Download Agilent Technologies EXG User manual
Transcript
Notice WARNING The printer and its accessories (e.g. the infrared to parallel converter) may not meet the leakage current requirements specified in IEC 60601-1 safety standards. According to IEC 60601-1-1 safety standard, do not place the printer and its accessories within 1.5 m (4.5 ft.) of the patient. WARNING If you use the cardiograph during defibrillation, please check the electrode instructions first. The cardiograph recovery time after defibrillation specified in IEC 60601-2-25 is affected by the type of electrodes used. Use of conductive gels with the electrodes improves ECG recovery times after defibrillation. The ECG recovery time for defibrillation does not meet the requirements of IEC 60601-2-25 when using reusable electrodes. You can achieve the ECG recovery time required by IEC 60601-2-25 with some disposable electrodes by setting the cardiograph to 0.5Hz Baseline Wander filter setting. WARNING The AC/DC adapter supplied with this cardiograph is an integral part of the cardiograph’s safety features. Using any other AC/DC adapter may compromise the cardiograph’s safety as well as performance. CAUTION [US] Federal law restricts this device to sale by or on the order of a physician. WARNING Use of accessories other than those recommended by Agilent Technologies may compromise product performance. WARNING If non-Agilent recommended data transmission cable is used, the cardiograph may not meet the Radiated Emission Standard found in CISPR 11. viii Revision 0.1