7 :ACQuire Commands :ACQuire:COMPlete (see page 790) Command Syntax :ACQuire:COMPlete <complete> <complete> ::= 100; an integer in NR1 format The :ACQuire:COMPlete command affects the operation of the :DIGitize command. It specifies the minimum completion criteria for an acquisition. The parameter determines the percentage of the time buckets that must be "full" before an acquisition is considered complete. If :ACQuire:TYPE is NORMal, it needs only one sample per time bucket for that time bucket to be considered full. The only legal value for the :COMPlete command is 100. All time buckets must contain data for the acquisition to be considered complete. Query Syntax :ACQuire:COMPlete? The :ACQuire:COMPlete? query returns the completion criteria (100) for the currently selected mode. Return Format <completion_criteria><NL> <completion_criteria> ::= 100; an integer in NR1 format See Also • "Introduction to :ACQuire Commands" on page 183 • ":ACQuire:TYPE" on page 195 • ":DIGitize" on page 159 • ":WAVeform:POINts" on page 640 Example Code ' AQUIRE_COMPLETE - Specifies the minimum completion criteria for ' an acquisition. The parameter determines the percentage of time ' buckets needed to be "full" before an acquisition is considered ' to be complete. myScope.WriteString ":ACQuire:COMPlete 100" See complete example programs at: Chapter 38, “Programming Examples,” starting on page 799 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 185
Related documents
VDO MS 2000X - Technical data
Agilent Technologies InfiniiVision 3000 DSO-X 3054A Technical data
Agilent Technologies InfiniiVision 5000 Series Technical data
Agilent Technologies Ininiium 90000 Technical data
Agilent Technologies Ininiium 9000 Technical data
Agilent Technologies Sprinkler 54753a User's Manual
Agilent Technologies 6000 Series Sprinkler User Manual
Bedienung und Grundeinstellungen
Agilent VisualBasic VisualC#による オシロスコープ 制御入門
KEYENCE Vision Sensor MODEL: CV-X100