Download Sencor SDF 760 Specifications
Transcript
2. Specifications Oxygen Measurement Ranges: mg/L: 0.0 to 60.0 mg/L % air saturation: 0.0 to 600.0% air saturation mbar: 0 to 1500 Accuracy: mg/L: ±0.1% plus 1 Least Significant Digit (LSD) % air saturation: ±0.1% plus 1 LSD mbar: ±1% plus 1 LSD Resolution: mg/L: 0.1%, or 0.01 mg/L , whichever is greater % air saturation: 0.1% air saturation mbar: 1 mbar Temperature Measurement Range: -5.0 to +50.0°C Accuracy: ±0.1°C Resolution: 0.01°C Barometric Pressure Measurement Range: 450 to 825 mmHg (600 to 1100 mbar) Accuracy: ±1% plus 1 LSD within ±10°C ambient temperature from calibration point Resolution: 1 mmHg Temperature Compensation The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and 45°C. The % air saturation mode is automatically temperature-compensated to an accuracy of ±0.5% of calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values between 5 and 45°C. 3