Operating instructions SIMDOS02 EN Cleaning and maintenance 8.2 Cleaning the pump In order to maintain the service life of the pump, after each use and before any long periods of inactivity the pump must be flushed through with neutral media. IMPORTANT: To ensure satisfactory start-up when the unit is again required, it is important to ensure that the pump is free of any crystallising, adhesive or curing media. 8.2.1 Metering medium: acids Flush pump with a suitable neutralising liquid in a circular flow for 10 minutes Then flush with distilled water for 5 minutes 8.2.2 Metering medium: bases Flush pump with a suitable neutralising liquid in a circular flow for 10 minutes Then flush with distilled water for 5 minutes 8.2.3 Metering medium: organic solvents Flush pump with isopropyl alcohol (C3H8O) in a circular flow for 10 minutes NOTE Never flush non-polar solvents immediately with water 8.2.4 Metering medium: biological solution Flush pump with 10% hydrogen peroxide in a circular flow for 10 minutes Then flush with distilled water for 10 minutes NOTE The procedure described in Section 8.2.4 is not a sterilisation KNF Flodos BA_SIMDOS02_EN_02_166267 Translation of original operating instructions, English 51