ATI Model A15/82 Fluoride Monitor with Auto-Calibration 3. Replacement tubes are precut to the proper length. For acid feed tubes, place the tubing into the retaining blocks and reinstall the bottom block. Pull the tube around the roller and install the top block. 4. Sample tube replacement is essentially the same. However, first disconnect the sample pickup tubing from the inlet fitting and the outlet tubing from the outlet side of the pump. Remove the tubing clamp plate and push the tube holders out of their slots. Push on the tubing from inside the retaining block so that the fitting slides out of the block. Pull the old tubing off the fitting and replace with a new tube, making sure that the tubing is pushed onto the connector up to the stop. Push the fittings back into the retaining blocks and put the blocks back into the slots. Put the clamp plate onto the back of the pump and reconnect the sample inlet and outlet tubing. Figure 11 - Pump Tubing Assy 30 O&M Manual Rev-G (6/10)