Maintenance 4 Replacing Parts Replacing a Level Sensing Cable Removing the level sensing cable 1 Detach the transfer tubing from the top of the isolation probe holder by using the supplied headless nut extender (part number 49041032) to loosen and remove the 1/4"-28 nut and ferrule. 2 Unplug the right-angle connector for the liquid level sensing cable (LLD cable) from the socket on the back of the Z-arm. 3 Detach the LLD cable from the LLD cable guide. 4 Loosen the small hexagonal nut on the probe holder for probe one. Detach the splittongue connector from the nut. 5 Remove the transfer tubing from the LLD cable casing. 4-11