13 Troubleshooting
Issue
Low event count
Low assay signal or
sensitivity
High Background
Low Precision
Clogged filter plate
Possible cause
Bead aggregate
Bead settle on
the well bottom
Vacuum too
strong
Standard not
reconstituted
well
Incubation time
too short
Excess exposure
to light
Well-to-well
contamination
Poor pipetting
precision
Contamination
from
adjacent wells
High lipid
content in the
serum, plasma or
bodily fluid
samples
Recommended actions
Vortex stock and working bead suspensions well
before pipetting.
Shake plates at 1000 rpm for 30 seconds prior to
acquisition or transferring to sample tubes.
Adjust the vacuum pressure so that 100 µL of 1x
Wash Buffer in the wells can be clear in 3-5 second.
Standard should be incubated on ice for 5min after
the addition of standard diluent
Follow recommended incubation time in each step
During incubation, cover the plate with aluminum
foil to minimize exposure of the beads to light
Change pipette tips after every transfer. Remove
plate seal carefully and avoid contents from one well
to mix with another.
Use calibrated pipettes.
Avoid well-to-well contamination during pipetting
and removal of plate seal
Centrifuge the samples at 10,000 x g for 10 min at 28 °C. Collect the serum, plasma or bodily fluid
fraction.