Head kV Effective mAs Slice collimation Slice width Feed/Rotation Rotation time Kernel Increment Image order CTDIw Effective dose ThinSliceSpi 120 320 0.75 mm 4.0 mm 6.8 mm 0.5 sec. H31f 4.0 mm ca-cr 50.6 mGy Male: 0.7 mSv Female: 0.8 mSv RoutineSpi 120 320 1.5 mm 8.0 mm 13.7 mm 0.5 sec. H31f 8.0 mm ca-cr 45.6 mGy Male: 1.1 mSv Female: 1.2 mSv PosteriorFossaOptimization Filter PFO: To reduce beam-hardening artifacts in head images, particularly in the base of the skull, use the Posterior Fossa Optimization (PFO) filter. Contrast medium IV injection Start delay 60 sec. Flow rate 2 ml/sec. Total amount 50 – 60 ml If you want to set up the scan with only one range, i. e. either 16 x 0.75 mm or 16 x 1.5 mm, you can easily delete one mode by clicking the chronicle with the right mouse button, and select cut. 97