16 Symbols, messages, problems and error messages 16.3 Problems The meter will not turn on. The batteries are empty or none are inserted. Insert new batteries. you have inserted the batteries the wrong way round. Remove the batteries and reinsert them as shown in the battery compartment. The ambient temperature is low and the batteries are almost empty. Move to a place where the temperature is between +10 and +40 °C and wait for the temperature of the meter to adjust to this temperature. Insert new batteries. 228 Condensation has affected the electronics. Allow the meter time to dry slowly. The meter is defective. Contact Accu-Chek Customer Care at 1-800-363-7949.