Section 3.2 Prepare a 2 × 25 bp mate-paired library
Shear the DNA
Table 32
3
Recommended shearing conditions for mate-paired library insert sizes
Shearing method
Shearing conditions
600 to
800 bp
Covaris™ shearing in 20% glycerol
•
•
•
•
•
•
•
•
•
Number of Cycles: 75
Bath Temperature: 5 °C
Bath Temperature Limit: 12 °C
Mode: Frequency sweeping
Water Quality Testing Function: Off
Duty cycle: 2%
Intensity: 7
Cycles/burst: 200
Time: 10 seconds
800 to
1000 bp
Covaris™ shearing in 20% glycerol
•
•
•
•
•
•
•
•
•
Number of Cycles: 30
Bath Temperature: 5 °C
Bath Temperature Limit: 12 °C
Mode: Frequency sweeping
Water Quality Testing Function: Off
Duty cycle: 2%
Intensity: 5
Cycles/burst: 200
Time: 10 seconds
1 to 2 kb
HydroShear® Standard Shearing
Assembly
• SC5 ‡
• 20 cycles
2 to 3 kb
HydroShear® Standard Shearing
Assembly
• SC9
• 20 cycles
3 to 4 kb
HydroShear® Standard Shearing
Assembly
• SC13
• 20 cycles
4 to 5 kb
HydroShear® Standard Shearing
Assembly
• SC15
• 5 cycles
5 to 6 kb
HydroShear® Standard Shearing
Assembly
• SC16
• 25 cycles
(13 mm × 65 mm borosilicate tube)
(13 mm × 65 mm borosilicate tube)
‡ Speed code (SC): 5.
IMPORTANT! If you are using the Covaris™ S2 System, set the chiller
temperature to between 2 to 5 °C to ensure that the temperature reading in
the water bath displays 5 °C. The circulated water chiller should be
supplemented with 20% ethylene glycol.
2. If the DNA source is not limiting, ensure that the shearing conditions result in the
desired insert sizes. Shear 5 μg DNA and run 150 ng sheared DNA on a 1%
E-Gel® EX Gel according to the manufacturer's specifications.
SOLiD™ 4 System Library Preparation Guide
95
Chapter 3 Mate-Paired Library
Preparation
Insert size