Download Operating instructions SIMDOS02 EN
Transcript
RC version (remote control) Operating instructions SIMDOS02 EN Flow rate range selection for analog input There are three flow rate ranges for different set points (see Fig. 59). 100%: 1 – 100% 30%: 0.3 – 30% 0.06 – 6 ml/min 15%: 0.15-15% 0.2– 20 ml/min 0.03 – 3 ml/min Fig. 59: Flow rate range selection 100% 30% 15% Tab. 27: Analog Input START / STOP analog signal Starting and stopping the pump with an analog signal 1. Set analog input to correct signal type 2. Press the “START” button. Pump switches to "Pause" condition. 3. Apply analog signal. The pump now starts operating, and switches to “Run” condition. 4. Reduce analog signal to below cut-off threshold. Pump stops and switches to "Pause" condition. NOTE After the pump has been stopped, there is a pause time of 0.2 seconds before it can be restarted. Master Stop Stopping the pump with the “STOP” button is called a “Master Stop”. The pump can then only be restarted manually (see Section 7.6). NOTE The pump is supplied with Master Stop. For initial start-up with external drive, it is therefore essential to press “START” once to cancel Master Stop. For an external restart after a power cut, use autostart (see Section 7.17) or Reset on digital input 2 (see Section 9.3). 62 KNF Flodos BA_SIMDOS02_EN_02_166267 Translation of original operating instructions, English