Download Agilent Technologies 6220B Technical data
Transcript
Methods with Acquisition Parameters Setting parameters to acquire a data file in All Ions MS/MS mode 3 Setting parameters to acquire a data file in All Ions MS/MS mode In order to acquire a data file in All Ions MS/MS mode, it is necessary to set up a method that has a Time Segment with at least two and a maximum of four different Experiments containing different Collision Energy values on a Q-TOF or different Fragmentor voltages on a TOF. The low energy Experiment provides MS information, and the high energy Experiments provide MS/MS fragment information. See the Qualitative Analysis Familiarization Guide for LC/MS for more information on analyzing a data file that was acquired in All Ions mode. If you have a Q-TOF instrument, you set the Collision Energy to 0 in the first Experiment. In the subsequent Experiments (channels), you set the Collision Energy to a higher value (for example, 20 V). If you are creating a method for a Q-TOF with different collision energies, you set the Fragmentor voltage at the same value in each experiment. It should be set to a low enough setting that no fragments are generated in the source but at a high enough setting that you get enough precursor ions transmitted to the skimmer. This value depends on the compounds you are analyzing and their stability. You set the Collision Energy and the Fragmentor voltage on the Source tab. Figure 43 First Experiment has a Collision Energy of 0 V Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 91