Download A&D OP06 Instruction manual
Transcript
7-6 Correcting the Internal Mass Value The GP series balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed. The internal mass value is corrected as follows: Model GP-12K GP-20K GP-30K/GP-30KS GP-32K/ GP-32KS GP-61K/ GP-61KS GP-40K GP-60K/ GP-60KS GP-100K/GP-100KS GP-102K Correction reference value 10 kg 20 kg Correction range -5.0 g to +5.0 g 40 kg 60 kg -50 g to +50 g 100 kg -50 g to +50 g Example: Using the GP-20K 32