COMMAND REFERENCE 4 - 33 CALCulate<n>:FEED Syntax: CALCulate<n>:FEED "<trace_name>" Note: The parameter "<trace_name>" is <string_data>. Therefore, it may be specified between single quotes as well, i.e., ’<trace_name>’. <n> [1] | 2 <trace_name> A trace name which is a predefined <acquisition_trace> or <memory_trace>. <acquisition_trace> CH1 | CH2 | CH3 | CH4 <memory_trace> Mi_1 | Mi_2 | Mi_3 | Mi_4 Note: - i = 1 .. 8 (standard memory) - i = 9 .. 50 (extended memory) Query form: CALCulate<n>:FEED? Response: "CHn" | "Mi_n" Note: - n = 1 .. 4 (n=1 .. 2 for PM33x0B) - i = 1 .. 8 (standard memory) - i = 9 .. 50 (extended memory) Description: The CALCulate:FEED command controls the source for the calculate function. The trace specified by <trace_name> is selected as source for the calculate block. After a *RST command, CH1 becomes the source for the CALCulate1 and CALCulate2 functions. Limitations: • • • • A channel must be ON before it can be selected. An empty trace may not be used as source in a CALCulate command. M1_i is not allowed as source for a CALCulate1 command. M2_i is not allowed as source for a CALCulate2 command. Example: Send → CALCulate2:FEED "CH3" Channel 3 becomes the source for MATH2. Send → CALCulate:FEED ’M8_4’ M8_4 becomes the source for MATH1. Front panel compliance: The CALCulate1 and CALCulate2 commands use the MATH1 and MATH2 features of the CombiScope instrument.