Download ID-H0530/0560 Digimatic Indicator User`s Manual

Transcript
User’s Manual
No. 99MAH016B3
SERIES No. 543
ID-H0530/0560
Digimatic Indicator
User’s Manual
Read this User’s Manual thoroughly
before operating the instrument. After reading,
retain it close at hand for future reference.
FOREWORD
In order to use this Digimatic Indicator properly, read this manual thoroughly before
operating the instrument. After reading this manual, store it in a safe place for future
reference.
Mitutoyo assumes no liability to any party for any loss or damage, direct or indirect,
caused by use of this instrument not conforming to this manual.
Information in this document is subject to change without notice.
 2013 Mitutoyo Corporation. All rights reserved.
PRECAITIONS
Precautions against the Power Supply
WARNING
Wrong handling that neglects the following precautions could result in serious injury or
death.
 This instrument is externally powered from the AC power supply via the AC adapter.
Be sure to use the Mitutoyo specified AC adapter (standard accessory: No.06AEG180).
 If this instrument is used near the machines that require a high voltage, high frequency,
or large current, do not share the power line of the AC adapter with those machines.
 Use the AC power supply that absolutely conforms to the voltage and frequency
indicated on the AC adapter. If the AC adapter is used under the condition exceeding
the permissible voltage range, it could cause an explosion or ignition.
 If this instrument is not used for an extended period of time, make sure that the AC
adapter is unplugged from the power supply to prevent fire, etc.
Precaution in Discarding This Instrument
WARNING
Wrong handling that neglects the following precautions could result in serious injury or
death.
 This instrument uses liquid crystal. If discarding this instrument, observe the ordinance
and regulation by each local government.
 The liquid crystal includes an irritating substance. Should any liquid content be
accidentally applied to your eyes or skin, wash them with clean flowing water. If you
swallow the liquid content, immediately rinse your mouth, and after drinking a large
amount of water, vomit it, and then consult a doctor.
Precaution for Use
The following acts or circumstances will cause a failure or a malfunction in this instrument.
Care should be exercised.
IMPORTANT  Do not apply a sudden impact such as a drop or excessive force to this instrument.
 Do not disassemble or retrofit the instrument.
 Do not operate keys with a sharp tip (such as the tip of a screwdriver or ball point pen).
 Avoid using and storing this instrument at sites where it is subject to direct sunlight and
excessively hot or cold environments.
 The use of this instrument at a site with tenuous air or high pressure could cause a
failure due to deterioration in material.
No. 99MAH016B
i
 Avoid storing the instrument at sites surrounded with high humidity and dust, and avoid
using it at sites where it is subject to direct splashes of water and oil.
 If a high-voltage device such as an electric pen is used, electronic parts may be
damaged. Also, using the instrument at a site where large electric noises are
generating could cause a malfunction.
 Use this instrument at a site with minimum vibration while securely fixing it on the
comparator stand.
 Do not use this instrument in such a way that a perpendicular load or a twisting force is
applied to the spindle.
 To clean the instrument, use a dry soft cloth or cotton swab, or one soaked in diluted
neutral detergent. Use of organic solvents (such as thinner or benzene) may result in
failure.
 The dirt in the spindle and cylinder will cause a malfunction.
 Wipe off dirt on the spindle with a cloth damped with alcohol, and then lightly wipe the
spindle with a cloth damped with a small amount of low-viscosity oil.
 Clean the inside of the cylinder periodically. After removing the cylinder by rotating it,
brush off the dust and dirt on and around the piston and in the cylinder using a brush or
a blower brush.
 Take sufficient damage-preventive measures (safety measures) for an accidental
failure of this instrument.
The following precautions are important to obtain correct measurement results. Care
should be taken.
IMPORTANT  Perform sufficient thermal stabilization of this instrument and an object to be
measured before starting measurement. Use this instrument at a site which is subject
to as little thermal fluctuation as possible.
 To perform stabilized measurement, allows approximately 20 minutes after turning on
the power. The base line could be drifted approximately 0.5μm.
 Do not use this instrument with either stroke end specified as the reference point.
 Note that, if this instrument is used in other than the specified conditions, the functions
and performance will not be guaranteed.
WARRANTY
This Digimatic indicator has been manufactured under Mitutoyo’s rigorous quality control
system. In the event that this instrument should fail within one year from the original date
of purchase through normal use, we will repair or replace it at our option, free of charge,
upon its prepaid return to Mitutoyo. Contact your dealer or the nearest Mitutoyo Service
Center.
However, the following failure or damage may be subject to a repair charge even within
the warranty period:
1 Unit failure or damage arising from improper handling, or unauthorized retrofit or
repair by the user.
2 Unit failure or damage as the result of moving, dropping, or transporting after
purchase.
3 Unit failure or damage due to fire, salt, gas, abnormal voltage, or natural catastrophe.
This Warranty is valid only in the country of original purchase.
NOTES ON OVERSEAS TRANSFER
Please contact Mitutoyo before transferring it abroad.
ii
No. 99MAH016B
Contents
FOREWORD .......................................................................................................................................................i
PRECAITIONS ....................................................................................................................................................i
WARRANTY .......................................................................................................................................................ii
NOTES ON OVERSEAS TRANSFER ...............................................................................................................ii
1
2
NAME OF RACH PART ......................................................................................................................... 1-1
1.1
Part Names of the Main Unit ........................................................................................................ 1-1
1.2
Details of the LCD Unit ................................................................................................................. 1-2
INSTALLATION...................................................................................................................................... 2-1
2.1
3
2.2
Attaching the Clamp Filter............................................................................................................ 2-1
Mounting on a Stand or a Jig ....................................................................................................... 2-1
2.2
Attaching the Lifting Lever ........................................................................................................... 2-1
2.3
Attaching the Release ................................................................................................................... 2-2
2.4
Mounting the Lug or Special Back .............................................................................................. 2-2
2.5
Replacing the Contact Point ........................................................................................................ 2-2
OPERATING PROCEDURE .................................................................................................................. 3-1
3.1
Starting/Stopping This Instrument .............................................................................................. 3-1
3.1.1
Connecting to the power supply .......................................................................................... 3-1
3.1.2
Start / Stop .............................................................................................................................. 3-1
3.2
Parameter Setup Mode ................................................................................................................. 3-1
3.2.1
Switching the unit system between inch and mm during display of “unit” (only in inch spec) .... 3-2
3.2.2
Switching the digital resolution during display of “res.” .................................................. 3-2
3.2.3
Switching the analog range (analog resolution) during display of “range” .................... 3-3
3.2.4
Switching the count direction during display of “dir” ....................................................... 3-3
3.2.5
Switching the I/O format between SPC(Digimatic) and RS-232 during display of “output” ......... 3-3
3.2.5.1
RS-232 setup.................................................................................................................... 3-3
3.2.6
Setting ID No. for RS-232 communication and remote control during display of “id” .. 3-4
3.2.7
Switching the length measuring system during display of “start” .................................. 3-4
3.2.8
Returning all parameters to factory defaults during display of “reset” .......................... 3-4
3.3
Measurement Mode ....................................................................................................................... 3-5
3.3.1
3.3.1.1
Setting the length measuring system and reference point ............................................... 3-5
INC system (Zero set) ...................................................................................................... 3-5
3.3.1.2
Preset system (setting the preset value) .......................................................................... 3-5
3.3.2
Switching a measurement type in the measurement mode .............................................. 3-6
3.3.2.1
Tolerance judgment (setting/checking the upper and lower limit values) ......................... 3-6
3.3.2.2
Run-out peak hold during display of “TIR” ....................................................................... 3-7
3.3.2.3
Maximum value peak hold during display of “Max” .......................................................... 3-7
3.3.2.4
Minimum value peak hold during display of “Min” ............................................................ 3-8
3.3.3
Data output / Data hold ......................................................................................................... 3-9
No. 99MAH016B
iii
3.3.4
Function lock .......................................................................................................................... 3-9
3.3.5
Centering the pointer ............................................................................................................. 3-9
3.4
Operation with the Remote Control (Option) .............................................................................. 3-9
DATA INPUT/OUTPUT ........................................................................................................................... 4-1
4
4.1
SPC (Digimatic) Input/Output ....................................................................................................... 4-1
4.1.1
Connecting procedure ........................................................................................................... 4-1
4.1.2
I/O connector .......................................................................................................................... 4-1
4.1.3
Output data format (DATA1) .................................................................................................. 4-2
4.1.4
Timing chart ............................................................................................................................ 4-2
4.2
RS-232 Input/Output ...................................................................................................................... 4-4
4.2.1
Connecting procedure ........................................................................................................... 4-4
4.2.2
I/O connector .......................................................................................................................... 4-4
4.2.3
Communication specification ............................................................................................... 4-4
4.2.4
Communication command (I/O format) ............................................................................... 4-5
4.2.5
Communication sample ........................................................................................................ 4-7
5
ERROR MESSAGE AND ACTION ........................................................................................................ 5-1
6
SPECIFICATION .................................................................................................................................... 6-1
6.1
Specifications of the main unit .................................................................................................... 6-1
6.2
Standard accessories ................................................................................................................... 6-1
6.3
Optional accessories .................................................................................................................... 6-2
iv
No. 99MAH016B
1
1
NAME OF RACH PART
This chapter describes the name of each part and the LCD display
contents.
1.1
Part Names of the Main Unit
(1) Cylinder
(2) Lug mounting screw
(3) Stem
(4) Spindle
(5) Contact point
mm Type : Parts No.901312
inch Type : Parts No.21BZB005
7
1
6
(6) DC jack
8
(7) I/O connector
(8) Power switch
9
2
(9) LCD unit
(10) Remote control reception
(11) MODE key
10
(12) SET/ZERO key
11 1 13
2
(13) DATA/Fn. key
14
(14) Release attaching hole
(15) AC adapter
(16) Spindle lifting lever
(17) Clamp filter
3
4
16
5
17
15
NOTE  Be sure to attach the clamp filter (ferrite core) to the AC adapter. For detailed
information about attachment, refer to section “2.1 Attaching the Clamp Filter”.
No. 99MAH016B
1-1
1.2
Details of the LCD Unit
2
6
5
4
1
7
3
9
8
11
10
12
17
16
15
14
13
(1) Function lock
(2) Output format display
(3) Upper over-range
(4) Upper pointer
(5) Pointer
(6) Lower pointer
(7) Lower over-range
(8) Analog range
(9) Tolerance judgment
(10) Maximum value hold measurement
(11) Runout hold measurement
(12) Minimum value hold measurement
(13) Display value unit
(14) Preset value setup mode
(15) Comparison measurement
(16) Data hold mode
(17) Reverse direction measurement
NOTE  When tolerance judgment measurement is performed, the upper pointer (4) and lower
pointer (6) blink at slow intervals to indicate the upper limit value and lower limit value,
respectively.
 In the total runout, maximum value, or minimum value hold measurement mode, the
upper pointer (4) and lower pointer (6) blink to indicate the maximum value and
minimum value, respectively.
 If the pointer (5) coincides with the upper pointer (4) and lower pointer (6), these
pointers blink at quick intervals.
1-2
No. 99MAH016B
22
INSTALLATION
This chapter explains how to mount this Digimatic indicator on a
stand or a jig and how to attach or replace optional accessories.
2.1
Attaching the Clamp Filter
Use this instrument by wrapping the AC adapter cord around the clamp filter (ferrite core)
as shown in the figure below.
2.2
Mounting on a Stand or a Jig
Use this instrument with its stem being clamped on the comparator stand (option) or an
appropriate jig.
IMPORTANT  If possible, avoid fixing the stem by directly clamping it with a clamp screw, etc. If the
stem is clamped with a tightening torque of 150cN•m or more, the indicator could cause
a malfunction.
NOTE  Clamp the stem so that the spindle is oriented perpendicularly with respect to the
reference plane and a measuring surface. If the axis line (spindle) is not perpendicular
to the reference plane (measuring surface), an error is added to a measured value.
For example, assuming a tilt angle of the axis line from the reference plane as , error 
per measuring length 30mm will be the following according to the tilt angle:  = 1°:  =
0.0045mm,  = 2°:  = 0.0185mm, and  = 3°:  = 0.0410mm.
 To mount this indicator on a jig, it is recommended that the stem be clamped with a
slotted holder that has a coupling portion of approximately 8G7 (AGD: 9.52) +
0.02/+0.005mm.
2.3
Attaching the Lifting Lever
While holding the spindle from the counter side (B) so as not to bend it, engage the
spindle with the groove on the lifting lever (No.137693/standard accessory).
No. 99MAH016B
2-1
2.4
Attaching the Release
Remove the rubber cap in the release attaching hole, and then screw in the release
(No.540774/option) by hand.
IMPORTANT  Do not pull or hit the release strongly. Some screw thread at its end may break.
 Inserting any other than the release or applying excessive force to the release mount
could result in failure.
NOTE  The amount of lifting the spindle when using the release is approximately 30mm from
the bottom dead center.
2.5
Mounting the Lug or Special Back
This Digimatic indicator can be equipped with the back with lug (No.101040: ISO/JIS type
or No.101306: ANSI/AGD type, both options) for Mitutoyo standard dial indicators or either
of various special backs.
Remove the four screws on the rear and mount the back
with lug with these screws.
NOTE
2.6
 For information about the assortment of accessories such as special backs, special
contact points, and extension rods, refer to the Mitutoyo general catalog, measuring
tool general catalog, or dial indicator catalogs.
Replacing the Contact Point
Various special contact points and extension rods of Mitutoyo dial indicator options are
available.
Hold the spindle with pliers while placing a waste cloth between them so that the
spindle will not rotate, and then detach/attach the contact point by pinching and
rotating it with other pliers.
IMPORTANT  Unless the spindle is held unturned in the above procedure, a failure could be caused
in the Digimatic indicator. Causing damage to the spindle could result in a malfunction.
 Upon replacement of the contact point the external dimensions, measuring force,
measuring orientation restriction, etc., of the indicator may be changed.
 When the contact point other than the standard contact point is used, the error may be
added due to the shape of the contact point. (e.g. Perpendicularity of the contact point
and parallelism with the reference surface for the flat contact point, runout of the roller
contact point, etc.)
[2.5]
Back with lug
[2.4]
2-2
[2.6]
Screw (4pieces)
No. 99MAH016B
3
OPERATING PROCEDURE
3
3.1
This chapter explains the operating procedure of the Digimatic
indicator.
Starting/Stopping This Instrument
3.1.1
Connecting to the power supply
1
Remove the DC jack cover located at the upper on the Digimatic indicator ID-H, and
then push in the DC plug of the AC adapter (standard accessory) all the way until it
stops.
2 Plug the AC plug at the other end of the AC adapter firmly into an AC outlet or
extended outlet strip.
IMPORTANT  To supply or shut off the power to this instrument, be sure to turn it on or off with the
[ON/OFF] switch on the side of the main body. If the power to the instrument is shut off
during operation, damage to its internal parts could result.
3.1.2
Start / Stop
1
2
3
3.2
Slide the [ON/OFF] switch on the side of the main body upward or downward to start
(restart) or stop (terminate) measurement with this instrument.
To start measurement, slide the [ON/OFF] switch upward.
To stop measurement, slide the [ON/OFF] switch downward.
Parameter Setup Mode
Parameters such as “Digital resolution” and “Counting direction” can be changed in
the Parameter Setup mode.
To put this instrument in the Parameter Setup mode, observe the following procedure.
1
2
3
4
NOTE
No. 99MAH016B
Press and hold the [MODE] key for a long time (2 seconds or more) in the
Measurement mode. Then, the mode is switched to the Parameter Setup mode, the
“res.” (“unit” for inch spec) symbol is blinking on the LCD.
Press and release the [MODE] key quickly (for less than 2 seconds) to change setup
items.
When the item to be set is displayed, press the [SET/ZERO] key quickly, then change
the setting.
When the setting of a desired setup item has been changed, press and hold the
[MODE] key for a long time to return to the Measurement mode.
 The setting contents remain in memory even when the power is turned off.
3-1
Parameter Setup List
3.2.1
”unit”
(Only for inch spec)
Switching the unit system
between inch and mm
Default
(Factory defaults)
“in”
(Only for inch
spec)
3.2.2
“res.”
Switching the digital
resolution
mm spec: 0.0005
Inch spec: 0.00002
3.2.3
“range”
Switching the analog range
mm spec: 0.01
Inch spec: 0.0004
3.2.4
“dir”
Switching the count
direction
“ ”
(No display: +
direction)
3.2.5
“output”
Switching the I/O format
between SPC and RS232
SPC
(Digimatic)
3.2.6
“id”
3.2.7
“start”
3.2.8
“reset”
Reference
3.2.1
Notation in
the manual
Setup content
Setting ID No. for RS232
communication remote
control
Switching the length
measuring system between
INC and preset systems.
00
INC
Returning parameters to
defaults.
Switching the unit system between inch and mm during display of “unit”
(only in inch spec)
1
2
3
4
3.2.2
Display
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “unit” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “unit” from blinking to lighting-up and
blinks the currently-set unit system.
Press the [MODE] key quickly to switch to a desired unit (in or mm).
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “res.”.
Switching the digital resolution during display of “res.”
1
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “res.” on the LCD starts blinking.
2 Press the [SET/ZERO] key quickly to blink the currently-set digital resolution.
3 Press the [MODE] key quickly to switch the setting to a desired resolution.
Unit system
Selectable resolution
mm
0.0005mm
0.001mm
inch (only for inch spec)
.00002”
.00005”
.0001”
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “range”.
IMPORTANT  Upon switchover of the current unit system or digital resolution, this instrument
automatically makes conversions of the preset values and tolerance limit values. After
changing such a parameter, be sure to check the converted values.
3-2
No. 99MAH016B
3. OPERATING PROCEDURE
3.2.3
Switching the analog range (analog resolution) during display of
“range”
1
2
3
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “range” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “range” from blinking to lighting-up
and blinks the currently-set analog range.
Press the [MODE] key quickly to switch the setting to a desired analog range.
Digital
resolution
0.0005mm
0.001mm
.00002”
.00005”
.0001”
RS-232
command *
4
Selectable analog range
±0.01
±0.02
±.0004
±.001
±.002
±0.02
±0.04
±.0008
±.002
±.004
±0.05
±0.1
±.002
±.004
±.01
±0.1
±0.2
±.004
±.01
±.02
±0.2
±0.4
±.008
±.02
±.04
±40
±40
±2
±2
±2
±80
±80
±4
±4
±4
A01
A02
A05
A10
A20
AFS
AFL
*: For information about RS-232 commands, refer to section 4.2.4.
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “dir”.
Switching the count direction during display of “dir”
3.2.4
1
2
3
4
NOTE 
3.2.5
Switching the I/O format between SPC(digimatic) and RS-232 during
display of “output”
1
2
3
4
3.2.5.1
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “output” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “output” from blinking to lighting-up
and blinks the currently-set I/O format.
Press the [MODE] key quickly to switch the setting to a desired I/O format (between
SPC and RS-232).
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “id” (section 3.2.6) if SPC is specified. If RS-232 is specified as the I/O format,
the procedure is continuously switched to the RS-232 setup (section 3.2.5.1).
RS-232 setup
1
2
3
4
5
No. 99MAH016B
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “dir” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “dir” from blinking to lighting-up. Only
“dir” is displayed if the currently-set direction is “+” count, and “REV” blinks at the left
of “dir” if the direction is “-" count.
Press the [MODE] key quickly to switch the setting to a desired direction (between +
and -).
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “output”.
Upon change of the count direction, the sign (+/-) of a display value is also changed.
Before calling a preset value, perform setup of the count direction.
When the procedure is switched to the RS-232 setup, “bps” starts blinking.
Press the [SET/ZERO] key quickly to blink the currently-set baud rate.
Press the [MODE] key quickly to switch the setting to a desired baud rate (4800 or
9600).
Quickly press the [SET/ZERO] key to apply the communication setting and switch the
setup item to the next “p-bit.”.
Referring to the following table, set the parity bit and data bit according to the
3-3
procedure in steps 2 to 4.
Setup item
Display
Baud rate
(Bit rate)
6
NOTE
3.2.6
3
4
5
4
NOTE 


Parity bit
0 (none)
1 (odd)
Data bit
7 bits
8 bits
2 (even)
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “id” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “id” from blinking to lighting-up and
blinks the currently-set ten’s place of ID No.
Press the [MODE] key quickly to switch the ten’s place to a desired number (0 to 9).
Press the [SET/ZERO] key quickly to apply the specified number and blink the current
unit’s place of ID No.
Press the [MODE] key quickly to switch the unit’s place to a desired number (0 to 9).
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “start”.
When the ID No. of the instrument has been set, only the signals from the remote
control and RS-232 communication with the same ID No. are received. If ID No. on the
remote control and RS-232 communication sides is set to “00”, however, the
instrument receives any signals irrespective of its ID No.
The remote control (option) ID No. can be set to one of “00” to “07” or “09” to “15”.
The RS-232 communication ID No. can be set to one of “00” to “09”.
Switching the length measuring system during display of “start”
1
2
3
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “start” on the LCD starts blinking.
Quickly pressing the [SET/ZERO] key changes “start” from blinking to lighting-up and
blinks the currently-set length measuring system.
Press the [MODE] key quickly to switch the setting to a desired length measuring
system.
Display
Setup content
Baud rate
Always starts up under the INC system.
(Bit rate)
“
4
“(no display)
Starts up under the length measuring system at the most
recent shutdown.
Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “reset
Returning all parameters to factory defaults during display of “reset”
1
2
3
3-4
9600bps
Setting ID No. for RS-232 communication and remote control during
display of “id”
2
3.2.8
4800bps
When the data bit setting has been applied, the setup item is switched to the next “id”.
 Initial setting (factory defaults) for RS-232 is as follows:
Bit rate: 9600bps
Parity bit: even
Data bit: 7bit
1
3.2.7
Setup content
Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “reset” on the LCD starts blinking.
When pressing the [SET/ZERO] key quickly, “no” on the LCD starts blinking.
To return the parameters to defaults, press the [MODE] key quickly to switch to “yes”
under blinking.
No. 99MAH016B
3. OPERATING PROCEDURE
4
IMPORTANT
3.3
Quickly press the [SET/ZERO] key to restore the defaults and switch the setup item to
the next “res.” (“unit” in inch spec).
 When resetting the parameters, the preset values and the upper and lower tolerance
limit values are also cleared. Set these values again.
 Do not reset during RS-232 communication. Since the “I/O format” and “ID No.” are
changed, RS-232 communication will not be conducted any more.
Measurement Mode
3.3.1
Setting the length measuring system and reference point
This instrument has two length measuring systems, INC (Comparison) system and Preset
system.
Press and hold the [SET/ZERO] key for a long time in the normal measurement mode or
the tolerance judgment mode to switch the length measuring system between INC and
Preset systems.
NOTE  The “INC” symbol lights up under the INC system and goes off under the Preset system.
 To set the reference point, make sure that this instrument is clamped securely with the
spindle pushed in slightly.
3.3.1.1
INC system (Zero set)
In the INC system, the instrument measures a distance from the zero-set position.
When the [SET/ZERO] key is pressed quickly in the normal measurement mode or
the tolerance judgment mode, the instrument is zero-set at the current position.
3.3.1.2
Preset system (setting the preset value)
In the Preset system, the instrument measures a distance from the reference point
that has been preset with an arbitrary value (Calling the preset value).
1
2
NOTE
No. 99MAH016B
When the [SET/ZERO] key is pressed shortly in the normal measurement mode or
the tolerance judgment mode, the most recent preset value is displayed, blinking the
“P” symbol.
Press the [SET/ZERO] key quickly to set the current spindle position to the preset
value, putting the instrument in the measurable state.
To change a preset value, press and hold the [SET/ZERO] key for a long time to blink
the “+” or “-” symbol, entering the state where a preset value can be set. Enter a sign
or a numeral observing the following procedure in steps (1) to (4).
(1) Quickly press the [MODE] key repeatedly until a desired sign (+ or -) or numeral is
displayed.
(2) Quickly press the [SET/ZERO] key to apply the sign or numeral and switch the
setup item to the next digit.
(3) Repeat the procedure in steps (1) and (2) to enter a numeral in all digits. When
the least significant digit has been decided, the “P” symbol blinks up.
(4) Press the [SET/ZERO] key quickly to set the current spindle position to the preset
value, putting the instrument in the measurable state.
 The setting contents remain in memory even when the power is turned off.
 Press and hold the [MODE] key for a long time in the process of presetting to restore
the state before presetting.
 Press the [SET/ZERO] key while holding down the [DATA/Fn] key to reverse the order
of digits.
 Press the [MODE] key while holding down the [DATA/Fn] key to reverse the order of a
numeric value.
 Preset value can be called only for the first time after switching to the preset system. It
is impossible to call the preset value for the second time and afterward.
3-5
3.3.2
Switching a measurement type in the measurement mode
This instrument is provided with measurement mode described in the table below in
addition to normal measurement mode.
Press the [MODE] key quickly to switch the measurement mode to a desired
measurement type.
3.3.2.1
Reference
---
Measurement mode
Normal
Display
Setup content
Performs normal measurement.
3.3.2.1
Tolerance judgment
“<○>”
3.3.2.2
Runout peak hold
“TIR”
3.3.2.3
Maximum value peak
hold
“Max”
Performs peak hold
measurement of the maximum
value.
3.3.2.4
Minimum value peak
hold
“Min”
Performs peak hold
measurement of the minimum
value.
Performs tolerance judgment
measurement.
Tolerance judgment is
performed for each hold value
in the runout, maximum value,
and minimum value hold
measurements.
Performs peak hold
measurement of the runout
(total indicator reading).
Tolerance judgment (setting/checking the upper and lower limit values)
Tolerance judgment is performed for the current value in the normal measurement
mode, whereas it is performed for each peak hold value in each peak hold mode.
It is possible to set the upper and lower limit values separately for individual INC
system and Preset system.
It is not possible to set the upper and lower limit values separately for each
measurement mode.
The tolerance judgment is displayed with “<  >” and the backlight (red for NG and
green for OK).
Quickly press the [MODE] key repeatedly until “<  >” on the LCD starts blinking.
When pressing the [SET/ZERO] key quickly, symbol “off” or “on” starts blinking.
Press the [MODE] key quickly to switch the blinking symbol to “on”.
Press the [SET/ZERO] key quickly. Then, the most recently-set upper limit value is
displayed and the “>” symbol starts blinking.
5 Quickly press the [SET/ZERO] key again to apply the upper limit value and switch the
setup item to the lower limit value.
To change the upper limit value, press and hold the [SET/ZERO] key for a long time to
blink the “+” or “-” symbol, entering the state where the upper limit value can be set.
1
2
3
4
3-6
No. 99MAH016B
3. OPERATING PROCEDURE
The same as the procedure in section 3.3.1.2(Setting the preset value), enter a
numeric value, then apply it with the [SET/ZERO] key to set the upper limit value.
When the upper limit value has been decided, the most recently-set lower limit value
is displayed and the “<” symbol starts blinking.
Quickly press the [SET/ZERO] key again to apply the lower limit value, thereby
entering the state where tolerance judgment can be performed.
To change the lower limit value, press and hold the [SET/ZERO] key for a long time to
blink the “+” or “-” symbol, entering the state where the lower limit value can be set.
The same as the procedure in section 3.3.1.2(Setting the preset value), enter a
numeric value, then apply it with the [SET/ZERO] key to set the lower limit value.
6
7
To cancel the tolerance judgment, switch the blinking symbol to “off” when “off” or “on”
8
is blinking, and then quickly press the [SET/ZERO] key.
NOTE  If tolerance limit values have been erroneously set so as to be an upper limit value < a
lower limit value, “Error90” will be displayed. Press the [SET/ZERO] key shortly to clear
the error, and then set tolerance limit values so as to be an upper limit value > a lower
limit value.
 Press and hold the [MODE] key for a long time in the process of tolerance setting to
restore the state before setting.
Run-out peak hold during display of “TIR”
3.3.2.2
Hold the runout of fluctuating measurement values according to the following
procedure.
When the tolerance judgment condition has been set, the instrument displays the
tolerance judgment result for the measured runout.
1
2
3
NOTE 


Quickly press the [MODE] key repeatedly until “TIR” is displayed.
Press the [SET/ZERO] key shortly to blink the “TIR” symbol. Then, the runout being
held is canceled and a new run-out hold measurement starts.
When the spindle is displaced, the “TIR” symbol changes from blinking to lighting-up
and a runout is held.
In the tolerance judgment measurement in the run-out hold mode, tolerancing is
performed by comparing the width between the upper and lower limit values that have
been set (upper limit value – lower limit value) and each TIR measurement value.
Start a run-out, maximum value, or minimum value hold measurement with the contact
point placed on a work-piece to be measured. Note that this instrument detects even a
displacement due to vibration or impact with a maximum response speed of 1m/sec.
When switching to “TIR”, “Max”, and “Min” sequentially by pressing the [MODE] key
quickly under the state where the run-out, maximum value, and minimum value have
been held, each peak value can be confirmed.
Maximum value peak hold during display of “Max”
3.3.2.3
Hold the maximum value of fluctuating measurement values according to the following
procedure.
When the tolerance judgment condition has been set, the instrument displays the
tolerance judgment result for the measured maximum value.
Quickly press the [MODE] key repeatedly until “Max” is displayed.
Press the [SET/ZERO] key shortly to blink the “Max” symbol. Then, the maximum
value being held is canceled and a new maximum value hold measurement starts.
3 When the spindle is displaced, the “Max” symbol changes from blinking to lighting-up
and the maximum value is held.
NOTE  Press and hold the [SET/ZERO] key for a long time during measurement in the Preset
system. Then, the “INC” symbol lights up and the preset position is canceled. It is
possible, thereby, to perform measurement on the basis of the maximum value.(peak
zero set)
1
2
No. 99MAH016B
3-7
3.3.2.4
Minimum value peak hold during display of “Min”
Hold the minimum value of fluctuating measurement values according to the following
procedure.
When the tolerance judgment condition has been set, the instrument displays the
tolerance judgment result for the measured minimum value.
Quickly press the [MODE] key repeatedly until “Min” is displayed.
Press the [SET/ZERO] key shortly to blink the “Min” symbol. Then, the minimum value
being held is canceled and a new minimum value hold measurement starts.
3 When the spindle is displaced, the “Min” symbol changes from blinking to lighting-up
and the minimum value is held.
NOTE  Press and hold the [SET/ZERO] key for a long time during measurement in the Preset
system. Then, the “INC” symbol lights up and the preset position is canceled. It is
possible, thereby, to perform measurement on the basis of the minimum value.(peak
zero set)
1
2
TIP
 The difference of display and tolerance judgment in each measurement mode is shown
below.
c
10.00
0
Upper limit 8.000
a
0
d
Lower limit
-3.000
0.00
0
-5.00
0
b
Difference of display in TIR, Max., Min. mode
→
→
0
a
0.00
5.00
Normal
TIR
0.00
5.00
Max.
0.00
5.00
Min.
0.00
5.00
0
b
-5.00
10.00
→
c
10.00
→
d
0.00
15.00
10.00
-5.00
Difference of tolerance judgment in each mode(Upper limit 8.00, Lower limit -3.00)
0
a
b
c
d
Normal
TIR
Max.
Min.
3-8
No. 99MAH016B
3. OPERATING PROCEDURE
3.3.3
Data output / Data hold
Output data to an external device or hold data according to the following procedure.
1
Press the [DATA/Fn] key quickly. Data is outputted to an external device if it is
connected with the connecting cable.
2 Data is held unless an external device is connected.
TIP  For detailed information about data input/output, refer to chapter 4.
 Data hold is available only when the SPC (digimatic) is set.
3.3.4
Function lock
To prevent wrong key operation, disable key entries except for peak hold cancel, data
output, data hold, and hold cancel according to the following procedure.
1
Press and hold the [DATA/Fn] key for a long time in each measurement mode. Then,
the “Lock” symbol lights up, putting the instrument in the function lock state.
2 To cancel the function lock state, press and hold the [DATA/Fn] key again for a long
time. The “Lock” symbol goes off, releasing the instrument from the function lock
state.
NOTE  Function lock is canceled automatically when turning off the power of the instrument.
3.3.5
Centering the pointer
If the pointer is located outside the analog display range, move the pointer display
position to the center of the scale. This centering is the same operation in which the
bezel of a dial indicator is shifted around an arbitrary range of the dial scale.
NOTE 



3.4
Press the [SET/ZERO] key shortly while holding down the [DATA/Fn] key in each
measurement mode to move the pointer to the center of the scale.
In the normal measurement mode, locate the pointer so that the current measurement
position comes to the center of the scale.
In the runout hold mode, bring the center position of runout to the center of the scale.
In the maximum value peak hold mode, bring the center position of the maximum value
to the center of the scale.
In the minimum value peak hold mode, bring the center position of the minimum value
to the center of the scale.
Operation with the Remote Control (Option)
This instrument can be operated using the remote control (option).
Name of each
remote control
Function
switch
INC system (when normal measurement): Zero-setting
SET
Preset system (when normal measurement): Preset recalling
Each peak hold mode: Canceling peak hold (starting measurement)
DATA
External device connected: Data output
External device unconnected: Data hold
MODE
Switching a measurement type in the measurement mode
NOTE  The remote control can identify and operate up to 14 Digimatic indicators in
combination by allocating ID No. to each indicator. For information about ID No. setup
method of this instrument, refer to section 3.2.6.
 Data hold is available only when the SPC (digimatic) is set.
 In the function lock state, switching zero-setting / presetting / measurement mode is
disabled.
No. 99MAH016B
3-9
MEMO
3-10
No. 99MAH016B
44
DATA INPUT/OUTPUT
This chapter explains the data input/output methods of the Digimatic
indicator.
This instrument provides two data communication methods of SPC (Digimatic)
input/output and RS-232 input/output. Select one of these methods to use the data
communication function.
IMPORTANT  Use only the output cable of Mitutoyo specific accessories. The use of an improper
cable or deteriorated cable may disable data output.
 Prior to data output carefully read the user’s manual of the data processing unit to use
it correctly.
4.1
SPC (Digimatic) Input/Output
Connect this instrument with a data processing unit such as Digimatic Mini Processor
DP-1VR using the M-SPC cable (optional accessory) to perform measurement transfer
and data processing including tabulation and record.
4.1.1
Connecting procedure
1
Switch the input/output format to SPC in the Parameter Setup mode. (Refer to section
3.2.)
2 Remove the cap on the I/O connector, and then plug the cable into the connector
firmly. (Pack the removed cap in a small bag and store it in a safe place so as not lose
it.)
TIP  For the method of switching the input/output format, refer to the section “3.2 Parameter
Setup Mode”.
4.1.2
I/O connector
Pin#
9
1
10
2
No. 99MAH016B
Signal
I/O
GND
-
2 (*1)
DATA1
O
3 (*1)
CK
O
4 (*1)
RD
O
5 (*2)
REQ
I
1
6
(N.C)
-
7
(N.C)
-
8
VDD(5V)
O
9
(N.C)
-
10
GND
-
External device side
ID-H side
Vcc
*1
R1
R2
74HC14, etc.
C
For the Vcc=5 system
R1,R2=22kΩ±10%
C=330pF±20%
5V
22kΩ
1kΩ
1000pF
*2
*3
Tr
4-1
IMPORTANT  Since the power voltage differs between the ID-H and external device sides, absolutely
use an open-collector output circuit or open-drain output circuit on the external device
side. Do not use a CMOS output circuit.
 Pin 8 is a terminal dedicated to the RS-232 communication. It is not possible to
transmit from the external device or transmit to it.
4.1.3
Output data format (DATA1)
Decimal position
X.X.X.X.X.X
1(1000)
2(0100)
3(1100)
4(0010)
5(1010)
d1
d2
d3
d4
d5
d6
d7
d8
d9
d10 d11 d12 d13
MSD
All ”F”(1111)
Sign
+:0(0000)
-:8(0001)
LSD
Measurement value
Unit
mm
inch
:0(0000)
:1(1000)
NOTE • Up to 6 digits of numeric values can be outputted as the SPC (Digimatic) output. If the 7
digits of numeric value such as “123.4565mm” is outputted, for example, the
instrument output the SPC data as “23.4565mm”.
4.1.4
Timing chart
0.7 ms < T1 < 1.2 ms
100 µs < T2 < 200 µs
200 µs < T3 < 300 µs
100 µs < T4 < 200 µs
T5: It depends on the capacity of
RD
REQ
DATA
CK
T5
T1
T2
T3
T4
the data processing device
connected.
IMPORTANT
• If the output request (REQ) signal is received during the spindle displacement or if
the inverter receives output request (REQ) signals repeatedly at short intervals, the
instrument may not output data.
NOTE
• Retain the REQ signal at Low level until the CK signal is outputted. Also, return the
REQ signal to High level before the last CK signal (at the 52nd bit) is outputted.
4-2
No. 99MAH016B
4. DATA INPUT/OUTPUT
4.2
RS-232 Input/Output
Connect this instrument with an external device such as a PC using the RS-232 dedicated
cable (optional accessory). With this connection it is possible to perform initial setup of the
instrument, control of switching a measurement mode, etc., and processing such as
measurement data transfer.
Also, multiple Digimatic indicators can be controlled from one RS-232 port on the external
device by setting ID No. on each indicator.
4.2.1
Connecting procedure
1
2
Switch the I/O format to RS-232 in the Parameter Setup mode, and then set the
communication speed (baud rate), parity, and data bit according to the used external
device. (Refer to section 3.2.)
Replace the input/output connector cap and plug the cable securely. (The replaced
cap should be kept in a small bag to prevent losses.)
NOTE  Before a cable is connected with or disconnected, be sure to turn off this instrument.
 If data input/output is not performed properly, there is a possibility of an error in
communication setup. Check the settings of this instrument and the connected device.
TIP  For the method of switching the input/output format, refer to the section “3.2 Parameter
Setup Mode”.
4.2.2
I/O connector
Connector specification: D-sub 9-pin female (receptacle)/inch thread specification
Pin#
1
2
3
4
5
6
7
8
9
Signal
(N.C)
TxD
RxD
DSR
S.G
DTR
CTS
RTS
(N.C)
I/O
O
I
I
O
I
O
-
Definition
Transmit data
Receive Data
Data Set Ready
Signal Ground
Data Terminal Ready
Clear to Send
Request to Send
-
NOTE  The above table lists the pin assignment of the dedicated cable connector.
4.2.3
Communication specification
Item
Home position
Compliant standard
Communication method
Communication speed (bit rate)
Data length
Parity control
Stop bit
Control signal
Specification
DCE (Modem definition),
Dedicated cable must be used.
EIA/TIA-232F (RS-232)
Half duplex
4800 or 9600 bps
7 or 8bit / ASCII / Upper case
None, even, or odd
2 bit
CTS, DSR, or no control sequence
TIP  For information about the switching method of bit rate, data length, and parity control,
refer to section 3.2.5.
No. 99MAH016B
4-3
4.2.4
Communication command (I/O format)
1
Measurement command
Operation content
Zero-setting
(Moves to the INC
system)
Input
(External device -> this instrument)
CR**CRLF
Measurement Moves to the Preset
DS**,PCRLF
system
system
Output
(This instrument -> external device)
CH**CRLF
DH**,PRESETCRLF
(*Refer to the “NOTE” below)
DH**CRLF
Measurement
mode
switching
Preset value setting CP**,+0016.2345CRLF
CH**CRLF
Preset value output DP**,OUTCRLF
DH**,+0016.2345CRLF
Normal
CN**CRLF
Max.
CX**CRLF
Min.
CM**CRLF
TIR (Run-out)
CW**CRLF
Normal value
Data output
Peak hold
Max. value
Min. value
GN**,+0016.2345CRLF
GA**CRLF
CH**CRLF
Peak zero set
DS**,XM-ZEROCRLF
DH**CRLF
Judgment ON
DJ**,ONCRLF
Judgment cancel
DJ**,OFFCRLF
GW**,+0016.2345CRLF
CD**,-0016.2345CRLF
Upper limit value output DJ**,GOUTCRLF
Lower limit value
output
DJ**,DOUTCRLF
Result output (OK)
Result output (-NG)
Status output
(under cancel)
Centering
DH**CRLF
CH**CRLF
DG**,+0016.2345CRLF
DD**,-0016.2345CRLF
DH**,OKCRLF
Result output (+NG)
Function lock
GM**,-0016.2345CRLF
CL**CRLF
Lower limit value input
Analog display
GX**,+0016.2345CRLF
TIR (run-out) value
Cancel
(Peak hold start)
Upper limit value input CG**,+0016.2345CRLF
Tolerance
judgment
CH**CRLF
DH**,+NGCRLF
DJ**,OUTCRLF
DH**,-NGCRLF
DH**,JOFFCRLF
DA**CRLF
Lock
DF**,LOCKCRLF
Cancel
Status output
(under lock)
Status output
(under cancel)
DF**,FREECRLF
DH**CRLF
DH**CRLF
DH**,F-LOCKCRLF
DF**,OUTCRLF
DH**,F-FREECRLF
NOTE  For the detail of each item in the operation content, refer to Chapter 3.
 “CRLF” represents “CR” (carriage return) and “LF” (line feed).
 “ * * ” indicates an ID No. For information about ID No. setup method of this
instrument, refer to section 3.2.
 Set an ID No. between 00 and 99.
 Digimatic indicators with a specified ID No. between 01 and 99 can receive and execute a
command.
4-4
No. 99MAH016B
4. DATA INPUT/OUTPUT
 A Digimatic indicator with the ID No. of 00 can receive and execute a command irrespective
of specification of any ID No.
 After power is turned on and moves to the preset system for the first time, the
communication command of “DH**, PRESETCRLF” is indicated (at this time, the preset
value is not determined). Continue to input the preset value.
 When performing the setting of the preset value in the INC system, the measurement mode
is moved to the preset system automatically.
 When moving from the INC system to the preset system, each peak hold mode is canceled.
 Specify a preset value and an upper/lower limit value with 10 digits of sequence including a
sign and a decimal point. However, the number next to the sign is fixed to “0”. A display
value is also outputted with the same sequence.
 For the presetting and the tolerance value setting, input the value and the decimal point
position with respect to the resolution.
 "Error 90" judgment is performed by comparing the new tolerance value with the current
tolerance value.
 After receiving the output in response to a command, transmit a next command. If there is
no response to a command, clear the communication buffer, then transmit the command
again after a lapse of one second or more.
 This instrument executes key operation on a top-priority basis. The instrument temporarily
stops the RS-232 communication function during key operation, and then executes
command and data output upon restoration of the countable state.
2
Parameter setup command
Operation content
Switching unit
Switching
resolution
Switching
analog range
Switching
count
direction
Setup when
startup
mm
inch
Status output
(mm)
Status output
(inch)
0.0005mm
0.001mm
.00002”
.00005”
.0001”
Status output
Range setting
Status output
+ direction
- (rev.) direction
Status output (+)
Status output (-)
Input
(External device -> this
instrument)
DU**,MMCRLF
DU**,INCRLF
DH**CRLF
DH**,MMCRLF
DU**,OUTCRLF
DH**,INCRLF
DR**,D0.000500CRLF
DR**,D0.001000CRLF
DR**,D0.000020CRLF
DR**,D0.000050CRLF
DR**,D0.000100CRLF
DR**,DOUTCRLF
DR**,A##CRLF
DR**,AOUTCRLF
DD**,NORMCRLF
DD**,REVCRLF
DD**,OUTCRLF
DH**CRLF
DH**,D0.000500CRLF
DH**CRLF
DH**,A##CRLF
DH**CRLF
DH**,NORMCRLF
DH**,REVCRLF
Start setting (INC) DS**,SINCRLF
Start setting
DS**,SFREECRLF
(Setup during stop)
INC / INC start
Measurement INC / start during
stop
system/Setup
DS**,OUTCRLF
output when
P / INC start
startup
P / start during
stop
DE**,RESETCRLF
Return initial setting
No. 99MAH016B
Output
(This instrument -> external
device)
DH**CRLF
DH**,INC-SINCRLF
DH**,INC-SFREECRLF
DH**,P-SINCRLF
DH**,P-SFREECRLF
DE**,RESETCRLF
4-5
IMPORTANT
 For the “Initial setting” by RS-232 input, setting of both “Input/output format” and “ID
No.” are retained.
NOTE
 For the detail of each item in the operation content, refer to Chapter 3.
 “A##” indicates the setting command of the analog range. For information about the
analog range, refer to section 3.2.3. Switching the analog range (analog resolution)
during display of “range”
3
Error command
Operation content
Input
Output
(External device -> this instrument)
(This instrument -> external device)
Over-speed
CH**,Error 20CRLF
Overflow
CH**,Error 30CRLF
Communication
CH**,Error 52CRLF
command error
Tolerance setup
Error
―
error
CH**,Error 95PCRLF
Preset value error
Upper limit
CH**,Error
value error
Lower limit
CS**CRLF
CH**CRLF
 “ ” indicates a space.
 For detailed information about errors, refer to “5 ERROR MASSAGE AND ACTION”.
NOTE
4.2.5
95GCRLF
CH**,Error 95DCRLF
value error
Cancel
CH**,Error 90CRLF
Communication sample
The example of RS-232 communication by using the hyper terminal (attached in the
Windows) is described below.
1
Start up the hyper terminal and select the setting of the digimatic indicator (ID-H side)
to the “Port setting”.
It is also possible to select “Port setting” by clicking on “File”  ”Properties”
”Connect to” and press the “Configure…” button.
When it is necessary to set the initial setting, set “Port setting” as follows.
“Bits per second”:9600bps, “Data bit”:7bit, “Parity”: even
“Stop bits”:2bit, “Flow control”:Hardware
2
Continue to select some of the settings for the hyper terminal.
Click on “File”  ”Properties”  ”Settings” to change the setting as follows.
「Emulation」:ANSI
Click on “ASCII setup” to put the checking mark for check boxes: “Send line ends
with line feeds”, “Echo typed characters locally”, “Append line feeds to incoming line
ends” and “Wrap lines that exceed terminal width”.
3
Example of communication
The procedure below is the case when the tolerance judgment is performed after presetting
4-6
No. 99MAH016B
4. DATA INPUT/OUTPUT
+12.3455mm and setting the upper limit for 12.35mm and the lower limit 12.34mm in the
preset system. For this time, ID No. is “00” (default).
(1) Perform presetting: Input the command “CP00,+0012.3455” from PC. This
instrument performs presetting and responds the command “CH00”.
(2) Set Upper limit: Input the command “CG00,+0012.3500” from PC. This instrument
sets the upper limit and responds the command “CH00”.
(3) Set Lower limit: Input the command “CD00,+0012.3400” from PC. This instrument
sets the lower limit and responds the command “CH00”.
(4) Set the tolerance judgment mode. Input the command “DJ00,ON” from PC. This
instrument changes the measurement mode to the tolerance judgment mode and
TIP
No. 99MAH016B
responds the command “DH00”.
 For the example of RS-232 communication described above, perform presetting (change of
the measurement system) at the beginning. This instrument changes the upper and the
lower limit of the measurement mode when inputting the command.
 For the procedure (2) to (4), it is allowed to proceed with the random order.
 When using the hyper terminal, it is not possible to correct the command typing error. Once
input the incorrect command (this instrument responds the command “CH**,Error 52”)
and then input the correct command.
4-7
MEMO
4-8
No. 99MAH016B
55
ERROR MESSAGE AND
ACTION
This chapter describes the error messages and their
corrective actions.
If any error occurs in this Digimatic indicator, a corresponding error message is
displayed. (While RS-232 communication is being performed, an error message is
also outputted. In the Digimatic code output mode an error message will not be
outputted.)
If the instrument does not recover the normal conditions after corrective actions,
contact your dealer or Mitutoyo sales office.
1. Errors possibly caused in normal operation
RS-232
Definition
Action
Message
output
Error20
Error30
(Normal
count)
Error52
Error90
Error95P
Error95G
Error95D
Over-speed
 The spindle operation
speed has exceeded
the response speed of
this instrument.
Display overflow
 The display value has
exceeded the
maximum number of
digits that can be
displayed.
RS-232 command input
error
 A wrong command
was inputted.
Tolerance setup error 
 The limit values are set
so as to be an upper
limit value < a lower
limit value.
Preset value setting
error
 The preset value has
over flown.
Upper limit value setting
error
 The upper limit value
has over flown.
Lower limit value setting
error
 The lower limit value
has over flown.
 Press the [SET/ZERO] key.
 Use this instrument at a speed so that
the spindle operation speed is within the
response speed.
 Return the display value to the
permissible number of digits to
automatically reset the error.
 Press the [SET/ZERO] key quickly or
hold it down slowly. *1
 Enter the setup mode with the [MODE]
key, and then reduce the resolution. *2
 Re-enter the correct command.
 Press the [SET/ZERO] key quickly, and
then set tolerance limit values so as to
be an upper limit value > a lower limit
value.
 Set a preset value again. *3
 Reduce the resolution. *3
 Set an upper limit value again. *4
 Reduce the resolution. *3
 Set a lower limit value again. *4
 Reduce the resolution. *3
*1. Press the [SET/ZERO] key quickly (less than 2 seconds) to switch to the
Preset Setup mode, or press and hold the key slowly (2 seconds or
more) to perform zero-setting in the INC measurement system.
No. 99MAH016B
5-1
*2. Press and hold the [MODE] key slowly to switch to the Parameter Setup
mode.
*3. Press the [MODE] key quickly to clear the error using the procedure in *1
and *2 “Error 30” message.
*4. Press and hold the [MODE] key slowly to blink the “<>” symbol and
make tolerance setup ready, or press and hold the [SET/ZERO] key
slowly to display a maximum permissible display value and set tolerance
limit values.
NOTE  If data input/output is not performed properly, there is a possibility of an error
in communication setup. Check the settings of this instrument and the
connected device.
 For detailed information about the RS-232 output command, refer to section
4.2.
 If any of the above errors occurs during function lock, press and hold the
[DATA/Fn] key for 2 seconds or more to cancel the function lock, and then
perform the error clear processing.
5-2
No. 99MAH016B
66
6.1
SPECIFICATION
Specifications of the main unit
Model name
Order No. *1
ID-H0530
543-561-1
ID-H0560
543-563-1
Resolution
0.0005mm / 0.001mm
Measuring range 30.48mm
Accuracy(20℃) 0.0015mm or less
*2
Standards
ID-H0530E
543-562-1
ID-H0560E
543-564-1
0.0005mm / 0.001mm
.00002” / .00005” / .0001”
60.96mm
30.48mm = 1.2”
60.96mm = 2.4”
0.0025mm or less
0.0015mm = .00006”
or less
0.0025mm = .0001”
or less
Stem diameter
ISO R463 / JIS B7503 / DIN 878
φ8mm
ANSI B89.1.10 / AGD
φ9.52mm = 3/8”DIA
Contact point
Carbide SR1.5 (M2.5x0.45)
Carbide SR1.5 (#4-48UNF)
Contact force
Maximum
response speed
2.0N or less
2.0N or less
Protection
Equivalent IP-30(at conditions ex-works)
Protection level according to IEC60529 / JIS D0207, C0920.
Plunger direction
Below the horizon
AC adapter(6V, 2A)
Power supply
2.5N or less
1000mm/sec
Operating temp
200mA or less
0℃~40℃
Storage temp
-10℃~60℃
Net weight
Approx. 305g (0.67 lbs)
Consumption current
2.5N or less
Approx. 290g (0.63 lbs)
Approx. 305g (0.67 lbs)
Approx. 290g (0.63 lbs)
*1:This Order No. mean only the main unit without AC adapter.
*2:Not including the quantizing error.
6.2
Standard accessories
# 99MAH016B
User’s manual
# 99MAH018B
Quick reference manual
# 137693
Lifting lever
# 21EAA150
Clamp filter
# 06AEG180JA AC adapter (100-240V : Japan, USA, Canada, & Co.)
# 06AEG180D
AC adapter (100-240V : Germany, & Co.)
# 06AEG180E
AC adapter (100-240V : UK, & Co.)
# 06AEG180K
AC adapter (100-240V : Korea, for KC)
# 06AEG180DC AC adapter (100-240V : China, for CCC)
Certification of inspection
No. 99MAH016B
6-1
6.3
6-2
Optional accessories
#21EZA099
Remote Controller
#540774
Spindle Lifting cable
#21EZA101
Lifting knob
#936937
M-SPC cable 1m/40”
#965014
M-SPC cable 2m/80”
#21EAA131
RS-232 cable 2m/80”
#101040
Lug-on-center back(for ISO/JIS model)
#101306
Lug-on-center back(for ANSI/AGD model)
No. 99MAH016B
SERVICE NETWORK
Europe
Mitutoyo Europe GmbH
Borsigstrasse 8-10, 41469 Neuss, GERMANY
TEL:49(2137)102-0 FAX:49(2137)102-351
Mitutoyo CTL Germany GmbH
Neckarstrasse 1/8, 78727 Oberndorf, GERMANY
TEL:49(7423) 8776-0 FAX:49(7423)8776-99
KOMEG Industrielle Messtechnik GmbH
Zum Wasserwerk 3, 66333 Völklingen, GERMANY
TEL: 49(6898)91110 FAX: 49(6898)9111100
Germany
Mitutoyo Deutschland GmbH
Borsigstrasse 8-10, 41469 Neuss, GERMANY
TEL:49(2137)102-0 FAX:49(2137)86 85
M3 Solution Center Hamburg
Tempowerkring 9·im HIT-Technologiepark 21079 Hamburg, GERMANY
TEL:49(40)791894-0 FAX:49(40)791894-50
M3 Solution Center Leonberg GmbH
Steinbeisstrasse 2, 71229 Leonberg, GERMANY
TEL:49(7152)6080-0 FAX:49(7152)608060
M3 Solution Center Berlin
Paradiesstrasse 208, 12526 Berlin, GERMANY
TEL:49(30)2611 267 FAX:49(30)26 29 209
M3 Solution Center Eisenach
im tbz Eisenach, Heinrich-Ehrhardt-Platz, 99817 Eisenach, GERMANY
TEL:49(3691)88909-0 FAX:49(3691)88909-9
M3 Solution Center Ingolstadt
Marie-Curie-Strasse 1a, 85055 Ingolstadt, GERMANY
TEL:49(841)954920 FAX:49(841)9549250
Mitutoyo-Messgeräte Leonberg GmbH
Heidenheimer Strasse 14, 71229 Leonberg, GERMANY
TEL:49(7152)9237-0 FAX:49(7152)9237-29
U.K.
Mitutoyo (UK) L.td.
Joule Road, West Point Business Park, Andover, Hampshire SP10 3UX,
UNITED KINGDOM TEL:44(1264)353123 FAX:44(1264)354883
M3 Solution Center Coventry
Unit6, Banner Park, Wickmans Drive, Coventry, Warwickshire CV4 9XA,
UNITED KINGDOM TEL:44(2476)426300 FAX:44(2476)426339
M3 Solution Center Halifax
Lowfields Business Park, Navigation Close, Elland, West Yorkshire HX5 9HB,
UNITED KINGDOM TEL:44(1422)375566 FAX:44(1422)328025
M3 Solution Center East Kilbride
The Baird Bulding, Rankine Avenue, Scottish Enterprise Technology Park, East
Killbride G75 0QF, UNITED KINGDOM
TEL:44(1355)581170 FAX:44(1355)581171
France
Mitutoyo France
Paris Nord 2-123 rue de la Belle Etoile, BP 59267 ROISSY EN FRANCE 95957
ROISSY CDG CEDEX, FRANCE TEL:33(1) 49 38 35 00 FAX:33(1) 48 63 27 70
M3 Solution Center LYON
Parc Mail 523, cours du 3éme millénaire, 69791 Saint-Priest, FRANCE
TEL:33(1) 49 38 35 70 FAX:33(1) 49 38 35 79
M3 Solution Center STRASBOURG
"Parc Mail" 523, cours du 3ème millénaire, 69791 Saint-Priest Cedex, FRANCE
TEL:33(1) 49 38 35 80 FAX:33(1) 49 38 35 89
M3 Solution Center CLUSES
Espace Scionzier 480 Avenue des Lacs, 74950 Scionzier, FRANCE
TEL:33(1) 49 38 35 90 FAX:33(1) 49 38 35 99
M3 Solution Center TOULOUSE
Aeroparc Saint-Martin ZAC de Saint Martin du Touch 12 rue de Caulet, 31300
Toulouse, FRANCE TEL:33(5)82 95 60 69
Italy
MITUTOYO ITALIANA S.r.l.
Corso Europa, 7 - 20020 Lainate (Ml), ITALY
TEL: 39(02)935781 FAX:39(02)9373290•93578255
M3 Solution Center VERONA
Via A. Volta, 37062 Dosso Buono (VR), ITALY
TEL:39(045)513012 FAX:39(045)8617241
M3 Solution Center TORINO
Via Brandizzo, 133/F - 10088 Volpiano (TO), ITALY
TEL:39(0)11 9123995 FAX:39(0)11 9953202
M3 Solution Center CHIETI
Contrada Santa Calcagna - 66020 Rocca S. Giovanni (CH), ITALY
TEL/FAX:39(0872)709217
Netherlands
Mitutoyo Nederland B.V.
Storkstraat 40, 3905 KX Veenendaal, THE NETHERLANDS
TEL:31(0)318-534911 FAX:31(0)318-534811
Mitutoyo Research Center Europe B.V.
De Rijn 18, 5684 PJ Best, THE NETHERLANDS
TEL:31(0)499-320200 FAX:31(0)499-320299
Belgium
Mitutoyo Belgium N.V.
Hogenakkerhoek straat 8, 9150 Kruibeke, BELGIUM
TEL:32(0)3-2540444 FAX:32(0)3-2540445
Sweden
Mitutoyo Scandinavia AB
Släntvägen 6, 194 54 Upplands Väsby, SWEDEN
TEL:46(0)8 594 109 50 FAX:46(0)8 590 924 10
M3 Solution Center Alingsas
Kristineholmsvägen 26, 441 39 Alingsas, SWEDEN
*As of October 2013
TEL:46(0)8 594 109 50 FAX:46(0)322 63 31 62
M3 Solution Center Värnamo
Storgatsbacken 9, 331 30 Värnamo, SWEDEN
TEL:46(0)8 594 109 50 FAX:46(0)370 463 34
Finland
Mitutoyo Scandinavia AB Finnish Branch
Viherkiitäjä 2A, FI-33960, Pirkkala, Finland
TEL: +358 207 929 640
Switzerland
Mitutoyo Schweiz AG
Steinackerstrasse 35, 8902 Urdorf, SWITZERLAND
TEL:41(0)447361150 FAX:41(0)447361151
Poland
Mitutoyo Polska Sp.z o.o.
ul.Minska 54-56, 54-610 Wroclaw, POLAND
TEL:48(71)354 83 50 FAX:48(71)354 83 55
Czech Republic
Mitutoyo Cesko, s.r.o.
Dubska 1626, 415 01 Teplice, CZECH REP
TEL:420-417-579-866 FAX:420-417-579-867
Hungary
Mitutoyo Hungária Kft.
Németvölgyi út 97, H-1124 Budapest, HUNGARY
TEL:36(1)2141447 FAX:36(1)2141448
Romania
Mitutoyo Romania SRL
1A, Drumul Garii Odai Street, Ground Floor, Room G03, 075100
OTOPENI-ILFOV, ROMANIA TEL:40(0)311012088 FAX:40(0)311012089
Russian Federation
Mitutoyo RUS LLC
13 Sharikopodshipnikovskaya, bld.2, 115088 Moscow, RUSSIAN FEDERATION
TEL:(7)495 7450 752 FAX:(7)495 745 0752
Mitutoyo Austria GmbH
Johann Roithner Straße 131 A-4050 Traun
TEL:+43(0)7229/23850 FAX:+43(0)7229/23850-90
Singapore
Mitutoyo Asia Pacific Pte. Ltd.
Head office / M3 Solution Center
24 Kallang Avenue, Mitutoyo Building, SINGAPORE 339415
TEL:(65)62942211 FAX:(65)62996666
Malaysia
Mitutoyo (Malaysia) Sdn. Bhd.
Kuala Lumpur Head Office / M3 Solution Center
Mah Sing Intergrated Industrial Park, 4, Jalan Utarid U5/14, Section U5, 40150
Shah Alam, Selangor, MALAYSIA TEL:(60)3-78459318 FAX:(60)3-78459346
Penang Branch office / M3 Solution Center
No.30, Persiaran Mahsuri 1/2, Sunway Tunas, 11900 Bayan Lepas, Penang,
MALAYSIA TEL:(60)4-6411998 FAX:(60)4-6412998
Johor Branch office / M3 Solution Center
No. 70, Jalan Molek 1/28, Taman Molek, 81100 Johor Bahru, Johor, MALAYSIA
TEL:(60)7-3521626 FAX:(60)7-3521628
Thailand
Mitutoyo(Thailand)Co., Ltd.
Bangkok Head Office / M3 Solution Center
No. 76/3-5, Chaengwattana Road, Anusaowaree, Bangkaen, Bangkok 10220,
THAILAND TEL:(66)2-521-6130 FAX:(66)2-521-6136
Cholburi Branch / M3 Solution Center
No.7/1, Moo 3, Tambon Bowin, Amphur Sriracha, Cholburi 20230, THAILAND
TEL:(66)3-834-5783 FAX:(66)3-834-5788
Amata Nakorn Branch / M3 Solution Center
No. 700/199, Moo 1, Tambon Ban Kao, Amphur Phan Thong, Cholburi 20160,
THAILAND TEL:(66)3-846-8976 FAX:(66)3-846-8978
Indonesia
PT. Mitutoyo Indonesia
Head Office / M3 Solution Center
Ruko Mall Bekasi Fajar Blok A6&A7 MM2100 Industrial Town, Cikarang Barat,
Bekasi 17520, INDONESIA TEL:(62)21-8980841 FAX:(62)21-8980842
Vietnam
Mitutoyo Vietnam Co., Ltd
Hanoi Head Office / M3 Solution Center
No.34-TT4, My Dinh-Me Tri Urban Zone, My Dinh Commune, Tu Liem District,
Hanoi, VIETNAM TEL:(84)4-3768-8963 FAX:(84)4-3768-8960
Ho Chi Minh City Branch Office / M3 Solution Center
31 Phan Xich Long Street, Ward 2, Phu Nhuan District, Ho Chi Minh City,
VIETNAM TEL:(84)8-3517-4561 FAX:(84)8-3517-4582
India
Mitutoyo South Asia Pvt. Ltd.
Head Office / M3 Solution Center
C-122, Okhla Industrial Area, Phase-l, New Delhi-110 020, INDIA
TEL:91(11)2637-2090 FAX:91(11)2637-2636
Mumbai Region Head office
303, Sentinel Hiranandani Business Park Powai, Mumbai-400 076, INDIA
TEL:91(22)2570-0684, 837, 839 FAX:91(22)2570-0685
Pune Office / M3 Solution Center
G2/G3, Pride Kumar Senate, F.P. No. 402 Off. Senapati Bapat Road, Pune-411 016,
INDIA TEL:91(20)6603-3643, 45, 46 FAX:91(20)6603-3644
Vadodara office
S-1&S-2, Olive Complex, Nr. Haveli, Nizampura, Vadodara-390 002, INDIA
TEL: (91) 265-2750781 FAX: (91) 265-2750782
Bengaluru Region Head office / M3 Solution Center
No. 5, 100 Ft. Road, 17th Main, Koramangala, 4th Block, Bengaluru-560 034,
INDIA TEL:91(80)2563-0946, 47, 48 FAX:91(80)2563-0949
SERVICE NETWORK
Chennai Office / M3 Solution Center
No. 624, Anna Salai Teynampet, Chennai-600 018, INDIA
TEL:91(44)2432-8823, 24, 27, 28 FAX:91(44)2432-8825
Kolkata Office
Unit No. 1208,Om Tower, 32,J.L.Nehru Road, Kolkata-700 071, INDIA
Tel: (91) 33-22267088/40060635 Fax: (91) 33-22266817
Taiwan
Mitutoyo Taiwan Co., Ltd.
4F., No.71, Zhouzi St., Neihu Dist.,Taipei City 114, TAIWAN (R.O.C.)
TEL:886(2)8752-3266 FAX:886(2)8752-3267
Taichung Branch
16F.-3, No.6, Ln.256, Sec.2, Xitun Rd., Xitun Dist., Taichung City 407, TAIWAN
(R.O.C.) TEL:886(4)2707-1766 FAX:886(4)2451-8727
Kaohsiung Branch
13F.-3, No.31, Haibian Rd., Lingya Dist., Kaohsiung City 802, TAIWAN (R.O.C.)
TEL:886(7)334-6168 FAX:886(7)334-6160
M3 Solution Center Taipei
4F., No.71, Zhouzi St., Neihu Dist., Taipei City 114,TAIWAN (R.O.C.)
TEL:886(2)8752-3266 FAX:886(2)8752-3267
M3 Solution Center Tainan
Rm.309, No.31, Gongye 2nd Rd., Annan Dist., Tainan City 709, TAIWAN (R.O.C.)
TEL:886(6)384-1577 FAX:886(6)384-1576
South Korea
Mitutoyo Korea Corporation
Head Office / M3 Solution Center
(Sanbon-Dong, Geumjeong High View Build.), 6F, 153-8, Ls-Ro, Gunpo-Si,
Gyeonggi-Do, 435-040 KOREA TEL:82(31)361-4200 FAX:82(31)361-4202
Busan Office / M3 Solution Center
Donghum Build. 1F, 559-13 Gwaebop-Dong, Sasang-Gu, Busan, 617-809, KOREA
TEL:82(51)324-0103
FAX:82(51)324-0104
Daegu Office / M3 Solution Center
371-12, Hosan-Dong, Dalseo-Gu, Daegu, 704-230, KOREA
TEL:82(53)593-5602 FAX:82(53)593-5603
China
Mitutoyo Measuring Instruments (Shanghai) Co., Ltd.
12F, Nextage Business Center, No.1111 Pudong South Road, Pudong New
District ,Shanghai 200120, CHINA TEL:86(21)5836-0718 FAX:86(21)5836-0717
Suzhou Office / M3 Solution Center China (Suzhou)
No. 46 Baiyu Road, Suzhou 215021, CHINA
TEL:86(512)6522-1790 FAX:86(512)6251-3420
Wuhan Office
RM. 1206B Wuhan World Trade Tower, No. 686, Jiefang Ave, Jianghan District,
Wuhan 430032, CHINA TEL:86(27)8544-8631 FAX:86(27)8544-8227
Chengdu Office
1-705, New Angle Plaza, 668# Jindong Road, Jinjiang District, Chengdu, Sichuan
610066,CHINA TEL:86(28)8671-8936 FAX:86(28)8671-9086
Hangzhou Office
RM. A+B+C 15/F, TEDA Building, No.256 Jie-fang Nan Road Hexi District,Tianjin
300042, CHINA TEL:86(22)5888-1700 FAX:86(22)5888-1701
Tianjin Office / M3 Solution Center Tianjin
No.16 Heiniucheng-Road, Hexi-District, Tianjin 300210, CHINA
TEL:86(22)8558-1221 FAX:86(22)8558-1234
Changchun Office
RM.1801, Kaifa Dasha, No. 5188 Ziyou Avenue, Changchun 130013, CHINA
TEL:86(431)8461-2510 FAX:86(431)8464-4411
Qingdao Office / M3 Solution Center Qingdao
No.135-10, Fuzhou North Road, Shibei District, Qingdao City, Shandong 266034,
CHINA TEL:86(532)8066-8887 FAX:86(532)8066-8890
Xi’an Office
RM. 805, Xi’an International Trade Center, No. 196 Xiaozhai East Road, Xi’an,
710061, CHINA TEL:86(29)8538-1380 FAX:86(29)8538-1381
Dalian Office / M3 Solution Center Dalian
RM.1008, Grand Central IFC, No.128 Jin ma Road,Economic Development
Zone,Dalian 116600, CHINA TEL:86(411)8718 1212 FAX:86(411)8754-7587
Zhengzhou Office
Room1801,18/F,Unit1,Building No.23, Shangwu Inner Ring Road, Zhengdong
New District,Zhengzhou City,Henan Province, 450018,CHINA
TEL:86(371)6097-6436 FAX:86(371)6097-6981
Mitutoyo Leeport Metrology (Hong Kong) Limited
Rm 818, 8/F, Vanta Industrial Centre, No.21-33, Tai Lin Pai Road, Kwai Chung,
NT, Hong Kong TEL:86(852)2992-2088 FAX:86(852)2670-2488
Mitutoyo Leeport Metrology (Dongguan) Limited / M3 Solution Center Dongguan
No.26, Guan Chang Road, Chong Tou Zone, Chang An Town, Dong Guan, 523855
CHINA TEL:86(769)8541 7715 FAX:86(769)-8541 7745
Mitutoyo Leeport Metrology (Dongguan) Limited – Fuzhou office
Rm 2104, City Commercial Centre, No.129 Wu Yi Road N., Fuzhou City, Fujian
Province, CHINA TEL (86) 0591 8761 8095 FAX (86) 0591 8761 8096
Mitutoyo Leeport Metrology (Dongguan) Limited – Changsha office
Rm 2121, Dingwang Building, No.88, Section 2, Furong Middle Road, Changsha
City, Hunan Province, CHINA TEL (86) 731 8872 8021 FAX (86) 731 8872 8001
Mitutoyo Measuring Instruments (Suzhou) Co., Ltd.
No. 46 Baiyu Road, Suzhou 215021, CHINA
TEL:86(512)6252-2660 FAX:86(512)6252-2580
U.S.A.
Mitutoyo America Corporation
965 Corporate Blvd., Aurora, IL 60502, U.S.A.
TEL:1-(630)820-9666 Toll Free No. 1-888-648-8869 FAX:1-(630)820-2614
M3 Solution Center-Illinois
945 Corporate Blvd., Aurora, IL 60502, U.S.A.
M3 Solution Center-Ohio
6220 Hi-Tek Ct., Mason, OH 45040, U.S.A.
TEL:1-(513)754-0709 FAX:1-(513)754-0718
*As of October 2013
M3 Solution Center-Michigan
44768 Helm Street, Plymouth, MI 48170, U.S.A.
TEL:1-(734)459-2810 FAX:1-(734)459-0455
M3 Solution Center-California
16925 E. Gale Ave., City of Industry, CA 91745, U.S.A.
TEL:1-(626)961-9661 FAX:1-(626)333-8019
M3 Solution Center-Massachusetts
1 Park Dr., Suite 11, Westford, MA 01886, U.S.A.
TEL:1-(978)692-8765 FAX:1-(978)692-9729
M3 Solution Center-North Carolina
11515 Vanstory Dr., Suite 150, Huntersville, NC 28078, U.S.A.
TEL:1-(704)875-8332 FAX:1-(704)875-9273
M3 Solution Center-Alabama
2100 Riverchase Center Suite 106 Hoover, AL 35244, U.S.A
TEL;1-(205)-988-3705 FAX;1-(205)-988-3423
CT-Lab Chicago
965 Corporate Blvd., Aurora, IL 60502, U.S.A.
TEL:1-630-820-9666 FAX:1-630-820-2614
Micro Encoder, Inc.
11533 NE 118th St., bldg. M, Kirkland, WA 98034, U.S.A.
TEL:1-(425)821-3906 FAX:1-(425)821-3228
Micro Encoder Los Angeles, Inc.
16925 E. Gale Ave. City of Industry, CA 91745 USA
TEL:1-626-961-9661 FAX:1-626-333-8019
Canada
Mitutoyo Canada Inc.
2121 Meadowvale Blvd., Mississauga, Ont. L5N 5N1., CANADA
TEL:1-(905)821-1261 FAX:1-(905)821-4968
Montreal Office
7075 Place Robert-Joncas Suite 129, Montreal, Quebec H4M 2Z2, CANADA
TEL:1-(514)337-5994 FAX:1-(514)337-4498
Brazil
Mitutoyo Sul Americana Ltda.
AV. Joao Carlos da Silva Borges, 1240 - CEP 04726-002 - Santo Amaro São Paulo - SP, BRASIL TEL:55(11)5643-0000 FAX:55(11)5641-3722
Regional Office
Belo Horizonte - MG
TEL:55(31)3531-5511 FAX:55(31)3594-4482
Rio Grande do Sul / PR, SC
TEL/FAX:55(51)3342-1498 TEL:55(51)3337-0206
Rio de Janeiro - RJ
TEL:55(21)3333-4899 TEL/FAX:55(21)2401-9958
Santa Barbara D’Oeste - SP
TEL:55(19)3455-2062 FAX:55(19)3454-6103
Norte, Nordeste, Centro Oeste
TEL:55(11)5643-0060 FAX:55(11)5641-9029
Escritorio BA / SE
TEL/FAX:55(71)3326-5232
Factory(Suzano)
Rodovia Índio Tibirica 1555, BAIRRO RAFFO, CEP 08620-000 SUZANO-SP,
BRASIL TEL:55(11)4746-5858 FAX:55(11)4746-5936
Argentina
Mitutoyo Sul Americana Ltda.
Argentina Branch
Av. B. Mitre 891/899 – C.P. (B1603CQI) Vicente López –Pcia. Buenos Aires –
Argentina TEL:54(11)4730-1433 FAX:54(11)4730-1411
Sucursal Cordoba
Av. Amadeo Sabattini, 1296, esq. Madrid Bº Crisol Sur – CP 5000, Cordoba,
ARGENTINA TEL/FAX:54 (351) 456-6251
Mexico
Mitutoyo Mexicana, S. A. de C. V
Prolongación Industria Eléctrica No. 15 Parque Industrial Naucalpan
Naucalpan de Juárez, Estado de México C.P. 53370, MÉXICO
TEL: 52 (01-55) 5312-5612, FAX: 52 (01-55) 5312-3380
M3 Solution Center Monterrey
Av. Morones Prieto No 914. Ote., Local 105 - Plaza Malz Col. La Huerta, C.P.
67140 Guadalupe, N.L., México
TEL: 52 (01-81) 8398-8228, 8398-8227 and 8398-8244 FAX: 52 (01-81) 8398-8226
M3 Solution Center Tijuana
Av. 2o. eje Oriente-Poniente No. 19075 Int. 18 Col. Cd. Industrial Nueva Tijuana
C.P. 22500 Tijuana, B. C., México
TEL: 52 (01-664) 624-3644 and 624-3645 FAX: 52 (01-664) 647-5024
M3 Solution Center Querétaro
Acceso “C” No. 107 Col. Parque Industrial Jurica C.P. 76100 Querétaro, Qro.,
México
TEL: 52 (01-442) 340-8018, 340-8019 and 340-8020 FAX: 52 (01-442) 340-8017
Aguascalientes Office / M3 Solution Center
Av. Aguascalientes No. 622, Local 12 Centro Comercial El Cilindro Fracc. Pulgas
Pandas Norte, C.P. 20138, Aguascalientes, Ags. México
TEL: 52 (01-449) 174-4140 FAX: 52 (01-449) 174-4143
Irapuato Office / M3 Solution Center
Av. Héroes de Nacozari No. 1655, local A-14 esq. con Boulevard Villas de lrapuato
"Plaza Delta" Col. San Miguelito, C.P. 36557 Irapuato. Gto., México
TEL: 52 (01-462) 144-1200