Download Bacharach H25-IR Specifications
Transcript
Specifications H25-IR H25 IR FAULT CODE TROUBLE SHOOTER 0020 INTERNAL CAL ERROR YES Fix loose or broken connection And recalibrate Indicates one or more of the 4 flow parameters collected during the internal cal cycle, deviate from the “factory” value by 25% or more NO Break in sample line In probe assembly NO Probe removed during Cal cycle NO YES NO YES Repeat internal cal cycle Bad or weak pump Probe tip filter clogged YES Replace internal filter and store new “factory” calibration Replace probe filter and recalibrate NO Break or restriction in Internal plumbing NO YES YES Replace pump and store New “Factory” calibration Internal filter clogged Internal leak empty or Defective (check leak Pressure) YES Replace internal leak, enter new leak Parameters in cal menu and store New “factory cal” Correct and recalibrate Bacharach Inc. CONFIDENTIAL for internal use only H25 IR FAULT CODE TROUBLE SHOOTER 0040 EXTERNAL CAL ERROR Indicates leak rate too low during External calibration cycle Probe removed or too far from leak during Cal cycle YES Repeat external cal cycle NO External leak rate too Low or empty NO YES NO YES Fix loose or broken connection And recalibrate Correct and recalibrate Bad or weak pump? Check flow on exhaust Break in sample line In probe assembly NO YES Replace pump and store New “Factory” calibration Break or restriction in Internal plumbing NO YES Correct and recalibrate Bacharach Inc. CONFIDENTIAL for internal use only 56 Instruction 3015-4342