4. INSTALLATION
CONNECTING AN
EXTERNAL
WATER FILLING
TANK
To avoid having to periodically fill the water tank (see Chapter 5, “First Start-Up”), it is
possible to connect the sterilizer to an external filling tank (supplied as an option), that the user
will periodically fill, or to a commercially-available, reverse-osmosis water purification system
with accumulation tank.
(OPTIONAL, automatic
filling function)
In that case, when the internal water tank reaches the MIN level, the autoclave activates a
pump that automatically fills the internal tank.
Follow the instructions below for the correct connection of the external tank:
– Install the tap provided on the filling tank; use Teflon tape or connector sealant for a perfect
seal.
Filling
tank
Clip
Teflon
Tap
Clip
Silicon pipe
–
–
–
–
–
–
Use the filling tanks silicone tube (or other suitable tube, max length 2 m) and insert it on
the filling connector taking care to push it completely on.
Lock the tube to connector with the plastic tie provided.
Insert the other end of the tube on the tap of the filling tank.
Make sure that the tube runs freely from the device to the filling tank, without being bent,
crushed or obstructed in any way.
Loosen the upper plug to facilitate the flow of water (also remove any gasket or underplug);
Open the tap on the filling tank.
NOTE
REFER TO THE CHAPTER, “CONFIGURING THE DEVICE – AUTOMATIC FILLING OPTION ”.
CONNECTING
DEMINERALIZER
The sterilizer can be connected to a demineralizer (water purifier) to assure the tank is
automatically filled continuously with high quality demineralized water.
(OPTIONAL, automatic
filling function)
Consult the relative User’s manual for instructions on how to install the demineralizer.
NOTE
FOR THIS OPTION SETTING,
AUTOMATIC FILLING OPTION”.
REFER TO CHAPTER
“CONFIGURING
THE
DEVICE –
For additional information and advice about the correct connection of the sterilizer to the
various water purification systems, contact M.O.COM. customer support (see Appendix Z).
NOTE
BACKFLOW FROM THE MACHINE TO THE WATER CIRCUIT MUST BE PREVENTED BY USING
TOOLS THAT ARE IN CONFORMITY WITH LAW IEC 61770.