Download Melco EMT 10T Specifications
Transcript
3 - 40 Rotary Hook 32. Next press . The head rotates one revolution. The LCD displays ONE REVOLUTION 110.0. The needle bar should be down with the presser foot about 1/4 inch from the needle plate. NOTE: If the needle does not come down the upper dead stop clamp may need adjustment as described later in this procedure. For now, however, complete the needle depth adjustment by grasping the presser foot and pulling down until the needle bar catches securely. At this time the presser foot will be about 1/4 inch from the needle plate. 33. Press . The head rotates to the ‘‘needle depth’’ position and the LCD displays: NEEDLE DEPTH 180.0. The needle has reached its lowest point. needle eye 34. Look directly into the hook assembly and check the location of the needle’s eye compared to Figure 3-38. Not more than 1/8 of the needle’s eye should be visible in the hook assembly. If the needle’s eye is in the acceptable range, skip the next few steps and go to Step 43. needle hook assembly (bobbin removed) Figure 3 - 38 If the needle’s eye is not in the acceptable range, continue this procedure with the next step. 35. Loosen the needle bar clamp screw (silver clamp) shown in Figure 3-39. 36. Slide the needle bar up or down until not more than 1/8 of the needle’s eye should be visible in the hook assembly. If you twisted the needle bar, return it to its original rotation before continuing. 37. Re-tighten the needle bar clamp screw. 38. Loosen the screw on the upper-dead-stop clamp (black clamp). Push the upper dead stop down to the needle bar clamp. Tighten slightly (but do not turn the upper dead stop). 39. Press and simultaneously. The head will rotate to the full needle up position (needle at its highest point) and the LCD displays TOP DEAD CENTER. EMT 10T Technical Manual Melco Embroidery Systems