Download Barnstead NANOpure DIamond UF Specifications
Transcript
MAINTENANCE AND SERVICING Note A small amount of water may drain from the ultrafilter when disconnected. 7. Using the tool supplied with the ultrafilter, simultaneously push up and hold the collet on the ultrafilter in while pulling the bottom fitting out of the connector. See photo. 8. Install new ultrafilter according to “Ultrafilter Installation” in the “Installation” section of this manual. 9. Sanitize the NANOpure DIamond TOC Life Science (UV/UF with TOC) according to the “System Sanitization” section of this manual. Tool used for ultrafilter removal Collet Top U-Shaped Tubing Connection Tubing with luer fitting Collet Velcro strap Collet Bottom connector Ultrafilter Installation 44 Ultrafilter Connections - Right side view