Download Siemens DataNET Hub (DNH) User`s manual
Transcript
DataNET Hub Specifications, Continued The Table below compares key communication features between using Advance Data Hiway, Ethernet, or DataNET. Communication Comparisons Feature Advance Data Hiway Ethernet DataNET Speed 14.2 KB/Sec 10 M baud 2. 048 M baud Redundant Yes No Yes Number of Units in Maxum System 248 1200 1200 DataNET Cable The following Table should be used when evaluating or selecting cable for an existing or new DataNET installation. Cable Type Vendors & P/N Installation Notes Fiber Optic SEA P/N 1270002-002 MM 62.5/125; 2 Fiber Conductors Industry Available from Multiple Vendors New Installation as an alternative to copper cable 2- Conductor Copper Wire SEA P/N 1686002-001 Belden P/N 9182 for Belden Armored Cable order SEA P/N 1680000-002 4-Conductor Copper Wire SEA P/N 1681000-003 Belden P/N SSD1743 Max distance from Analyzer to Hub or between Hubs 1520 meters (5000 ft) Recommended for Single Channel Max distance from Analyzer to Hub or between Hubs 1520 meters (5000 ft) Recommended for Dual Channel Max distance from Analyzer to Hub or between Hubs 1520 meters (5000 ft) Ethernet Cat 5 Unshielded Twisted pair (UTP) 2000585-001 SEA P/N 1681003-006 Industry Available from Multiple Vendors Recommended for short distances or within shelters. Max distance from analyzer to hub or between hubs 600 meters (2000 ft) Introduction • 9