Measurement Uncertainty Closed Channel Resistance Test For the performance verification tests in this manual, measurement uncertainties are calculated based on the Agilent 3458A Digital Multimeter. The measurement uncertainty shown in Table 2-1 is the accuracy of the Agilent 3458A using 90-day specifications. The calculations follow. Conditions: • 4-wire ohms function, 10 Ω range • 90-day specifications • Worst-case reading = 2.0 Ω M.U. = (15 ppm of Reading + 5 ppm of Range) = (15x10-6 * 2.0) + ( 5x10-6 * 10) Ω = 8.0x10-5 Ω DC Isolation Test Conditions: • 2-wire ohms function, 1 G Ω range • 90-day specifications • Worst-case reading = 1.2 G Ω (highest resistance that can be measured with the Agilent 3458A) M.U. = (0.5% of Reading + 10ppm of Range) = (0.005 * 1.2x109) + ( 10x10-6 * 1x109) Ω = 6.1x106 Ω Test Accuracy Ratio (TAR) 32 Verification Tests Test Accuracy Ratios are not defined for single-sided measurements, so all measurements have "NA" (Not Applicable) in the TAR column. Chapter 2