Download Agilent Technologies U1242A Technical data
Transcript
5 Performance Tests and Calibration Error from nominal 1 year Step Test Function Range 5520A Output U1241A 5 6 7 8 Turn the rotary switch to Turn the rotary switch to Turn the rotary switch to position Press function to go to position position μA μA 1000 W 1000.0 W ± 3.3 W 10 kW 10.000 kW ± 33 W [1] 100 kW 100.00 kW ± 330 W 1000 kW 1000.0 kW ± 3.3 kW 10 MW 10.000 MW ± 83 kW 100 MW 100.00 MW ± 1.53 MW [2] 1000 nF 1000.0 nF ± 12.4 nF 10 μF 10.000 μF ± 0.124 μF 100 μF 100.00 μF ± 1.24 μF 1000 μF 1000.0 μF ± 20.4 μF 10 mF 10.000 mF 0.204 mF 1000 μA 1000.0 μA ± 1.3 μA 10000 μA 10000 μA ± 13 μA 1000 μA 1000.0 μA, 500 Hz ± 10.5 μA 1000.0 μA, 1 kHz ± 15.5 μA 10000 μA, 500 Hz ± 105 μA 10000 μA, 1 kHz ± 155 μA 100 mA 100.0 mA ± 0.23 mA 440 mA 400.0 mA[3] ± 2.3 mA 100 mA 100.00 mA, 500 Hz ± 1.05 mA 100.00 mA, 1 kHz ± 1.55 mA 400 mA[3], 500 Hz ± 4.5 mA 400 mA[3], 1 kHz ± 6.5 mA 10000 μA 9 10 Turn the rotary switch to position Press function to go to mA mA 440 mA 44 U1242A [1] U1241A/U1242A User’s and Service Guide