Download Avalanche Transportable Sampler User Manual
Transcript
Avalanche Portable Refrigerated Sampler Section 7 SDI-12 Sondes 7.7 Sonde Calibration and Validation You should calibrate your sonde in the lab, according to manufacturer’s recommendations. Note The AQ700 sonde’s PC software program provides calibration for its sensors. Refer to the AQ700 user manual for complete instructions for calibration of each parameter. The Avalanche allows for field calibration of level, pH, Dissolved Oxygen (DO), and Oxidation Reduction Potential (ORP); however, field calibrations are stored in the sampler, not in the sonde. They are used to modify readings from the sonde until a new sonde is connected. When a sonde is moved to another sampler, it will retain internal calibration constants set in the lab, but will not retain any that were set in the field. Although field calibration is limited to four parameters, you can validate any of the available parameters. Validation allows you to immerse the sonde in a known solution and verify the sonde’s readings. The validation process does not modify any readings like calibration does; it merely alerts you to inaccurate ones. No data will be stored, nor will enable conditions be updated, while calibration or validation is being performed. 7.7.1 Calibration Screens To access the calibration screens from the main menu, select Other Functions > Hardware. Advance to SDI-12 DATA: NO SDI-12 SONDE SETUP KNOWN SONDES SCAN SDI-12 SONDES Select SETUP KNOWN SONDES and DONE. The screen will list the parameters you can calibrate, similar to the example below. The sonde’s address will appear after the parameter (such as pH0 and pH5). SELECT PARAMETER TO CALIBRATE: pH5 DO5 ORP5 DONE The parameter selected will determine the content of the calibration screens: • Level - Enter the current measured level. The sampler will use this value to offset the level reading of the sonde. • pH - You will be given the choice of three different pH solutions: 4.00, 7.00, and 10.00. After making your selection, rinse the probe and place in the pH buffer. If the reading is off by more than 0.5 pH units, you will be asked to verify if the reading is correct. • DO - For dissolved oxygen, you can choose between two methods: DO Standard and Saturation. The saturation 7-8