Download COX SO-300 Specifications
Transcript
a) Connection of a 3-wire sensor to an INDUSTRIAL SCIENTIFIC central controller 1 wire (+) continuous power supply → No. 3 1 wire (-) continuous power supply (0 volt mass) → No. 2 1 output signal wire → No. 1 - Detector Control Unit + 24 Vcc 3 2 1 3 2 1 GND Signal b) Connection of a 2-wire sensor to an INDUSTRIAL SCIENTIFIC central controller - 1 wire (+) → No. 3 1 signal wire → No. 1 Formant la boucle de courant 4/20 mA sur 2 fils Detector Control Unit + 24 Vcc 3 2 1 Signal 3 2 1 c) Connection of a 3-wire CTX300 sensor to a non-INDUSTRIAL SCIENTIFIC controller and to an internal power supply. Control Unit« other » + 24 Vcc + Detector Power (1) GND R Signal (R) Maximum resistance = 200 Ω. (1) 15 = VCC = 32 18 = VCC = 30 for CO2 Max I: 130 mA 8 3 2 1