Download adept technology Adept XUSR User`s guide
Transcript
Chapter 9: Cleanroom Robots 9.1 Connections Figure 9-2. Cleanroom Connections 9.2 Requirements Table 9-2. Cleanroom Robot Requirements Vacuum source 0.80 m 3/min (28 ft3/min) minimum volumetric flow rate 6 mm of water (0.2 inches of water) differential pressure measured between the robot and the vacuum source 3/4 inch NPT female thread pipe fitting at the back of the robot Compressed air source Clean, dry, oil-free compressed air 75 psi (0.52 MPa) 1.4 SCFM (.04 m 3/min.) flow rate 3/8 inch NPT female thread pipe fitting at the back of the robot, flow regulator not supplied Quill inside diameter The inside diameter of the quill must be plugged by the user’s end-effector in order for sufficient vacuum to develop in the outer link. 9.3 Exclusions and Incompatibilities Table 9-3. Internally Mounted Hand Valves Installation considerations The internal air line normally used to supply the internally-mounted hand valves (Adept Option Kit P/N 02853-000) is instead used to provide vacuum to the bellows/outer link. One of the passive 6 mm user air lines would need to be used instead. Performance considerations The air exhausting from the internally-mounted hand valves (Adept Option Kit P/N 02853-000) may be of sufficient quantity/ quality to cause the robot to exceed Class 10 Particulate Limits. Recommendation For these reasons, Adept recommends mounting hand valves externally. Adept Cobra ePLC600/800 Robot, User’s Guide, Rev A Page 104 of 108