HOW IT WORKS 7 polar blue dye will have a stronger partition affinity for the stationary phase. The more polar red dye favors the mobile phase, moves more rapidly down the column than the blue dye, and emerges first from the column into the detector. If we could see into the column we would see a purple disc move down the column, gradually separating into a fast moving red disc followed by a slower moving blue disc (Fig. 1.3). 1.1.2 Basic Hardware: A Quick, First Look The simplest HPLC system is made up of a high-pressure solvent pump, an injector, a column, a detector, and a data recorder (Fig. 1.4). Note: The high pressures referred to are of the order of 2000–6000 psi. Since we are working with liquids instead of gases, high pressures do not pose an explosion hazard. Leaks occur on overpressurizing; the worse problems to be expected are drips, streams, and puddles. Solvent (mobile phase) from a solvent reservoir is pulled up the solvent inlet line into the pump head through a one-way check valve. Pressurized in Figure 1.3 Separation model 2 (HPLC column). Figure 1.4 An isocratic HPLC system.