Integral XT
7.12
Programmer
Almost any temperature/time profile can be created with the programmer. A desired bath temperature
can be approached as quickly as possible or via a defined ramp. Furthermore, the pump level and the
behavior of the switching outputs can be defined. Five temperature/time programs are provided for free
programming. Each program consists of a number of temperature/time segments. Also included are
details of how often the program is to be executed (loops). The sum of all segments of all programs
may be up to a maximum of 150.
Typical segments are:
Ramp: If a time is specified, then the segment is a ramp which is described by the target temperature,
i.e. the temperature at the end of the segment, and the duration from the start to the end of the segment.
Step: Without any specified time the final temperature is approached as quickly as possible.
Temperature hold phase: No temperature change (i.e. the temperatures at the start and end of a
segment are the same).
+
The programmer can be controlled or changed via the RS 232 interface, the timer or switching contacts.
7.12.1 Program example
°C
80
70
60
50
40
30
Start
20
10
0
20
40
60
Pump
T end °C
30,00°C
30,00°C
50,00°C
70,00°C
70,00°C
60,00°C
40,00°C
Menu
YAWE0028 / 29/01/2015
Time [h:m]
------00:20
00:20
00:40
00:10
00:30
00:00
End
100
120
t/min
Cooling time depends on
type of device, load, etc.
This is Segment No. 2
Real program example with 6 segments
No
Start
1
2
3
4
5
6
80
Tolerance
0,00°C
0,10°C
0,00°C
0,00°C
0,10°C
0,00°C
0,00°C
Insert
Delete
No
Start
1
2
3
4
5
6
Pump
------2
3
4
2
2
2
Pump
Starting up
Menu
Out 1
Out 2
Out 3
------------- ------------------- ------------------- ------------------- ------------------- ------------------- ------------------- -------
End
Insert
Delete