for North America and 8.3lbs for Europe. These tanks consist of three (3) parts: cylinder, dip-tube and valve. It is important that the gas supplier be aware of the stainless steel dip-tube. The gas supplier must connect to the N2O tank valve outlet connector “CEODEUX C10500275”. ¾ Provided that the gas refiller has the proper equipment and adaptors to attach to the Medtronic CryoCath LP N2O tank and that they provide an appropriate Certificate of Analysis on the refrigerant quality specifications, then it is acceptable for Medtonic customers to obtain their N2O tank refilled by a supplier of their choice. ¾ All parties handling the refrigerant tanks must be aware that when the N2O tanks have any refrigerant in them at all they are considered to be containing dangerous goods (I.A.T.A Class 2.2 Non Flammable, Non-toxic Gas and Class 5.1 Gases Oxidizer) and should be handled accordingly. Medtronic CryoCath LP Page 40 of 41