Download Bogen CM-810 Operating instructions
Transcript
Installation and Connection 3.4.1 Connecting the instrument air Connect the instrument air supply to the FIDOR. 3.4.2 Connecting the combustion air Connect the combustion air if an external combustion air supply is used 3.4.3 Connecting the fuel gas (hydrogen) Recommendation: – Fit a pressure monitor on the fuel gas cylinder which monitors the cylinder pressure and emits a signal at a certain residual pressure (e.g. 10 bar). – If the FIDOR is integrated in an enclosed system cabinet: Install an H2 flow limiter for the fuel gas supply in the system cabinet and, if required, install a hydrogen sensor (H2 Sensor) (< 25% LEL). Lay the lines. – Only use analytically pure tubes made of copper or stainless steel. – Do not contaminate the insides of tubes during assembly Connect the fuel gas to the “Fuel gas” connection. ● WARNING: Risk of explosions The screw fitting for the fuel gas supply contains a flow limiter. ● The fuel gas inlet is marked “Fuel Gas”. Do not modify this screw fitting. Do not connect the fuel gas supply to any other screw fitting. Connecting the test gases 3.4.4.1 Zero gas Connect the zero gas. 3.4.4.2 Reference gas Recommendation: Monitor the feed pressure of the reference gas with a pressure monitor which emits an electric signal at a certain minimum pressure (e.g. 10 bar). Connect the signal of the pressure monitor with the “External signal failure” control input. Connect the reference gas. Subject to change without notice 3.4.4 24 GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V 1-0 · © SICK AG