Download Hamilton Electronics EPHUM011 User's Manual
Transcript
pH ARC Sensors Modbus RTU Programmer’ s Manual (EPHUM011) Modbus address: 1568 Length: 8 Type: 16 Command: Info user Parameter: Text Format: Character Hello World Value: Figure 2.9.3.2: Example to write 16 ASCII characters to register 1568 with operator A or S. Write Modbus address: 1568 Length: 8 Command: Info user Parameter: Text Format: Character Hello World Value: Figure 2.9.3.3: Example to read the register 1568 (written in figure 2.9.1.6). Read Type: 3 2.10 System Commands 2.10.1 Recall Sensor’ s Factory Settings Using register 8192 you can recall the sensor manufacturer values (interfaces, calibration data and passwords), except the SIP and CIP data which remain unchanged. By sending the recall value “ 732255” , all configuration values will be set to default. Start Number of Reg1 / Reg2 register registers 8192 2 Recall by value 732255 Figure 2.10.1.1: Definition of register 8192. Modbus function code 16 Modbus address: 8192 Command: Recall Parameter: Recall Format: Decimal 732255 Value: Figure 2.10.1.2: Example to write the restore command. Length: 2 3 Abbreviations AO CP ECS PMC SMC 624300/01 Analog Output Interface Calibration Point Electrochemical Sensor Interface Primary Measurement Channel Secondary Measurement Channel page 66 / 67 Read access none Type: 16 Write access S Write