Download Agilent Technologies U3402A Technical data
Transcript
Application Tutorial 3 Dual Display Operation Examples This section describes some practical operations when using dual display feature. Measure DC Voltage and AC Ripple on a Rectification Circuit A single measurement for both DC voltage and AC ripple can be displayed through both display while testing a rectifier circuit. 1 Connect the red and black test leads to the input terminal and probe the test points as shown in Figure 3- 1. AC/DC voltage source + – Figure 3-1 Terminal connection when measuring DC voltage and AC ripple on a rectification circuit 2 Press 3 Press 4 Press U3402A User’s and Service Guide to select DC voltage measurement for primary display. to enable the secondary display. The 2ND is displayed. to select AC voltage measurement for secondary display. 53