Download DNA 5K Assay User Guide

Transcript
DNA 5K Assay
User Guide
For LabChip GX Touch/GXII Touch
Copyright 2014 PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of
PerkinElmer, Inc. All other trademarks are the property of their respective owners.
PN CLS140147, Rev. C
2
Contents
Specifications ......................................................................................................... 3
Assay Specifications .......................................................................................... 3
Sample Conditions ............................................................................................. 3
Kit Contents ....................................................................................................... 4
Safety Warnings and Precautions.......................................................................... 6
Preparation Procedures ......................................................................................... 7
Additional Items Required .................................................................................. 7
Preparing the Gel-Dye Solution .......................................................................... 7
Preparing the DNA Samples, DNA Ladder and the Buffer Tube .......................... 8
Preparing the Chip ............................................................................................. 9
Inserting a Chip into the LabChip GX Touch/GXII Touch Instrument ................. 10
Running the Assay ........................................................................................... 13
Storing the Chip ............................................................................................... 16
Chip Cartridge Cleaning ................................................................................... 17
Results .................................................................................................................. 18
DNA 5K Ladder Result ..................................................................................... 18
Troubleshooting ................................................................................................... 19
LabChip Kit Essential Practices........................................................................... 25
General ............................................................................................................ 25
Reagents .......................................................................................................... 26
Chips ................................................................................................................ 26
Samples ........................................................................................................... 29
Chip Well Aspiration Using a Vacuum ................................................................. 30
Customer Technical Support ............................................................................... 31
Licenses and Rights of Use ................................................................................. 32
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Specifications
3
Specifications
Assay Specifications
Table 1. Assay Specifications
Sizing Range
100-5000bp
Sizing Resolutiona
± 10% from 150-500 bp
± 15% from 100-150 bp, 500-1500 bp
± 20% from 1500-5000 bp
Sizing Accuracy
± 10%
Sizing Precision
5% CV
Linear
Concentration
Range
0.25 ng/µL - 50 ng/µL per fragment
Sensitivity
0.25 ng/µL
Maximum Total DNA 80 ng/µL total, 50 ng/µL per fragment
Concentration
Quantitation
Accuracy
± 30% or ± 1 ng/µL, whichever is greater
Quantitation
Precision
20% CV
a. Resolution is defined as half height or better separation of two peaks. Actual
separation performance can depend on the sample and application. Peaks that
are resolved less than half height can still be accurately identified by the system
software.
Sample Conditions
Table 2. Sample Conditions
Additives
PN CLS140147, Rev. C
PerkinElmer recommends that BSA and
detergents exceeding 0.05 mg/mL and 0.01%
(v/v) respectively in concentration not be
used. Higher concentrations can result in chip
failure. In addition, non-aqueous solvents are
not compatible with DNA LabChip protocols
DNA 5K Assay User Guide
Specifications
4
Table 2. Sample Conditions(Continued)
Particulates
All sample plates should be spun down prior
to analysis. All buffers should be filtered with
a 0.22 µm cellulose acetate filter.
Salt
Concentration
Total salt concentration must not exceed
125 mM.
Plasmids
Plasmid concentration in samples must be
below 20 ng/µL. Please note that although the
DNA Assays cannot analyze plasmids, the
presence of plasmids above 20 ng/µL can
interfere with assay results
Kit Contents
Storage: When not in use, store chips and reagents refrigerated at
4°C. Do not leave chips and reagents unrefrigerated overnight.
DNA 5K Reagent Kit
Part Number CLS760675
Each kit contains enough reagents for 20 Low-throughput (LT) or 10
High-throughput (HT) chip preparations. Up to 48 samples can be
tested with a LT chip preparation. Up to 384 samples can be tested
with a HT chip preparation.
Table 3. Reagent Kit Contents, PN CLS760675
Reagent
Vial
Quantity
DNA Dye Concentrate
Blue
1 vial, 0.09 mL
Chip Storage Buffer
White
9 vials, 1.8 mL each
DNA Gel Matrix
Red
5 vials, 1.1 mL each
10X DNA Ladder
Yellow
1 vial, 0.26 mL
DNA Marker
Green
1 vial, 1.6 mL
Table 4. Consumable Items
Item
Supplier and Catalog Number Quantity
Spin Filters
Costar, Cat. # 8160
10
Ladder Tubes
Genemate, Cat. # C-3258-1
20, 0.2 mL
Detection Window VWR, Cat. # 21912-046
Cleaning Cloth
PN CLS140147, Rev. C
1
DNA 5K Assay User Guide
Specifications
5
Table 4. Consumable Items (Continued)
Item
Supplier and Catalog Number Quantity
Swab
ITW Texwipe®, Cat. # TX758B
3
Buffer Tubes
E&K Scientific, Cat. # 697075NC
20, 0.75 mL
Table 5. DNA LabChips
Item
Catalog Number
DNA 5K/RNA/CZE Chip for use with GX
Touch/GXII Touch HT
Cat. # 760435
DNA 5K/RNA/CZE Chip for use with GX
Touch/GXII Touch 24
Cat. # CLS138949
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Safety Warnings and Precautions
6
Safety Warnings and Precautions
WARNING!
For Research Use Only. Not recommended or intended for
diagnosis of disease in humans or animals. Do not use internally or
externally in humans or animals.
CAUTION
We recommend that this product and components be handled only
by those who have been trained in laboratory techniques and that it
is used in accordance with the principles of good laboratory
practice. As all chemicals should be considered as potentially
hazardous, it is advisable when handling chemical reagents to wear
suitable protective clothing, such as laboratory overalls, safety
glasses, and gloves. Care should be taken to avoid contact with
skin or eyes. In case of contact with skin or eyes, wash immediately
with water
WARNING!
Dye Concentrate contains DMSO. S24/25: Avoid contact with skin
and eyes.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
7
Preparation Procedures
Additional Items Required
•
18 megohm, 0.22-µm filtered water (Milli-Q® or equivalent).
•
70% isopropanol solution in DI water.
•
Bio-Rad Hard-Shell® 384-well Skirted PCR Plates, Cat. # HSP38XX (recommended)
•
PerkinElmer Hard-Shell thin-wall 96-well skirted PCR plate
(blue), Cat. # 6008870 (recommended)
Note: Allow the chip and reagents to equilibrate to room
temperature for about 20-30 minutes before use.
Preparing the Gel-Dye Solution
Notes: The Dye Concentrate contains DMSO and must be thawed
completely before use.
One vial of DNA Gel Matrix (red cap
) is good for 4 Lowthroughput chip preparations or 2 High-throughput chip
preparations. Up to 48 samples can be tested with a LT chip
preparation. Up to 384 samples can be tested with a HT chip
preparation.
1
Vortex the thawed DNA Dye Concentrate (blue cap
15 seconds before use.
) for 10-
2
Transfer 13 µL of DNA Dye Concentrate (blue cap
of DNA Gel Matrix (red cap
).
) to 1 vial
3
Vortex the solution until it is well mixed and spin down for a few
seconds.
4
Transfer the mixture into two spin filters (approximately 550 µL
each).
5
Centrifuge at 9300 rcf for 7.5 minutes at room temperature.
6
Discard filters, label and date the tubes
7
Store in the dark at 4°C. Use within 3 weeks.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
8
Preparing the DNA Samples, DNA Ladder and the
Buffer Tube
Notes: DNA Ladder should be prepared in the same buffer as your
DNA samples. A buffer mismatch between sample and ladder may
lead to inaccurate quantitation and sizing.
DNA sample buffer solution is the user's DNA buffer such as PCR
buffer, etc.
Standard Sample Workflow
Buffer Tube
Ladder Tube
Sample
Well A1
Ladder
Tube
Buffer
Tube
Figure 1. Locations of the Buffer Tube and Ladder Tube in the
GX Touch/GXII Touch instrument.
Preparing the Ladder Tube
1
Gently vortex DNA Ladder (yellow cap
Briefly spin the ladder vial.
2
In the provided 0.2 mL Ladder Tube, add 12 µL of DNA Ladder
to 108 µL of 1X DNA sample buffer solution. Mix thoroughly by
pipetting the solution up and down several times. Ensure there
are no air bubbles in the Ladder Tube.
3
Insert the Ladder Tube into the ladder slot on the LabChip GX
Touch/GXII Touch instrument.
4
Recommended sample volumes are 25 µL for a 384-well plate
or 40 µL for a 96-well plate.
PN CLS140147, Rev. C
) for 10 seconds.
DNA 5K Assay User Guide
Preparation Procedures
9
Preparing the Buffer Tube
1
Add 750 µL of 1X DNA Sample Buffer solution to the 0.75 mL
Buffer Tube provided with the reagent kit. Ensure there are no
air bubbles in the Buffer Tube.
2
Insert the Buffer Tube into the buffer slot on the LabChip GX
Touch/GXII Touch instrument.
Preparing the Chip
1
Allow the chip to come to room temperature.
2
Use a pipette tip attached to a vacuum line to thoroughly
aspirate all fluid from the chip wells (see Figure 2). For more
details on how to set up a vacuum line see page 30.
Figure 2. Using a vacuum to aspirate the chip wells is more
effective than using a pipette.
3
Rinse and completely aspirate each active chip well (1, 3, 4, 7, 8
and 10) twice with water. Do not allow active wells to remain dry.
4
If any water spills onto the top and bottom chip surfaces during
rinsing, aspirate using the vacuum line. DO NOT run the tip over
the central region of the detection window. Use the provided
Detection Window Cleaning Cloth dampened in water (Milli-Q ®
or equivalent) or alcohol to clean chip detection window as
needed.
5
Using a reverse pipetting technique, add Gel-Dye solution to
chip wells 3, 7, 8 and 10 as shown in Figure 3 (Low-throughput)
or Figure 4 (High-throughput).
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
Figure 3. Reagent placement
for Low-throughput (up to 48
samples).
6
10
Figure 4. Reagent placement
for High-throughput (up to
384 samples).
Add DNA Marker (green cap
) to chip well 4 as shown in
Figure 3 (Low-throughput) or Figure 4 (High-throughput). For
the High-throughput chip preparation add 50 µL DNA Marker for
96-well plates and 120 µL DNA Marker for 384-well plates or
multiple 96-well plate analysis.
Note: The marker well may need to be replenished if the chip is in
idle mode on the instrument for an extended period of time.
7
Make sure the rims of the chip wells are clean and dry.
8
IMPORTANT: Ensure chip well 1 (waste well) is empty before
placing the chip into the instrument.
Note: Use the Low-throughput protocol when running the LabChip
GX Touch/GXII Touch 24 instrument.
Inserting a Chip into the LabChip GX Touch/GXII Touch
Instrument
1
Check that the sample plate, Buffer Tube, and Ladder Tube are
properly placed on the instrument.
2
Remove the chip from the chip storage container and inspect
the chip window. Clean BOTH sides of the chip window with the
PerkinElmer-supplied clean-room cloth dampened with a 70%
isopropanol solution in DI water.
3
Touch the Unload Chip button on the Home screen.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
11
Figure 5. Home screen.
4
Insert the chip into the LabChip GX Touch/GXII Touch
instrument (Figure 6) and close the chip door securely.
Figure 6. Chip in the LabChip GX Touch/GXII Touch instrument.
5
Touch the Load Plate button on the Home screen (Figure 5) to
retract the sample plate and send the sipper to the Buffer Tube.
Note: Do not keep the chip door open for any length of time. Dye is
sensitive to light and can be photobleached.
6
PN CLS140147, Rev. C
The Assay Choice window opens (Figure 7). Touch the desired
assay and then touch OK.
DNA 5K Assay User Guide
Preparation Procedures
12
Figure 7. Assay Choice menu.
Notes: If performing multiple runs in a day, in between chip
preparations the chip should be washed using the instrument and
Chip Storage buffer as described in “Storing the Chip” on page 16.
Be sure to periodically clean the O-rings on the top plate of the chip
interface on the LabChip GX Touch/GXII Touch. Use the provided
lint-free swab dampened with water (Milli-Q® or equivalent) to clean
the O-rings using a circular motion. Allow the O-rings to dry before
inserting a chip.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
13
Running the Assay
Note: Chips can be primed independently from running assays.
Touch the Prime button on the Home screen. Make sure the Buffer
Tube is placed on the instrument.
Figure 8. Chip priming screen.
1
Touch the Run button (see Figure 8).
2
Select the appropriate assay type (see Figure 7), plate name,
well pattern, and whether to read wells in columns or rows.
Select number of times each well is sampled under Adv.
Settings (Figure 9). Touch the green arrow button.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
14
Figure 9. Selecting wells.
3
PN CLS140147, Rev. C
In the Setup Run tab, select the operator name, the option to
read barcode, the destination of the file, the inclusion of sample
names, expected peaks, and excluded peaks and the filename
convention. Select Auto Export to export results tables
automatically (Figure 10). Touch the green arrow button.
DNA 5K Assay User Guide
Preparation Procedures
15
Figure 10. Run setup screen.
4
Touch Start to begin the run (Figure 11).
Figure 11. Starting a run.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
16
Storing the Chip
After use, the chip must be cleaned and stored in the chip container.
The procedure below can be conducted the following day when
running overnight.
1
Place the chip into the plastic storage container. The sipper
should be submerged in the fluid reservoir.
2
Remove the reagents from each well of the chip using vacuum.
3
Each active well (1, 3, 4, 7, 8 and 10) should be rinsed and
aspirated twice with water (Milli-Q® or equivalent).
4
Add 120 µL of DNA Storage Buffer (white cap
wells.
5
Place the chip in the LabChip GX Touch/GXII Touch instrument
and touch the Wash button in the upper right corner in the Home
Screen.
) to the active
Figure 12. Wash screen.
6
Remove the chip from the instrument and place it in the plastic
storage container. Add an additional amount of Storage Buffer to
well 1.
7
Cover the wells with Parafilm® to prevent evaporation and store
at 4°C. Storage of a chip with dry wells may cause it to become
clogged.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Preparation Procedures
17
Chip Cartridge Cleaning
1
2
PN CLS140147, Rev. C
Daily
a
Inspect the inside of the chip cartridge and O-rings for
debris.
b
Use the provided lint-free swab dampened with water (MilliQ® or equivalent) to clean the O-rings using a circular
motion. If the O-rings stick to the chip or a pressure leak is
detected, perform the more extensive monthly cleaning
procedure.
Monthly
a
To reduce pressure leaks at the chip interface, clean the Orings frequently. Remove the O-rings from the top plate of
the chip interface on the LabChip GX Touch/GXII instrument.
Soak O-rings in water (Milli-Q ® or equivalent) for a few
minutes. Clean the O-ring faces by rubbing between two
fingers. Wear gloves.
b
To reduce the occurrence of current leaks, clean the chip
interface frequently. Clean the top plate of the chip interface
using the provided lint free swab dampened with water (MilliQ® or equivalent).
c
Allow the O-rings and chip interface to air dry. Reinsert the
O-rings into the chip cartridge.
DNA 5K Assay User Guide
Results
18
Results
DNA 5K Ladder Result
The electropherogram of a typical DNA 5K ladder is shown in
Figure 13. Between the upper and lower markers, peaks in order of
increasing migration time correspond to ladder fragments of 100,
300, 500, 700, 1100, 1900, 2900, and 4900 bp.
.
Lower
Marker
Upper
Marker
Figure 13. DNA 5K ladder electropherogram produced using
the Standard Sample Workflow.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Troubleshooting
19
Troubleshooting
Note: Some of the data examples shown in this section were
generated with assays other than the assay described in this user
guide.
Symptom: No ladder or sample peaks but marker peaks detected.
Note: The lower marker peak height will most likely be greater than
normal height.
Possible causes:
1
Air bubble in sipper introduced during chip priming.
What to do:
1
Reprime the chip. See “LabChip Kit Essential Practices” on
page 25 for instructions on how to reprime the chip.
Symptom: Missing sample, ladder and marker peaks.
Possible causes:
1
Clog in sipper or marker channel of chip.
What to do:
1
Reprime the chip. See “LabChip Kit Essential Practices” on
page 25 for instructions on how to reprime the chip.
Symptom: Ladder detected but no sample peaks.
Possible causes:
1
The sipper is not reaching the sample due to low sample volume
in the well of the plate.
2
If the missing sample peaks occurred only in a few wells of the
plate, check those wells for air bubbles.
3
The sipper is not reaching the sample due to an incorrect
capillary height setting or incorrect plate definition.
4
If the plate has been uncovered for some time, sample
evaporation might have occurred.
5
Debris from the sample or sample prep is clogging the sipper.
What to do:
1
PN CLS140147, Rev. C
Add more sample to the well.
DNA 5K Assay User Guide
Troubleshooting
20
2
Manually insert a larger volume pipette tip (~100 µL) into the
sample well and dislodge the bubble. Rerun these sample wells.
3
Check the plate definitions.
4
Check the sample wells, especially around the edge of the plate
where evaporation is fastest, and make a fresh plate if volumes
are low.
5
If you suspect there may be debris in your samples, spin the
sample plate down in a centrifuge (e.g. 3000 rcf for 5 minutes).
Unclog the sipper by repriming the chip. See “LabChip Kit
Essential Practices” on page 25 for instructions on how to
reprime the chip.
Symptom: No ladder peaks but sample peaks and marker peaks are
present.
Possible causes:
1
Low or no ladder volume in the Ladder Tube.
What to do:
1
Add more ladder to the Ladder Tube and restart the run.
Recommended standard ladder volume is 120 µL (minimum
volume is 100 µL).
Symptom: No marker peaks but sample peaks are present.
Possible causes:
1
No marker added to chip well 4.
2
If there is marker solution in chip well 4, the problem may be
due to a marker channel clog.
What to do:
1
PN CLS140147, Rev. C
This may be due to not filling marker well or chip remaining idle
on instrument for extended period of time. Add or replenish the
marker solution in the chip using the following procedure:
• Touch the Unload Chip button on the Home screen to open
the chip door.
• Return the chip to the chip container ensuring the sipper is
immersed in fluid.
• Thoroughly aspirate all fluid from chip well 4 using a vacuum
line.
• Ensure that chip well 4 is rinsed and completely aspirated
twice with water (Milli-Q® or equivalent).
• Add Marker Solution (green cap
) to chip well 4.
DNA 5K Assay User Guide
Troubleshooting
•
•
2
21
Reinsert the chip back into the instrument.
Restart the run.
Perform a marker channel unclogging procedure by repriming
the chip. See “LabChip Kit Essential Practices” on page 25 for
instructions on how to reprime the chip.
Symptom: Ladder traces show up in the lanes following the ladders
(delayed sip).
Figure 14. Small ladder peaks in sample well caused by
delayed sip.
Possible causes:
1
Separation channel overloaded with sample.
2
Partial clog in the separation channel.
What to do:
1
Lower the starting sample concentration.
2
Reprime the chip. See “LabChip Kit Essential Practices” on
page 25 for instructions on how to reprime the chip.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Troubleshooting
22
Symptom: Unexpected sharp peaks.
Figure 15. Unexpected sharp peak.
Possible causes:
•
Dust or other particulates introduced through sample or
reagents.
What to do:
1
Do one or all of the following:
• Replace the 18 megohm, 0.22-µm filtered water (Milli-Q® or
equivalent) water used for chip preparation.
• Replace the buffer used for sample and reagent preparation.
• Use a 0.22-micron filter for all water and buffers used for
chip, sample, and reagent preparation.
• Spin down sample plate to pellet any particulates.
Symptom: Humps in several electropherograms which do not
correspond to sample data.
Figure 16. Humps in several electropherograms.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Troubleshooting
23
Possible causes:
1
Electrode 7 is dirty and has contaminated the Gel-Dye mixture
in well 7.
What to do:
1
Before restarting the run, clean electrode 7. Remove the chip
and follow the electrode cleaning procedure. We recommend
using the provided swab and isopropanol to manually clean
electrode 7.
Symptom: Peaks migrating much faster or slower than expected.
Note: Some migration time variance between chips or within a plate
is considered normal chip performance. All chips are QC tested at
PerkinElmer prior to shipment.
Normal migration time windows for the markers are:
•
•
DNA 5K assay Lower Marker (9 - 12 seconds)
DNA 5K assay Upper Marker (18 - 19 seconds)
Possible causes:
1
Incorrect Gel-Dye ratio. Migration time is sensitive to dye
concentration and peaks will migrate too fast or too slow if the
dye concentration in the gel is too low or too high, respectively.
Note: Excess dye within the separation channel will slow down
migration, and less dye in the separation channel will make peaks
migrate faster.
2
Particulates from the samples may be clogging the separation
channel (this will slow down migration).
3
Gel-Dye was not primed properly into the chip.
What to do:
1
Prepare a fresh Gel-Dye solution. Wash and reprime the chip
with the new Gel-Dye mixture. See “LabChip Kit Essential
Practices” on page 25 for instructions on how to wash and
reprime the chip.
2
If fast or slow migration is observed repeatedly on a new chip,
contact technical support to arrange return of the chip to
PerkinElmer. Please send a data file showing the failure along
with the return request.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Troubleshooting
24
3
Minimize the loading of particulates in the sample by performing
a centrifuge spin of the sample plate (e.g. 3000 rcf for
5 minutes) before starting a new run. The debris may be flushed
out of the chip by washing and re-priming the chip. See
“LabChip Kit Essential Practices” on page 25 for instructions on
how to wash and reprime the chip.
4
Check the O-rings on the top surface of the chip interface and
clean if necessary.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
LabChip Kit Essential Practices
25
LabChip Kit Essential Practices
To ensure proper assay performance, please follow the important
handling practices described below.Failure to observe these
guidelines may void the LabChip Kit product warranty.1
Note: It is important to keep particulates out of the chip wells,
channels and capillary. Many of the following guidelines are
designed to keep the chips particulate-free.
For assay and instrument troubleshooting, refer to the LabChip GX
Touch software Help file or call PerkinElmer Technical Support at 1800-762-4000.
General
•
Allow the chip, sample plate and all reagents to equilibrate to
room temperature before use (approximately 20 to 30 minutes).
•
Clean the O-rings in the chip interface weekly and the
electrodes daily. Refer to the Instrument Users Guide
Maintenance and Service section for procedures.
•
Avoid use of powdered gloves. Use only non-powdered gloves
when handling chips, reagents, sample plates, and when
cleaning the instrument electrodes and electrode block.
•
Calibrate laboratory pipettes regularly to ensure proper reagent
dispensing.
•
Only the PerkinElmer-supplied clean room cloth can be used on
the chip to clean the detection window.
•
Water used for chip preparation procedures must be
18 megohm, 0.22-µm filtered water (Milli-Q® or equivalent).
•
Using the “Reverse Pipetting Technique” (described next) will
help avoid introducing bubbles into the chip when pipetting the
gel.
1.
PerkinElmer, Inc. warrants that the LabChip Kit meets specification at the time of
shipment, and is free from defects in material and workmanship. LabChip Kits are
warranted for 90 days from the date of shipment. All claims under this warranty must be
made within thirty days of the discovery of the defect.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
LabChip Kit Essential Practices
26
Reverse Pipetting Technique
Figure 17. Reverse pipetting.
1
Depress the pipette plunger to the second stop.
2
Aspirate the selected volume plus an excess amount from the
tube.
3
Dispense the selected volume into the corner of the well by
depressing plunger to the first stop.
4
Withdraw the pipette from the well.
•
Store reagents at 4°C when not in use.
•
The LabChip dye contains DMSO and should be thawed
completely before use. It is recommended that you prepare
aliquots to reduce the time required for thawing.
•
Gently vortex all kit reagents before use.
•
Dispense reagents into chip wells slowly without introducing air
bubbles. Insert the pipette tip vertically and to the bottom of the
chip well.
•
Protect the dye and Gel-Dye mixture from light. Store in the dark
at 4°C when not in use.
•
The Gel-Dye mixture expires 3 weeks after preparation.
Reagents
Chips
Repriming Chips
Note: Buffer tubes filled with 1X DNA sample buffer or water should
be placed into the instrument while priming or washing chips.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
LabChip Kit Essential Practices
27
•
Touch the Unload Chip button on the Home screen to open the
instrument door. Place the chip into the instrument.
•
Close the chip door securely and choose the corresponding
assay.
•
Touch the Prime button on the Home screen to reprime the chip.
Washing and Repriming Chips
•
Touch the Unload Chip button on the Home screen to open the
instrument door.
•
Return the chip to the chip container ensuring the sipper is
immersed in fluid.
•
Thoroughly aspirate all fluid from the chip wells using a vacuum
line.
•
Ensure that each active well (1, 3, 4, 7, 8 and 10) is rinsed and
completely aspirated twice with water (Milli-Q® or equivalent).
Do not allow active wells to remain dry.
•
Add 120 µL of Chip Storage Buffer to each active well (1, 3, 4, 7,
8 and 10).
•
Place the chip in the LabChip GX Touch/GXII Touch instrument.
•
Close the chip door securely
•
Touch the Wash button on the Home screen (Figure 18).
Figure 18. Wash screen.
•
PN CLS140147, Rev. C
After the completion of the wash cycle return the chip to the chip
container ensuring the sipper is immersed in fluid.
DNA 5K Assay User Guide
LabChip Kit Essential Practices
28
•
Thoroughly aspirate all fluid from the chip wells using a vacuum
line.
•
Ensure that each active well (1, 3, 4, 7, 8 and 10) is rinsed and
completely aspirated twice with water (Milli-Q® or equivalent).
Do not allow active wells to remain dry.
•
Add Gel-Dye solution to chip wells 3, 7, 8 and 10 using a
Reverse Pipetting Technique as shown in Figure 17.
•
Add Marker (green cap
) to chip well 4. Use 50 µL for 96-well
or 120 µL for 384- well plate analysis. Please note that the
marker well may need to be replenished if the chip is in idle
mode on the instrument for an extended period of time.
•
Place the chip into the LabChip GX Touch/GXII Touch
instrument.
•
Close the chip door securely.
•
Touch the Run or Prime button on the Home screen.
•
If air bubbles are not dislodged after a reprime, apply a vacuum
to the sipper. Perform this by filling all active wells with 100 µL
of Chip Storage Buffer. Then suction the sipper with a vacuum
line as shown in Figure 19 until droplets of fluid flow out from the
sipper. When suctioning the sipper, be careful not to bend or
break the sipper. To facilitate this, cut the end of the pipette tip
attached to the vacuum line to widen the mouth.
Figure 19. Removing an air bubble or clog by suctioning the
sipper with a vacuum line.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
LabChip Kit Essential Practices
29
Other Considerations:
•
Chips should be stored refrigerated prior to first use.
•
Cover the active wells on the chip with Parafilm® and store at
4°C. If using the chip again within 24 hours it may be left at room
temperature.
•
Do not allow the liquid in the chip container to freeze, as this
may lead to poor chip performance. Do not submerge the chip in
any solution.
•
The entire chip surface must be thoroughly dry before use.
•
The sipper must be kept immersed in fluid at all times and
should not be exposed to an open environment for long periods
of time.
•
Use care in chip handling to prevent sipper damage. Damage to
the sipper can result in inconsistent sampling.
•
Avoid exposing the chips to dust by keeping them in a closed
environment such as in the chip container or in the instrument
before and after chip preparation.
•
Chips can be prepared and left idle on the instrument for up to 8
hours. This workflow allows analysis of samples as needed
throughout the day without having to re-prep the chip as long as
the maximum number of samples per chip prep is not exceeded.
•
PerkinElmer recommends the chip be re-prepared after it has
been idle for 8 hours.
•
Prepared sample plates should be free of gas bubbles and
particulate debris, both of which may inhibit sipper flow.
•
Sample plates containing gas bubbles and/or particulate debris
should be spun down at 3000 rpm (1250 rcf) prior to analysis.
•
Up to 96 samples in a 96-well or 384-well plate can be run with
a single chip preparation.
Samples
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Chip Well Aspiration Using a Vacuum
30
Chip Well Aspiration Using a Vacuum
Aspirating with a pipette can leave used reagents in the chip wells.
For this reason, PerkinElmer recommends vacuuming the wells
instead. This can be accomplished by attaching a permanent
pipette tip to a house vacuum line with trap (Figure 20). To avoid
contamination, use a new disposable pipette tip over the permanent
tip for each chip aspirated (Figure 21).
B
A
Figure 20. A: Permanent pipette tip attached to a house
vacuum line; B: vacuum line with trap.
Figure 21. Replacing the disposable pipette tip.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Customer Technical Support
31
Customer Technical Support
PerkinElmer, Inc.
68 Elm Street
Hopkinton, MA 01748-1668
PerkinElmer Technical Support
Phone (USA Toll Free): 800-762-4000
Phone (Worldwide): +1 203-925-4602
Fax: +1 203-925-4602
Email: [email protected]
Internet: www.perkinelmer.com
For additional assay and instrument troubleshooting, refer to the
LabChip GX Touch Software Help file.
PN CLS140147, Rev. C
DNA 5K Assay User Guide
Licenses and Rights of Use
32
Licenses and Rights of Use
Label Licenses
This product is provided under an intellectual property license from Life
Technologies Corporation. The purchase of this product conveys to the buyer the
non-transferable right to use the data for internal research and training purposes
solely in combination with PerkinElmer, Inc. instrumentation, and not to generate
revenue, which may include, but is not limited to use of the product: (1) in
manufacturing or in quality assurance or quality control of a commercial product;
(2) to provide a service, information, or data for a fee or other consideration; (3) for
therapeutic or prophylactic purposes; (4) for diagnostic use; and (5) for resale,
whether or not such items are resold for use in research. For information on
purchasing a license to this product for any other purposes, contact Life
Technologies Corporation, 5791 Van Allen Way, Carlsbad, CA 92008 USA or
[email protected].
Rights of Use
The chip and reagents supplied with this kit are sold with limited rights of use. The
chip may only be used with the specific quantity of reagents supplied with this kit.
The purchaser has no right or license to refurbish, reuse, remanufacture, or
otherwise use the chip with any other reagents than those specifically supplied in
this kit. For more information on the terms and conditions of use of these chips and
reagents, please read your LabChip GX Touch User Guide and refer to the
applicable label license. The reagent kit contains materials that are provided under
a license from Life Technologies Corporation, for research use only. PerkinElmer,
LabChip, and the LabChip logo are registered trademarks of PerkinElmer, Inc.
and/or its parent, affiliates, and/or subsidiary companies (collectively
"PerkinElmer"). The PerkinElmer logo is a registered trademark of PerkinElmer,
Inc. All other trademarks and registered trademarks are the property of their
respective holders.
© 2014 PerkinElmer, Inc.
http://www.perkinelmer.com
PN CLS140147, Rev. C
DNA 5K Assay User Guide
PerkinElmer, Inc.
68 Elm Street
Hopkinton, Massachusetts 01748 U.S.A.
TEL 508-435-9500
FAX 508-435-3439
http://www.perkinelmer.com