Download IMPORTANT !!!! Read this manual before attempting any
Transcript
IMPORTANT !!!! Read this manual before attempting any handling or installation of the meter. BME-M3000-09-07 Disclaimer The user/purchaser is expected to read and understand the information provided in this manual, follow any listed Safety Precautions and Instructions and keep this manual with the equipment for future reference. The information in this manual has been carefully checked and is believed to be entirely reliable and consistent with the product described. However, no responsibility is assumed for inaccuracies, nor does Badger Meter Europe assume any liability arising out of the application and use of the equipment described. Should the equipment be used in a manner not specified by Badger Meter Europe, the protection provided by the equipment may be impaired. Questions or Service Assistance If you have questions regarding the product or this document contact: Badger Meter Europa GmbH Nürtinger Str. 76 72639 Neuffen (Germany) Telephone: +49-7025-9208-0 Fax: +49-7025-9208-15 On the Web: www.badgermeter.de or call your local Badger Meter representative. Product Identification Information Record the product identification numbers from the nameplate. Modular Mag Meter Model Number _M3000______ Serial Number _____________ Tag Number _______________(if applicable) 2 Basic safety advices The electromagnetic flow meter is only suitable for the measurement of conductive fluids. The manufacturer is not liable for damages that result from improper or nor not in accordance with the requirements use. The meters are constructed according to state-of-the-art technology and tested operationally reliable. They have left the factory in a faultless condition concerning safety regulations. The mounting, electric installation, taking into operation and maintenance of the meter may only be carried out by suitable technicians. Furthermore the operating personnel has to be trained by the operating authority and the instructions of this manual have to be followed. Principally the regulations for the opening and repairing of electrical equipment in your country have to be payed attention. Repairs Should you send back a flow meter in operation, please take notice of the following points: - Please enclose a description of the fault as well as a precise statement of the measured medium (if necessary a safety specification sheet). - The meter has to be in a cleaned condition (outside and inside). Especially with harmful measuring mediums you have to pay attention that there are no impurities in the pipe or at the connections. - If it is not possible to clean the meter completely, particularly with harmful materials, do not send back the meter. We reserve the right to repair only cleaned meters. Costs, which result from insufficient cleaning, will be charged to you. CE mark The electromagnetic flow meter complies with the European regulation ATEX 94/9/EC, EMC regulation 2004/108/EC and PED regulation 97/23/EC. The device is so designed and manufactured that it cause no danger by direct or indirect contact. Badger Meter Europe confirm the conformity with the European regulations by printing the CE sign on the meter. 3 TABLE OF CONTENTS Disclaimer Questions or Service Assistance Product Identification Information Basic saftey and repair advise CE mark 2 2 2 3 3 SAFETY PRECAUTIONS and INSTRUCTIONS Equipment Unpacking, Inspection, Moving and Return Policies Rigging, Lifting, Moving Large Units Instructions Specific to Hazardous Area Installations 6 6 7 METER DESCRIPTION Empty Pipe Detection Amplifier Mounting Options Meter Mount Remote Mount 9 9 9 9 9 METER/AMPLIFIER LOCATION, ORIENTATION and APPLICATIONS Remote Amplifier Outdoor Location Temperature Pipelines and Fluid Flow Meter Orientation Vertical Placement . Horizontal Placement Straight Pipe Requirements .. Pipe Reducer Requirements . Chemical Injection Applications Partially Filled Pipe Situations . Meter Gaskets and Grounding 9 9 9 10 10 10 10 10 10 11 11 12 METER INSTALLATION Remote Mount Amplifier Remote Mount Amplifer/Detector Wiring Detector Junction Box Remote Mount Wiring Diagram Electrode and Coil Wiring from Detector Junction Box to Remote Mount Amplifier Junction Box Electrode Wiring Coil Wiring Output Wiring External Disconnect AC Power Wiring Adjustable Display/Control Card . 12 13 13 13 14 15 15 15 17 18 18 19 AMPLIFIER CONTROLS, CONTROL METHODS AND DATA DISPLAY PROGRAMMING Card Display and Controls Two Programming Methods Main Screen(s) Password Amplifier Access Switches/Buttons How to Program What to Program 20 20 20 20 21 22 22 26 PROGRAMMING REQUIRED PARAMETERS Input a Password Set Flow Units Set Full Scale Flow Set Flow Direction Set Damping Factor Empty Pipe Calibration . Pulse Output Analog Output 26 26 27 27 28 28 28 29 31 4 ADDENDUM A Detector Specifications Amplifier Specifications 34 34 35 ADDENDUM B Maintenance Flow Tube and Electrode Cleaning Fuse Replacement Amplifier PCB Stack Replacement 36 36 36 36 37 ADDENDUM C Menu Structure Flow Chart 40 ADDENDUM D ATEX Certificate 42 42 5 SAFETY PRECAUTIONS and INSTRUCTIONS NOTE: Do not remove the liner protector until installation. Safety considerations are emphasized by the placement of safety symbol icons on the product or next to important text, pictures or drawings throughout this manual. The symbols are: 8. Storage: If the meter is to be stored, place it in its original container in a dry, sheltered location. Storage temperature ranges are: -4°F to 158°F (-20°C to 70°C). Rigging, Lifting, Moving Large Units When and where this symbol is attached to the product it indicates a potential hazard. It means that documentation must be consulted to determine the nature of the potential hazard and any actions that need to be taken. DO NOT lift or move a meter via its amplifier, junction box or cables. Lift and move meters with flow tubes between 2" and 8" (50 and 200mm) with a crane rigged with soft straps. Place a strap around the detector body, between the flanges, on each side of the detector. Warning indicates an action or procedure, which, if not performed correctly, can result in sever personal injury, death, or substantial property damage. Comply with the instruction and proceed with care. Caution indicates an action or procedure, which if not performed correctly, will or can cause minor personal injury or property damage. Comply with the instruction and proceed with care. Equipment Unpacking, Inspection, Moving and Return Policies 1. If shipping container damage is evident at delivery, have a responsible person present when the meter is unpacked. 2. Inspect the shipping container for unpacking, lifting or moving instructions. 3. As the unit is opened and unpacked, verify the shipment packing list and your order form match the items in the shipment. 4. Open the container and remove all cushioning materials. Keep the shipping container and packing materials should the meter need to be returned or put in storage. 5. Remove the meter from the container. For meter flow tube diameters between 2” and 8” (50 and 200mm) use a crane or lifting device with soft straps placed around the detector body, at the flanges. For meter flow tube diameter sizes 10” (250mm) and larger, lift the unit via the lifting lugs with a crane, cables and hooks. 6. Inspect the meter for signs of shipping damage; scratches, loose or broken parts. Lift meter flow tube diameter sizes 10" (250mm) and larger via its lifting lugs. Use the proper size crane, cables and hooks. DO NOT lift the meter by the amplifier or, on remote style meter, by the junction box. NOTE: If damage is found, a formal claim for damage in transit is the responsibility of the customer. Within 48 hours of delivery request an inspection report from the carrier. File a claim with the carrier. Contact the Badger Meter factory to facilitate repair or replacement, +49 7025 92080. 7. All detectors with PTFE liners are shipped with a liner protector to maintain proper form of the PTFE material during shipping and storage. 6 Instructions Specific to Hazardous Area Installations These instructions apply to equipment covered by ATEX Certificate Number FM08ATEX0051X issued by FM. Use the sling-rigged method to lift large detectors into a vertical position while still crated. Use this method to position large detectors vertically into pipelines. 1. During any installation or repair, perform all procedures in accordance with the applicable code of practice. 2. Suitably trained personnel shall perform all installation or repair procedures. 3. If the equipment is likely to come into contact with aggressive substances, it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected, thus ensuring that the type of protection is not compromised. Aggressive Substances – e.g. acidic liquids or gases that may attack metals or solvents that may affect polymeric materials. Suitable Precautions – e.g. regular checks as part of routine inspections or establishing, from the material’s data sheet, that it is resistant to specific chemicals. Do not lift a detector with a forklift via the detector body between the flanges. The housing could be dented and/or damage caused to internal coil assemblies. NEVER place forklift forks or rigging chains, straps, slings, hooks or other objects inside or through the detector flow tube for lifting or handling purposes. The isolating liner could be damaged, rendering the unit inoperable. 7 Additional Information Certification markings are noted on the product label. For additional information regarding importation, equipment installation, equipment repair, equipment return or renewal parts, please contact Badger Meter Europe or your local Badger Meter representative. Meter mounted 85-240 VAC Meter mounted 24 VDC Remote mounted 85-240 VAC - Amplifier Remote mounted 24 VDC - Amplifier Remote mounted 85-240 VAC - Detector Remote mounted 24 VDC - Detector 8 METER DESCRIPTION Badger Model M3000 electromagnetic meters are ATEX approved for Zone 2 hazardous locations. To achieve hazardous location ratings electrodes in the flow tube are Intrinsically Safe, designed according to ATEX standards. Those standards limit the amount of energy that can be sent to electrodes to prevent a spark from occurring. This configuration can also provide a more convenient amplifier programming and display placement for monitoring meter readings. Empty Pipe Detection Badger Magnetoflow Mag meters are equipped with an Empty Pipe Detection feature. Empty Pipe Detection is accomplished by positioning a third electrode between 12 o’clock and 1 o’clock. Any time this electrode is not covered by fluid, for a minimum of five seconds, the meter displays an Empty Pipe Detection condition, sends out an error message if desired, and stops measuring to maintain accuracy. When the electrode is again covered with fluid, the error message disappears and the meter continues measuring. Amplifier Mounting Configuration Options Two amplifier-mounting configuration options are available to meet a variety of meter placement and environmental conditions. Meter Mount Configuration The Meter Mount configuration has the amplifier mounted directly on the detector. This compact, self-contained configuration minimizes installation wiring. METER/AMPLIFIER LOCATION, ORIENTATION and APPLICATIONS Remote Amplifier Outdoor Location The amplifier can be installed and operated outdoors. However, protection from the elements must be considered: 1. Be aware of the ambient environment and temperature ratings for the unit -4°F to 122°F (-20°C to 50°C). 2. If an indoor location is within 100 feet (30m) of the detector installation, consider increasing the cable length and mounting the amplifier indoors. 3. Consider mounting the amplifier in a IP 67 outdoor enclosure. 4. At a minimum, fabricate a roof or shield over and/or around the amplifier to protect the LCD display screen from sunlight and to keep the possibility of rain or moisture problems to a minimum. Temperature To prevent meter damage in any environment, minimum and maximum temperature ranges must be observed. 1. For remote amplifier applications, the fluid temperature range is -4°F to 221°F (-20°C to 105°C) at a maximum ambient temperature of 122°F (50°C) for the following liner materials: PTFE. Meter Mount 2. For remote amplifier applications, the fluid temperature range is 32°F to 178°F (0°C to 80°C) at a maximum ambient temperature of 122°F (50°C) for the following liner materials: Hard rubber. Remote Mount Configuration Remote Mount configuration places the amplifier and its functions at a location separate from the fluid flow and detector. This configuration is necessary in situations where process fluid temperature or environment exceeds amplifier ratings. A remote mounting bracket is supplied. 3. For meter mounted amplifier applications, the fluid temperature range is -4°F to 212° (-20°C to 100°C) at a maximum ambient temperature of 122°F (50°C) for the following liner materials: PTFE. Detector and amplifier are connected by wires, run through conduit, between junction boxes on the detector and remote mounted amplifier. The distance between the detector junction box and amplifier junction box can be up to 100 feet (30m). 9 4. For meter mounted amplifier applications, the fluid temperature range is 32°F to 178°F (0°C to 80°C) at a maximum ambient temperature of 122°F (50°C) for the following liner materials: Hard rubber. Horizontal Placement In a horizontal piping orientation, mount the detector to piping with the flow measuring electrode axis in a horizontal plane (3 and 9 o’clock). 5. The ambient temperature range surrounding the amplifier is -4°F to 122°F (-20°C to 50°C.) This arrangement prevents solids build-up or sediment deposit or accumulation on the electrodes. 6. The ambient temperature range surrounding a remote junction box mounted to the detector is -4°F to 122°F (-20°C to 50°C.) Pipelines and Fluid Flow Although the Model M3000 is rugged, there are some pipeline and fluid flow conditions that should be avoided: 1. Do not install the meter where extreme pipe vibrations exist. If vibrations are present, secure piping before and after the meter with appropriate pipe supports. If vibrations can’t be restrained, consider mounting the amplifier remotely. 2. Avoid installing the detector close to pipeline valves, fittings or impediments that can cause flow disturbances. 3. For detectors with PTFE liners, avoid installing the detector on suction sides of pumps. 4. Avoid installing the detector on outlet sides of piston or diaphragm pumps. Pulsating flow can affect meter performance. 5. Avoid locations near equipment producing electrical interference such as electric motors, transformers, variable frequency, power cables, etc. 6. Verify both ends of the signal cables are securely fastened. 7. Place power and signal cables in separate conduit. 8. Place the meter where there is enough access for installation/maintenance purposes. Straight Pipe Requirements Sufficient straight pipe runs are required at the detector inlet and outlet for optimum meter accuracy and performance. An equivalent of three (3) diameters of straight pipe is required on the inlet (upstream) side. Two (2) diameters are required on the outlet (downstream) side. Pipe Reducer Requirements With pipe reducers a smaller size meter can be mounted in larger pipelines. This arrangement may increase low flow accuracy. Meter Orientation Mag Meters can operate accurately in any pipeline orientation and can measure volumetric flow in forward and reverse directions. There are no special requirements for standard, concentric, pipe reducers. Custom fabricated pipe reducers must have a minimum slope angle of 15 degrees to minimize flow disturbances and excessive loss of head. If this is not possible, install the custom pipe reducers as if they were fittings and install the amount of straight pipe stated previously. NOTE: A Forward Flow direction arrow is printed on the detector label. Vertical Placement Mag meters attain optimal performance when placed vertically, with liquid flowing upward and meter electrodes in a closed, full pipe. Vertical placement allows the pipe to remain completely full, even in low flow, low pressure applications and it prevents any solids build-up or sediment deposit or accumulation on the liner and/or electrodes. NOTE: Carefully observe the “Forward Flow” label on the meter body and install the meter accordingly. 10 Chemical Injection Applications For water line applications with a chemical injection point, install the meter upstream of the injection point. This eliminates any meter performance issues. If a meter must be installed downstream of a chemical injection connection, the distance between the meter and the injection point must be significant; 50 to 100 feet (15 to 30 meters). When the water/chemical solution reaches the meter it must be a complete, homogeneous mixture. If the . injection point is too close, the meter senses two (2) different liquids (conductivity is different for each) and correct data output cannot be assured. The injection method: spaced bursts, continuous stream of drips, a liquid or gas can also affect downstream readings by the meter. Do not install the meter in a vertical, downward flow section of pipe Sometimes it’s difficult to specify the exact downstream placement distances because of the number of variables. Contact Badger Meter Technical Support, +49 7025 92080, to review your application if necessary. Partially Filled Pipe Situations It is possible to encounter situations where the process pipe is momentarily only partially filled. Examples include; lack of backpressure, insufficient line pressure, gravity flow applications, etc. ALWAYS locate ON/OFF valves on the downstream side of the meter To eliminate these situations: Do not install the meter in the highest point of the pipeline To minimize the possibility of partially full pipe flows in horizontal, gravity or low pressure applications, create a pipe arrangement that insures the detector remains full of liquid at all times. 11 12 For Remote Mount Units consider: 1. Amplifier location. 2. Remote amplifier mounting bracket. 3. Proper conduit and conduit fittings. 4. Wiring and conduit locations. REMOTE MOUNT AMPLIFIER NOTE: Screws are supplied to attach bracket to amplifier. Screws are not supplied to attach bracket at mounting location. NOTE: Not relevant to meter mount units. Remote Mount Amplifier Location Requirements: 1. A sturdy and safe mounting surface capable of holding the amplifier weight (20 pounds, 9 Kg). 2. Within the allowable temperature range: -4°F to 122°F, (-20°C to 50°C). 3. Access to amplifier covers, ports, terminals, screen and adjustments. 4. As close to the detector as possible. 5. Determine length and route of cable/conduit runs. (Refer to the Remote Mount Wiring Diagram on page 15.) Detector Junction Box The Detector Junction Box has one chamber and two wire ports. The Junction Box, Chamber and wiring ports provide openings for wire, conduit, tool and hands access to terminal blocks. Detector to remote mount amplifier electrode and coil wires connect to the detector through the chamber wire ports. Use the supplied remote mount amplifier bracket to secure the remote mount amplifier at a desired location. An amplifier unit weighs approximately 20 pounds (9 Kg). Mount Bracket to Amplifier 1. Align bracket-mounting holes with amplifier mounting holes. 2. Attach bracket to amplifier with supplied screws. Torque screws to 80 in. lb. Mount Bracket/Amplifier to Location 1. Position the bracket/amplifier in the desired orientation. 2. Secure bracket/amplifier to location. · Suitably trained personnel shall perform all installation or repair procedures. · Disconnect power to the unit before attempting any installation or maintenance. · Do not bundle or route signal wires with power wires. · Use proper conduit, connections and supplied cables in all wiring procedures. · Observe all local applicable electrical codes when wiring any equipment. REMOTE MOUNT AMPLIFIER/DETECTOR WIRING Remote Mount Amplifier The remote mount amplifier has three chambers and five wire ports. The Junction Box and Connections Chambers and wiring ports provide amplifier openings for wire, conduit, tool and hands access to amplifier terminal blocks. Detector to amplifier wires connect in the Junction Box Chamber. The Display/Programming Chamber provides access to fuses and circuit boards. They are discussed later. 13 14 Electrode Wiring in Amplifier Junction Box To connect the electrode wires in the Amplifier Junction Box: 1. Unscrew the amplifier junction box chamber cover. If necessary, use a strap wrench. 2. Remove the protective plastic cover to access the terminal block screws. Electrode and Coil Wiring From Detector Junction Box to Remote Mount Amplifier Junction Box A remote mount unit requires electrode and coil cables, from the detector junction box to the amplifier junction box, be enclosed in properly rated conduit. Use conduit fittings (not supplied) that are rated for Zone 2 hazardous locations. NOTE: Plastic cover must be reattached when wiring is complete to maintain hazardous location rating. Failure to use proper conduit fittings rated for Zone 2 hazardous locations, invalidates the ATEX rating and any warranties, expressed or implied, for this equipment. 3. Strip the cable jacket back 2 inches (50mm). 4. Strip the 4 wires back ¼ inch (6mm). 5. Thread wires through the proper cable access. Connect the wires to the compression style screw terminals of the Amplifier Junction Box. 1. Lay out the cable and conduit between the Detector Junction Box and the Amplifier Junction Box. 2. Run cables through the conduit, between Detector Junction Box and Amplifier Junction Box. 3. Place four IP 67, M20 fittings on conduit. 4. Remove the four junction box wire port screws, two on each junction box. Electrode Wiring in Detector Junction Box To connect electrode wires in the Detector Junction Box: 1. Unscrew the Detector chamber cover. If necessary, use a strap wrench. 2. Remove the protective plastic cover to access the terminal block screws. NOTE: Plastic cover must be reattached to maintain hazardous location rating. 3. Strip the cable jacket back 2 inches (50mm). 4. Strip the 4 wires back ¼ inch (6mm). 5. Thread wires through the proper cable access. Connect the wires to the compression style screw terminals of the Detector Junction Box. Red to terminal labeled -E1 Green to terminal labeled -SHLD Black to terminal labeled -E2 White to terminal labeled -EP Coil Wiring in Detector Chamber To connect coil wires in the Detector chamber: 1. Lay out the cable and conduit between the Detector Junction Box and the Amplifier Junction Box. Use Belden #8770 cable or equivalent for coils. NOTE: Plastic cover must be reattached to maintain hazardous location rating. 2. Strip the cable jacket back 2 inches (50mm). 3. Strip the 2 wires back ¼ inch (6mm). 4. Thread wires through the proper cable access. Connect the wires to the compression style screw terminals of the detector chamber. Red to terminal labeled Green to terminal labeled Black to terminal labeled White to terminal labeled E1 SHLD E2 EP Cable length, between Junction Boxes, may be up to 100 feet (30M). 6. Run cable and conduit to Amplifier junction box. 15 OUTPUT WIRING The Badger M3000 Meter converts liquid flow into electrical signal(s). With proper output wiring and amplifier programming, the signal(s) are sent to, and used by, processing equipment used in operations or other procedures. NOTE: Output wires and terminals are the same for meter mount or remote mount meters Output wiring requires 18 to 22 AWG maximum, shielded wire (not supplied). Signal wire insulation temperature class should exceed the maximum temperature where installed (typical, 185°F, 85°C). Use conduit and conduit fittings (not supplied) rated for Zone 2 hazardous locations. Output Wire Connections Red to terminal labeled – C1 Clear to terminal labeled – C2 · Properly trained personnel must perform all installation and/or repair procedures. · Disconnect power to the unit before attempting any installation or maintenance. 5. Connect conduit to junction box. Use a IP 67, M20 fitting. 6. Install protective plastic cover over terminal blocks. 7. Attach detector chamber cover. To connect control signal wires: 1. Remove the connections chamber cover. If necessary, use a strap wrench. 2. Remove the two terminal block wire port access screws. 3. Connect output wires to processing equipment. 4. Group and place output wires in conduit. Position conduit at amplifier terminal block wire ports. 5. Connect conduit to control output signal wires ports. 6. Run output wires through wire ports, into amplifier terminal chamber. 7. Strip output wires back ¼ inch (6mm). 8. Connect output wires to terminals (see below). Cable length between Junction Boxes may be up to 100 feet (30M). Coil Wiring in Amplifier Junction Box To connect the coil wires in the amplifier junction box: 1. Strip the cable jacket back 2 inches (50mm). 2. Strip the 2 wires back ¼ inch (6mm). 3. Connect the wires to the compression style screw terminals of the amplifier junction box. NOTE: Use twisted pair shielded wire for all output wiring. Belden #1266A or equivalent. Red to terminal labeled – C1 Clear to terminal labeled – C2 4. Connect conduit to junction box. Use a M20 fitting. 5. Install protective plastic cover over terminal blocks. 6. Attach the amplifier junction box chamber cover. 16 Amplifier Output Wire Terminal Block Connections: Reference Control Signal Wiring Diagrams on next two pages. Analog Input Wiring Diagram Function: 4-20 mA, 0-20mA, 0-10mA Analog Output (Loop voltage = 18VDC Sourced) mA Analog Output (+) 5 Common 6 Output 1 - Transistor Output (open collector) Functions: Forward Pulse, AMR Pulse, Flow Set Point, Error Alarm, Empty Pipe, Flow Direction Output 2 - Transistor Output (open collector) Functions: Reverse Pulse, Frequency Output Preset Output, Flow Set Point, Error Alarm, Flow Direction Terminal 1 Input 1 (+) Function: reset, positive zero return. Terminal 2 Output 1 (+) Programmable Transistor Output (open collector) - Passive Output to Badger external counter. Functions: forward pulse, AMR pulse, flow set point, error alarm, empty pipe, flow direction. Active Output to external counter. Terminal 3 Output 2 (+) Programmable Transistor Output (open collector) - Passive Output to Badger external counter. Functions: reverse pulse, frequency output, preset output, flow set point, error alarm, flow direction Active Output to external counter. Terminal 4 Common field ground. Terminal 5 mA Analog Output (+) Source. Terminal 6 Common field ground. 17 External Disconnect Position this device in an accessible location. Disconnect main power to the unit before attempting any device maintenance. Position and identify the disconnect device so as to provide safe and easy operation. To re-position or rotate the display/control card in the amplifier: Label the disconnect device as being for the Mag Meter. 1. Remove display chamber cover. Turn the cover counterclockwise to remove it from the amplifier. If necessary, use a strap wrench. Install an external disconnect switch or circuit breaker that meets local standards. AC Power Wiring For AC power use three wire, sheathed, cable with cable diameter of 18 AWG (not supplied). AC wire insulation temperature class must not exceed the maximum ambient temperature of its location (typical, 185°F, 85°C). Use conduit and conduit fittings (not supplied) that are rated for Zone 2 hazardous locations. To maintain a IP66 rating, use watertight fittings that are rated IP66 or better. 2. Remove the 2 card screws and washers. (NOTE: Use a split screwdriver to prevent dropping screw into To prevent accidents connect main power only after all other wiring has been completed. The amplifier is a microprocessor device. It is important that the power supply be as “clean” as possible. Avoid using power lines that feed heavy loads; pumps, motors, etc. If dedicated lines are not available, a filtering or isolation system may be required. 3. Tilt card up/out approximately 45 degrees at the holding clips. AC wiring is the same for meter mount and remote mount amplifiers. 1. Remove AC Wires Port screw from amplifier connections chamber. 2. Lay out AC cable and conduit to amplifier. 3. Place cable in conduit. 4. Strip AC cable back 2 inches (50mm). 5. Strip AC wires back ¼ inch (6mm). 6. Attach AC wires to amplifier terminal. Black to L White to N Green to G 4. Gently pull card down and out from between holding clips. 7. Connect conduit to amplifier. 8. Attach chamber cover. 5. Rotate card to appropriate position. Adjustable Display/Control Card Because meter positioning sometimes places the amplifier display/programming chamber in an awkward position, the display/control card is adjustable in 90-degree increments. 6. Angle card and position card holes between the holding clips. 18 AMPLIFIER CONTROLS, CONTROL METHODS and DATA DISPLAY PROGRAMMING The M3000 amplifier display/programming chamber contains a display/control card. This card and its display screen provide easy access to meter information and the ability to view, program and adjust meter data parameters. 7. Push card in, between holding clips. Lower card back into position and attach card with screws and washers. Card Display and Controls The M3000 uses a 2.5 X 1 inch (63 X 25mm) four line, 16character, backlit, LCD display. 8. Attach the chamber cover. Display screens and screen data are manipulated with magnet switches or push buttons. By operating the switches or buttons different screens are accessed, program parameters are selected and changed and settings are reprogrammed. If no contact is made with switches or buttons for 2 minutes, in any parameter, the display returns automatically to the Main Screen. 19 Two Main Screen Format settings are possible. Which setting to use is determined by the fluid Flow Direction through your meter. The settings are Uni-Directional flow or Bi-Directional flow (reference page 27/28 to set flow direction). ACCESS to AMPIFIER SCREENS, DATA and PARAMETER PROGRAMMING NOTE: The M3000 can be programmed to meter many flow situations and serve a variety of purposes during a production process. To meet diverse needs there are a wide variety of programming options and parameters. Your metering requirements probably do not require the use of all program screens, options and parameters. Both main screens display the rate of flow (R=) and flow units (reference page 26 to set flow units). Your M3000 amplifier is delivered preprogrammed, based on information available at the time the unit was ordered. In most instances it will not require any changes. Uni-Directional Flow Main Screen Uni-directional Flow Totalizes Pipe Flow in only one direction, the flow direction on the detector label. (reference page 27/28 to set uni-directional flow) Use this section to program flow signal outputs or reprogram your meter to specific requirements. Uni-directional readings, on the display screen, are identified as R=, T1, T2 and PS. Flow measurement and totalizing continues during amplifier programming. R= Flow Rate Two Programming Methods There are two methods for programming. In one method a magnet wand is used to manipulate +, - and E switches to view and change screens and program functions and settings. The amplifier display chamber cover stays on. T1 registers Forward Volume T2 registers Forward Volume and can be reset through Input 1. PS registers Preset Batch Amount The other method involves removing the display chamber cover and manipulating the +, -and E push buttons to view and change settings and program desired meter functions and settings. With this information an operator can tell at a glance the volume going through the meter. Bi-Directional Flow Main Screen Bi-Directional Flow totalizes pipe flow in both directions (reference page 27/28 to set bi-directional flow) Bi-directional totalizers readings, on the display screen, are identified as R=, T+, T- , TN and PS. AMPLIFIER SCREENS, MENUS and PROGRAMMING Amplifier programming requires planning, moving and activating the +, -and E selections on three types of menu/ parameter screens, List screen, Input Numbers screen and select ON or Off screen. R= T+ TTN MAIN SCREEN(S) The first screen is the Main Screen. It is always displayed when programming functions are not occurring. NOTE: When programming, if no contact is made with switches or buttons for 2 minutes, in any parameter, the display returns automatically to your programmed Main Screen. 20 Flow Rate registers Forward Volume registers Reverse Volume Net Total = (T+) - (T-) Amplifier Access Switches/Buttons Accessing amplifier menus, movement to and between screen, parameter selections and settings are all accomplished by pressing the +, - and E magnet operated switches or push buttons in conjunction with an on-screen location arrow. PS registers Present Batch Amount With this information an operator can tell at a glance the volume going through the meter in both directions. Switches Identifications With your Main Screen showing on the display, switch or press E. Screen PASSWORD??? 01 opens. Buttons PASSWORD??? O1 PASSWORD??? 01 is a typical Input Numbers screen. If a password had been entered into your program, this screen opens (reference page 24 to input a password) Switch/Buttons (+ and -) An underscore ( _ ) is positioned under the first 0. 1. Switch or press + to increase the number by one digit. Switch or press – to decrease the number by one digit. Select/Press + to: • move text up by one line, relative to the on screen arrow, for each select/press when choosing menu or sub-menu list names • move text up by one line, relative to the arrow, for each select/press when selecting a parameter from a list • increase a number by one digit for each select/press when inputting parameter numeric settings 2. After the correct number is entered for that digit, press E to move the underscore to the next 0. 3. Repeat the number selection process for this and all the 0’s. 4. After the last number is entered, press E. 5. The MAIN MENU appears. Select/Press – to: • move text down by one line, relative to the on screen arrow, for each select/press when choosing menu or sub-menu list names • move text down by one line, relative to the arrow, for each select/press when selecting a parameter setting from a list • decrease a number by one digit for each select/press when inputting parameter numeric settings However, if the wrong password was entered: Screen INVALID PSWD 02 opens. 1. Select +, -or E to return to the Main Screen. E Switch/Button 2. Press E again. 3. The screen returns to PASSWORD??? 01. Enter the correct password as described above. All passwords are factory set to 0000 (no password is programmed or required). If 0000 is the password, pressing E from the Main Screen opens the MAIN MENU screen. 21 Select/Press E to: • enter (open) a menu or sub-menu topic at which the arrow is pointing • select between ON/OFF parameter settings for each select/press • Move the parameter numeric setting underscore ( _ ) one place to the right for each select/press when inputting parameter numeric settings • SAVE a parameter numeric setting. After all numbers are input, press E and the setting is saved, the screen closes and the previous screen with the arrow pointing at “Exit this Menu” shows again. Select E again. That screen closes and the previous screen appears with the arrow pointing at its “Exit this Menu” • Continue selecting E to return to the Main Screen. • If your amplifier is not password protected, press E from the Main Screen to access the MAIN MENU. The MAIN MENU is a list that provides access to all amplifier menus and parameters • If your amplifier is password protected and E is pressed from the Main Screen, the PASSWORD screen opens (reference PASSWORD??? 01, on page 26) How to Program Amplifier screens, menus and program parameters are arranged in the familiar “branching” format where the MAIN MENU screen is a list of selections. The Main Menu list provides access to other lists, selections or parameter screens which provide input for meter programming. The amplifier program uses three types of screens; a List screen, an Input Numbers screen and a select ON or Off screen. MAIN MENU 00 >Exit this Menu Meter Setup Measurements Inputs/Outputs Clear Totals Communications Info/Help Logout Language Select Sample List Screen Amplifier Access Switches/Buttons Accessing amplifier menus, movement to and between screens, parameter selections and settings are all available using +, - and E buttons. · Sample Input Numbers Screen Move the parameter numeric setting underscore ( _ )one place to the right for each select/press when inputting parameter numeric settings. · SAVE a parameter numeric setting. After all numbers are input, press E and the setting is saved, the screen closes and the previous screen with the arrow pointing at "Exit this Menu" shows again. Select E again. That screen closes and Sample Select ON or OFF Screen the previous screen appears with the arrow pointing at its “Exit this Menu”. · · The following pages introduce you to the screen formats, describe how to maneuver to and through them and provide some specifics about programming terminology and parameters. Continue selecting E to return to the Main Screen. If your amplifier is not password protected, pressing E from the Main Screen accesses the MAIN MENU. The MAIN MENU is a list that If possible, have access to your amplifier display and controls and perform these screen manipulations. provides access to all amplifier menus and · parameters. If your amplifier is password protected and E is pressed from the Main Screen, the PASSWORD screen opens (reference PASSWORD??? 01, on page 26) 22 3. Next, press E. MAIN MENU The MAIN MENU contains a list of eight selections and an “Exit this Menu” choice. All amplifier programming and parameters are accessed from this list MAIN MENU 00 >Exit this Menu Meter Setup Measurements Inputs/Outputs Clear Totals Communications Info/Help Logout Language Select 4. Screen METER SETUP 10 opens METER SETUP 10 is another List screen. It provides access to common meter parameters. 1. Maneuver the + or – magnetic switches or push buttons to position the arrow at a selection. Only four lines of text are visible on the display screen. Operating the + and – switches/buttons moves text up or down and into view. When an item from the list on the MAIN MENU is selected, (press E when the item is in line with the arrow) either a screen opens that requires an action such as select a size or numeric unit for a parameter, turn a parameter on or off, etc. or a screen with another list may open requiring another selection be made to access and set a specific parameter size, numeric unit or on/off configuration. 2. For example, place the arrow at Pipe Dia. by pressing or activating + button once. Each screen has a name and number that displays at the top of the screen. Write down screen names, numbers and parameters that you access and change should other changes be needed later. 3. Press E to open the Pipe Dia. screen. NOTE: Your metering requirements may not require the use of all screens, options and parameters. MAIN MENU 00 is a typical List screen. It provides access to parameter screens. The MAIN MENU is a list of eight selections and an “Exit this Menu” choice. All amplifier programming and parameters are accessed from this list. 1. Press + or – to position the arrow at a selection. 4. Screen Pipe Dia. 11 opens. 2. For example, press + once to place the arrow at Meter Setup. 23 PIPE DIA. 11 is another List screen. It requires a parameter selection. 8. The screen automatically returns to METER SETUP 10 with the arrow pointed at Exit this Menu. PIPE DIA. 11 is a list of pipe diameter sizes. Transitions 1. Press + or – to position the arrow at another selection or E to return to the Main Menu. 2. Use the E key to return to screen Main Menu 00. 5. Press + or – to position the arrow at the appropriate pipe size. 3. Next, press + three times to position arrow at Inputs/ Outputs. 6. Press E. 4. Press E. 7. A screen, with the statement [(xx mm [xx”] ] **SELECTED** appears for about 2 seconds. 5. Press + three times to position arrow at Digital Out. #1 on screen In/Outputs 30. It verifies that the selected pipe diameter size parameter was entered into the amplifier settings. 24 2. Press the + once to place the cursor at Meter Setup. 6. Press E. 3. Press E to bring up Meter Setup 10 screen. 4. Press + five times to position the cursor at Empty Pipe. 7. Press + once to move the arrow across from Pulses / unit. 5. Select or Press E. 8. Press or activate E. 6. Screen Empty Pipe 15 opens. Empty Pipe 15 is a typical select ON/OFF screen. Pulses / Unit 3D is a typical Input Numbers screen. 1. When the screen opens, the arrow is pointing at a number. Empty Pipe is usually OFF. To turn Empty Pipe, On: 1. Press + once to move the cursor down one place, to Cal. empty pipe. Press E. 2. Use + or – selections to increase or decrease the underscored number to the desired digit. 3. Select E to move the underscore to the next digit and repeat the number selection process. 4. To skip a digit press E. The underscore moves to the next digit. 5. When all digits are set, press E. Pulses / Unit 3D is programmed into the system and the screen Dig Output 1 33 returns. 2. In screen EMPTY CALIB. 1E, press + twice. 3. Press E to turn Cal=[ON]. 4. Press + to move cursor to Exit WITH save. Empty Pipe is a “branch” list screen from the MAIN MENU. From here calibrate ON or OFF is made. 5. Press E to lock the selection into the program. 1. Press E to activate the Main Menu 00 screen. Continue pressing E to reverse through the screens to your next programming selection or to return to the Main Screen. You have now used the +, -and E selections to access the three types programming screens, List, Input Numbers and Select ON or Off and maneuver through some meter programming. 25 6. If the wrong number is input, INVALID PSWD comes up. Press +, -or E once. What to Program There are a variety of screens available. Go to 'Addendum C' for flow chart reference. That section describes how to determine and record what parameters need to be programmed. PROGRAMMING REQUIRED PARAMETERS All meters have required parameters that must be programmed. They include a Password, the Main Screen for Uni-or Bi-Directional Flow, Empty Pipe, Pulse Output and Analog Output, among others. 7. Main Screen appears. Press E again. This section presents keystroke details describing how to program required parameters. Input a Password 1. With Main Screen up, press E, to open MAIN MENU 00. 8. PASSWORD??? 01 screen reactivates. Input the 7. From TotalizerUnit 22 press + or -to select a totalizer correct password. unit. Press E. 2. With MAIN MENU 00 up, press + once to move cursor to Meter Setup. Press E. Set Flow Units 1. With Main Screen up, press E. 3. On METER SETUP 10 screen, press + six times or once to go to Chg. Password. 2. With MAIN MENU 00 active, press + twice (Measurements) and then E. 4. At CHG. PASSWORD 16 screen, input a password number. 3. On the MEASUREMENTS 20 screen press + once to place cursor at Flow Unit. Press E. 5. Or this screen comes up. Input the correct password. 26 2. At FullScaleFlow 23 input the proper flow. Select E to move cursor as described earlier. 4. From FLOW UNITS 21 press + or -to select a flow unit. Press E. 5. This type of screen shows for two seconds. 3. Screen returns to MEASUREMENTS 20. Press + four times, to place cursor at LowFlowCutOff. Press E. 6. The screen returns to MEASUREMENTS 20. With the cursor at Exit this Menu, press + twice to move the cursor to Totalizer Unit. Press E. 4. At LowFlowCutOff 24 screen input a percentage at the arrow. Press E. 7. From TotalizerUnit 22 press + or – to select a totalizer unit. Press E. 5. This type of screen shows for two seconds. 8. This type of screen shows for two seconds. 8. 6. The screen returns to MEASUREMENTS 20. With the cursor at Exit this Menu, press + twice to move cursor to Totalizer Unit. Press E. Screen goes back to MEASUREMENTS 20. Set Flow Direction 1. At MEASUREMENTS 20 press + five times to set the arrow Flow Dir. Press E. Set Full Scale Flow 1. Press + three times to move the cursor to FullScaleFlow. Press E. 27 2. At FLOW DIR 25 press + or - to select Uni-Directional or Bi-Directional. 6. Press E once to place cursor at Main Menu 00. 3. This type of screen shows for 2 seconds. 7. Press E once to move cursor back to the Main Screen. Set Damping Factor 1. At MEASUREMENTS 20 press + six times to move the arrow to Damping Factor. Press E. Empty Pipe Calibration 1. Press E once to move cursor to Main Menu. 2. In the MAIN MENU 00, press + once to move the cursor to Meter Setup. Next, press E. 2. At DampingFactor 26 press Exit this Menu, No Damping or press a Time Frame. 3. If No Damping is selected, this type of screen shows for 2 seconds. 3. In METER SETUP 10, press + five times to display Empty Pipe. Next, press E. 4. And this type of screen shows for 2 seconds when Damping Factor time is selected. 4. In EMPTY PIPE 15, press + once to place cursor at Cal. empty pipe. Next, press E. 5. MEASUREMENT 20 screen returns with the arrow pointing at Exit this Menu. 28 5. In EMPTY CALIB. 1E, press + once to place cursor at Volts = 0.000. Change the number as described earlier. Press E to move cursor to Cal[OFF] E=ON. 10.At Cal[OFF] E=ON, press or select E once to change to OFF to ON. (NOTE: Make sure the Flow Detector pipe is full of fluid.) NOTE: The full pipe voltage reading should be below 3.00 Volts. 6. At Cal[OFF] E=ON, press or select E once to change OFF to ON. NOTE: Make sure the Flow Detector Pipe is Empty. 11.Press + once to move cursor to Exit WITH save. Press E once to save the setting and return to EMPTY PIPE 15. NOTE: With the pipe empty the usage reading should be between 3.00 and 3.30 Volts. 12. In EMPTY PIPE 15, press + three times to place cursor at Enable/Disable. Next, press E. 7. Press + once to move cursor to Exit WITH save. Press E once to save the setting and return to EMPTY PIPE 15. 13.In EMPTY CONTROL 1G, press + twice to move cursor to Det[OFF] E=ON. Press E once to switch OFF to ON. 8. In EMPTY PIPE 15, press + twice to place cursor at Cal. full pipe. Then, press E. 14. Press + once to move cursor to Exit WITH save. Press E once to save the setting and return to EMPTY PIPE 15. 9. In FULL PIPE CAL 1F, press + once to move cursor to Volts = 0.00. Change the number as described earlier. Press E to move the cursor to Cal[OFF] E=ON. 29 6. The screen returns to DIG OUTPUT 1 33. Press or activate + twice, then press E. Pulse Output 1. Press E once to move cursor to Main Menu. 7. Into screen PULSE WIDTH 3E place a number. Press E. 2. In the MAIN MENU 00, press + three times to move the crusor to Inputs/Outputs. Next, press E. 8. The screen returns to DIG OUTPUT 1 33. Press + three times, then press E to see this screen. 3. At IN/OUTPUTS 30 press + three times to place the cursor at Digital Out. #1. Press E. This is only for Flow Set Point. (Reference Addendum C, Digital Output #1.) 9. Into screen SET PT. MIN 3F place a number as described earlier. Press E. 4. In DIG OUTPUT 1 33, press + once to move the cursor to Pulses / unit. Press E. You only need to do this if the function of output one (1) is to be Fwd Pulse or AMR (50ms pulse). Reference Addendum C, Digital Output #1. 10. The Screen returns to DIG OUTPUT 1 33. Press + four times, then press E. 5. In PULSES/UNIT 3D input a proper number as described earlier. Press E. 11. In SET PT. MAX 3G input a number as described earlier. This is only for Flow Set Point. (Reference Addendum C, Digital Output #1.) 30 12. Press or activate the letter E to return to screen DIG OUTPUT 1 33. 16. When Select Function is selected, screen Z1 appears for two (2) seconds. 17. It then switches to SELECT OUT #1 3J. Make a selection. Press E. 13. n DIG OUTPUT 1 33 press + five times, to Output Type. 14. In OUTPUT TYPE 3H press Normally Open or Normally Closed. • Select N.O. or N.C. for any output. (Reference Addendum C, Digital Out. #1.) 18. This type of screen shows for two (2) seconds and then returns to DIG OUTPUT 1 33. • Normally Opened or Closed **Selected** shows for two (2) seconds. Whichever selection is made from screen SELECT OUT #1 3J, the next screen always shows ** SELECTED**. • The screen then returns to DIG OUTPUT 1 33 where you will have to: 1. Select Functions screen 2. Screen Z1 appears 3. Select another screen. Do this until all required selections are programmed. The screen goes back to DIG OUTPUT 1 33. There are four Digital Outputs in this program. Perform the same for all of them, if applicable. 15. Pres + six times, or -once to Select Function. Analog Output 1. Press E once to move cursor to Main Menu. 31 2. In the MAIN MENU 00, press + three times to move the cursor to Inputs/Outputs. Next, press E. 8. Press the E button until the amp meter reads the desired No Flow set point. (The line to the left of the selection rotates as the E button is pressed, to show that there is activity.) 3. At IN/OUTPUTS 30 press + once to place the cursor at Analog Output. Press E. 4. In ANALOG OUTPUT 31 press + once to move the cursor to Range Select. Press E. 9. Press + to set the cursor at Exit With save. 10. Press E to Save. 5. In RANGE SELECT 3A select a range with the + button and then press E. 11. After pressing E, the menu goes back to ANALOG OUTPUT 31. Press + three times to place arrow at FS Calibrate. Press E once. 6. The screen returns to ANALOG OUTPUT 31. Press + twice to move the cursor to Zero Calibrate and then press E. 12. Screen ANALOG FS CAL 3C comes up. Press + to set the arrow across from Decrease AO or Increase AO. Press E button until amp meter reads the desired Full Scale set-point. Connect amp meter to mA analog output pins 5 and 6. 7. In ANALOG ZERO 3B make a selection of Decrease or Increase based on what the amp meter is reading. 32 13. Press + to set screen at Exit WITH save. 14. Press E to Save. 15. ANALOG OUTPUT 31 returns with the arrow pointing at Exit this Menu. Press E. 16. IN/OUTPUTS 30 returns with the arrow pointing at Exit this Menu. Press E. 17. MAIN MENU 00 returns with the arrow pointing at Exit this Menu. Press E. 18. The screen returns to the Main Screen. The above sections guided you through various screens and programming. These are just a few of the screen and screen settings. If you need help programming, call +49 7025 92080, to discuss screens and screen settings with a Technical Support Specialist. 33 Flow range: 0.1 to 39.4 ft/s (0.03 – 12 m/s) Pressure Limits: max. 16 bar Flow direction: Uni- or Bi- directional Ambient Temperature : -4°F to 122°F (-20°C to 50°C) Sizes : 1/4" to 12” (15 to 600 mm) Coil Power: Pulsed DC Conductivity: min. 5 µS/cm Pipe Spool Material: 316 stainless steel Accuracy: ±0,25% o.r. for velocites ≥ 1.64 ft/s (0.5 m/s) ±0.004 ft/s o.r. for velocities < 1.64 ft/s (0.5 m/s) Spool housing Material: Carbon Steel, welded Junction Enclosure Material Cast aluminium (powder coated paint) Electrode material: Alloy C, gold/platinum, tantalum, platinum/rhodium Flanges: Carbon Steel or 316 stainless steel Liner material: PTFE from ½” to 12” (15 to 600 mm) hard rubber from 1” to 12” (25 to 600 mm) Grounding Rings: 316 stainless steel or Alloy C remote mounted amplifier Grounding Electrode: Fluid temperature max ambient temp. 122°F (50°C) PTFE –4°F to 248°F (-20°C to 105°C) Hard rubber 32°F to 178°F (0°C to 80°C) Alloy C, gold/platinum, tantalum or platinum/rhodium AlSi12 acc. ISO3522 (CSN 424330) IP 67, 3 x M 20, painted Amplifier Enclosure: Electrical Classification: mounted amplifier PTFE –4°F to 212°F (-20°C to 100°C) Hard rubber 32°F to 178°F (0°C to 80°C) Mounted Version Ex II 3/1G EEx nA ia IIC T3 Remote Version Amplifier Ex II 3/1G EEx nA [ia] IIC T3 Remote Version Detector Ex II 3/1G EEx nA ia IIC T3 34 SPECIFICATIONS Power Supply: 85-265VAC, 45-65Hz or 24 VDC Power Consumption: 20W Accuracy: ± 0.25% of rate for velocities greater than 1.64 ft/s (0.50 m/s) ± 0.004 ft/s (± 0.001 m/s) for velocities less than 1.64 ft/s (0.50 m/s) Repeatability: <0.1% of full scale Flow Range: 0.10 to 39.4 ft/s (0.03 to 12 m/s) Fluid Conductivity: Min. 5.0 microSiemens/cm Flow Direction: Unidirectional or bidirectional (programmable) Totalization: 3 separate displayable totalizers – 10 digits (programmable - forward, reverse and net) Analog Outputs: 0-10mA, 0-20mA, 4-20mA (programmable and scalable) Voltage sourced (18VDC) – isolated. Max. loop resistance = 750 Ω Frequency Output: Open Collector – Max. full scale flow = 10Khz. Digital Outputs: (2) Open Collector, (programmable – scaled pulse, flow alarm, status, or frequency output) Max. 24VDC, 0.5W (2) AC solid state relay (programmable – flow alarm or status). Max. [email protected] Pulse Width: Open Collector, 5ms to 1 second (programmable) or automatic 50% duty cycle Min-Max Flow Alarm: Open collector or solid state relay (programmable – 0 to 100% of flow) Empty Pipe Detection: Field tunable for optimum performance based on specific application Excitation Frequency: Programmable - 3.75Hz, 7.5Hz or 15Hz Auxiliary Input: Max. 24VDC (programmable – positive zero return, external totalizer reset or preset batch start) Noise Dampening: 1 to 30 seconds (programmable) Units of Measure: U.S. gallons, Imperial gallons, MGD, cubic feet, cubic meters, liters, oil barrels, pounds, ounces, acre feet Low Flow Cut-Off: 0 to 100% of full scale (programmable) Zero-Point Stability: Automatic correction LC Display: 4 lines X 16 character alphanumeric – back light Displays: 3 totalizer values, flow rate, alarm status, output status, error / diagnostic messages Programming: Internal 3 button or external magnetic wand Galvanic Separation: £ 500 volts Electrical Classification: see namplate information page 8 Housing: Amplifier enclosure and remote junction enclosure: cast aluminum (powder coated paint) Material AlSi12Mn (ISO 3522) Housing Rating: Amplifier enclosure and remote junction enclosure – IP67 Mounting: Direct detector mount or remote wall mount – bracket included. (for remote mount, max. cable distance = 100 ft (30M) Field Wiring Entry Ports: (3) M20, internal thread Ambient Temperature: -40°F to 122°F (-40°C to 50°C) Communication: RS232C serial, standard ANSI terminal compatible data stream 35 ADDENDUM B Replace fuses with fuses of the same ampere rating and type. Refer to wiring diagrams when ampere ratings are unknown or questionable. MAINTENANCE Mandatory, routine or scheduled maintenance should not be required for the Badger M3000 Mag Meter electronics or flow tube after proper installation. Detector coil and incoming amplifier power supplies are each protected by a fuse in the amplifier. However, some occurrences may require personnel to perform the following: · Flow Tube and Electrode Cleaning · Fuse Replacement · Amplifier I&C Card Stack Replacement Detector coils are protected by a 630mA, 250VAC, slow blow fuse Incoming amplifier power is protected by a 500mA, 250VAC, slow blow fuse These maintenance procedures are discussed in this section. Fuse ratings are listed on the circuit board, next to the fuse holders. Refer to the M-Series, Model 4000 To access and replace fuses: Disconnect main power to the unit before attempting any device maintenance or cleaning. Do not clean components inside the amplifier or junction box. 1. Remove display chamber cover. Turn the cover counterclockwise to remove it from the amplifier. If necessary, use a strap wrench. Flow Tube and Electrode Cleaning At times flow tube, electrodes, amplifier/junction box housings and the amplifier window may need periodic cleaning, depending on process fluid properties, fluid flow rate and surrounding environment. Clean the flow tube and electrodes by following the material handling and cleaning procedures documented in MSD Sheets for the products(s) that were in contact with the flow tube and electrodes. Should flow tube and/or electrode cleaning become necessary: 2. Remove the 2 display card screws and washers. 1. Disconnect detector from pipeline. 2. Clean electrodes with isopropyl alcohol or fresh, clean water depending on the chemical compatibility of the measured fluid. 3. Reconnect detector to pipeline Fuse Replacement Disconnect main power to the unit before attempting any device maintenance. Risk of electrical shock. Replace fuse ONLY with the same type and rating. Authorized personnel must perform fuse replacements. 36 Because PCBs are complex circuits, with all meter functions enabled through multiple links and layers, determining the exact board and circuit that is causing a system problem is difficult and usually requires test equipment. 3. Tilt card up/out approximately 45 degrees at the holding clips. Should a meter problem occur: 1. Call Badger Meter at +49 7025 92080, and discuss the problem with a Technical Support Specialist. 2. If the problem appears to originate in a PCB, it will be recommended that the entire PCB stack be removed and returned to Badger Meter. 4. Fuses are in the round/black fuse holders. Use a flat screwdriver to turn the holder cap counterclockwise until the lock tab reaches the opening and the holder and fuse disengage. PCB Stack in Amplifier REMOVE PCB STACK Disconnect main power to the unit before attempting any device maintenance. 5. Replace the fuse. 1. Remove display/programming chamber cover. Turn the cover counterclockwise to remove it from the amplifier. If necessary, use a strap wrench. 6. Reverse steps 1 to 3 to assemble the unit. 2. Remove 2 display card screws and washers. Place in storage for reuse. Amplifier Printed Circuit Board (PCB) Stack Replacement All M3000 mag meters operate through printed circuit boards (PCBs) housed in the amplifier. The PCBs are grouped in a stack located behind the display/control card in the display/programming chamber. 37 3. Tilt display card up/out approximately 45 degrees at the holding clips. 6. Disconnect the Power, Coil, Electrode and I-O plugs and harnesses from the PCB interconnect card. 4. Gently pull card down and out from between holding clips. 5. Disconnect display card plug from left side of PCB display. 7. Remove the 4 screws holding the PCB stack in the amplifier housing. Place in storage for reuse. 38 When placing display card back into PCB Display, verify mark and red stripe are aligned with Number 1. 8. Gently remove the card stack from the amplifier housing. Do not tug or pull to remove the stack. Ease it out carefully. For safety reason, all replaced parts should be provided by Badger Meter Europe. Use only Equipment as designed and specified as in the manual. 9. Carefully wrap and package the PCB stack and display card. Send back to Badger Meter. 10. Reverse these steps to install a PCB stack. When replacing the PCB stack, be sure to place the two lower feet of the circuit board support along the edges of the amplifier housing. Gently move the stack back into the housing until the to holes at the top mate with the housing. (The stack is not connected to the circuit board in this picture.) 39 ADDENDUM C In the Flow Chart: 1. Each separate screen is inside a box. 2. If a screen “branches”, an arrow points right and each screen of the branch is in a box. 3. When there is text, but it’s not in a box, that signifies a parameter setting and a short explanation is given of the parameter. Mark on the chart what parameters need to be set. Note parameter settings on the chart. Program your amplifier accordingly. Keep the chart as a reference for other shifts and personnel and to monitor meter performance. Menu Structure 40 41 ADDENDUM D Atex Certificate 42 Please see our website at www.badgermeter.de for specific contacts. Due to continuous research, product improvements and enhancements, Badger Meter reserves the right to change product or system specifications without notice, except to the extent Badger Meter Europa GmbH Nürtinger Str. 76 72639 Neuffen (Germany) www.badgermeter.com Copyright © Badger Meter, Inc. 2004. All rights reserved 43