Device-related safety information
Safety
Effects of the magnetic field on
devices
B
A
A.3
Products and magnetizable data carriers (for example, storage
disks, magnetic tapes, diskettes, credit cards) and precision
devices (for example, watches, pagers, hearing aids) may be
damaged by the magnetic field.
A.3
Safety distances
A.3
This table shows the effects of the magnetic fringe field on
devices located in the vicinity of the magnet and the safety distances required. Observe the minimum distances to be maintained from the center of the magnet’s X, Y, and Z axes.
A.3
A.3
Magnetic flux density
3 mT
1 mT
0.5 mT
0.2 mT
0.1 mT
0.05 mT
A.3
A.3
A.3
A.3
A.3
A.3
A.3
Minimum distances
(x = y = radial, z = axial)
x = 2.2 m
z = 3.9 m
x = 2.8 m
z = 5.0 m
x = 3.1 m
z = 6.0 m
x = 3.5 m
z = 7.5 m
x = 3.7 m
z = 9.2 m
x = 7.3 m
z = 11.2 m
A.3
Devices affected
A.3
A.3
A.3
A.3
A.3
A.3
A.3
A.3
A.3
Small motors, watches, cameras,
credit cards, magnetic media
Oscilloscopes, computers, disk
drives, shielded color monitors
B/W monitors, magnetic media, cardiac pacemakers, insulin pumps
Siemens CT systems
A.3
A.3
A.3
A.3
A.3
A.3
Siemens linear accelerators
A.3
A.3
A.3
A.3
X-ray I.I.(I.I. = image intensifiers),
gamma cameras, third party linear
accelerators
A.3
0.0
A.3-2
syngo MR 2004A
Revision 03
MAGNETOM Trio System Manual
0.0