Download Operating instructions SIMDOS02 EN

Transcript
RC version (remote control)
Operating instructions SIMDOS02 EN
Flow rate range selection for analog input
There are three flow rate ranges for different set points (see Fig.
59).

100%: 1 – 100%

30%:
0.3 – 30% 0.06 – 6 ml/min

15%:
0.15-15%
0.2– 20 ml/min
0.03 – 3 ml/min
Fig. 59: Flow rate range selection
100%
30%
15%
Tab. 27: Analog Input
START / STOP analog signal
Starting and stopping the pump with an analog signal
1. Set analog input to correct signal type
2. Press the “START” button.

Pump switches to "Pause" condition.
3. Apply analog signal.

The pump now starts operating, and switches to “Run”
condition.
4. Reduce analog signal to below cut-off threshold.

Pump stops and switches to "Pause" condition.
NOTE
After the pump has been stopped, there is a pause time of 0.2
seconds before it can be restarted.
Master Stop
Stopping the pump with the “STOP” button is called a “Master
Stop”. The pump can then only be restarted manually (see Section
7.6).
NOTE
The pump is supplied with Master Stop. For initial start-up with
external drive, it is therefore essential to press “START” once to
cancel Master Stop.
For an external restart after a power cut, use autostart (see
Section 7.17) or Reset on digital input 2 (see Section 9.3).
62
KNF Flodos BA_SIMDOS02_EN_02_166267
Translation of original operating instructions, English