Download Agilent Technologies E6198B User manual

Transcript
B
Register Definitions
Card Type (R)Base + 0h
This register reads back the card type of the card. The N9378A
is uniquely identified by card type 05h.
Bit
7
6
5
4
3
2
1
0
Purpose
Card
Type
Card
Type
Card
Type
Card
Type
Card
Type
Card
Type
Card
Type
Card
Type
Setting
0
0
0
0
0
1
0
1
• Read Only
• Power On/Reset State = 05h
Card Configuration (R) Base + 1h
This register reads back the card configuration. The default
read back is FFh, but you can pull down bits to zero as required
to create a configuration byte unique to the loads installed.
(Note that you can use the mezzanine card registers described
below for additional load configuration information.)
Use the actions loadCardGetInfo to read this register.
Bit
7
6
5
4
3
2
1
0
Purpose
undefined
undefined
undefined
undefined
undefined
undefined
CC1
CC0
Setting
x
x
x
x
x
x
state
state
• Read Only
• x = undefined
Status & Control (R & W) Base + 2h
This register reads back the card's status. It also can be used to
reset the board or open all relays (OAR).
Card status is defined as the state of the relay timer. The relay
timer is designed for the armature relays (>4ms). The timer
restarts whenever the registers controlling the respective relays
are written to or the card is reset. The card remains in the busy
state until the timer has timed out.
B-36
Bit
7
6
5
4
3
2
1
0
Read
1
1
1
1
1
1
1
Busy~
E6198B Switch/Load Unit User Manual