Download HP Saw E4419B User's Manual
Transcript
HP4402.book : chapter2.fm 59 Fri Feb 12 15:32:45 1999 Power Meter Operation Configuring the Remote Interface Configuring the Remote Interface This section explains how to configure the power meter for remote control using either the HP-IB parallel interface or the RS232/422 serial interface. For further information on using the power meter remotely, refer to the HP E4418B/E4419B Programming Guide. HP-IB To select HP-IB as the remote interface press System , Inputs Remote Interface , Select Interface , HPIB (IEEE 488). HP-IB Address Each device on the HP-IB (IEEE-488) interface must have a unique address. You can set the power meter’s address to any value between 0 and 30. The address is set to 13 when the power meter is shipped from the factory. The address is stored in non-volatile memory, and does not change when the power is switched off, or after a remote interface reset. Your HP-IB bus controller has its own address. Avoid using the bus controller’s address for any instrument on the interface bus. Hewlett-Packard controllers generally use address 21. To set the HP-IB address from the front panel: 1. Press System , Remote Interface , Configure Interface , Inputs HPIB . The current setting of the HP-IB address is displayed under the HP-IB Addr softkey. 2. To change this setting press HP-IB Addr . The power meter displays the address in a pop up window. Modify this address (see below) as desired. ■ Use or to modify the digit on which the cursor is currently positioned. ■ Use or to move to other digits. 3. To confirm your choice press Enter . To set the HP-IB address from the remote interface use the: • SYSTem:COMMunicate:GPIB:ADDRess command. HP E4419B User’s Guide 2-59