Download Operating Instructions
Transcript
Proline Kryomats 6.5 Connecting external loads Operation as circulating thermostat − When used as circulation thermostat, care for shortest hose connections with largest inner diameter as possible. This gives the best flow. − Push hose with 11-12 mm internal width onto hose olive (accessories) or connect metal hoses ( 6.3) to pump connectors. − Pump connectors at side: Inlet and outflow see labeling housing. − Pump connectors at back: Inlet and outflow see labeling housing. − Set bypass valve to "external" ( 3). − The external application must not have a hydraulically blocking effect. − If cross-sectional area of tube is too low temperature gradient between bath and external load due to low flow rate. − Always ensure the largest possible passages in the external circuit. − When tightening the union nuts on the pump nipple AF 19, use a wrench AF 14 to counter the tightening torque (see figure). − If external control is to be used, provide a Pt100 temperature probe in the external load ( 7.5.2 and 7.5.4). − With loads at a higher position and with stationary pump and ingress of air into the thermostatic circuit, the external volume can drain away, even with closed circuits Risk of thermostat overflowing! − Secure hoses against slippage with hose clips. − Unused pump connectors must be closed off. YAUE0007 / 3/7/2013 Preparation 33