Download Agilent Technologies E1442

Transcript
Example: Differential
Divider or Filter
Configuration
Any channel can be configured as a differential divider (with optional filter)
connected to the normally open contact of the Form C relay. The differential
divider requires that two channels be used.
Figure 1-23 shows channel 24 and 25 in this configuration with the optional
filter. For resistors R24 and R25, add a cross-channel capacitor for a
differential filter or add a cross-channel resistor for a differential divider.
No configuration jumpers are required.
TERMINAL
MODULE
SWITCH
MODULE
COM
COM
NC
NC
Channel 25
NO
NO
R25
Cross-Channel
R or C
C
R
COM
COM
NC
NC
NO
Channel 24
NO
R24
COM
CH 25
User
Wiring
To
Terminal
Module
NC
NO
COM
CH 24
NC
NO
Channel Resistor
Solder Holes
Channel Capacitor
Solder Holes
Channel
Configuration
Jumpers
Channel
Number
Differential filter:
Add Capacitor
Differential Divider:
Add Resistor
RP24-31
RP16-23
V24-31
V16-23
RP8-15
V8-15
V0-7
RP0-7
User Supplied Resistor Packs (SIP) Locations
and associated pullup voltage (for optional standup resistors)
Figure 1-23. Example: Differential Divider or Filter Configuration
36 Getting Started
Chapter 1