Physiological effects MR system operation Normal Operating Mode Normal Operating Mode can be safely used for all patients. First Level Controlled Operating Mode C.3 C.3 C.3 In First Level Controlled Operating Mode, patients may experience noticeable stress levels depending on the measurement program selected. We recommend establishing a procedure to ensure medical monitoring of the patient. C.3 Switching between operating modes C.3 To switch from Normal to First Level Controlled Operating Mode, the user has to explicitly select and confirm the change. A corresponding prompt will appear in the dialog window of the MR console. In First Level Controlled Operating Mode, the patient has to be medically monitored. C.3 0.0 C.3-2 syngo MR 2004A Revision 03 MAGNETOM Trio System Manual 0.0